NA 8667961Recalls & reported problems
Richard Wolf GmbH · SCALPEL 20.5FR CERAMIC SEMICIRCULAR for optical urethrotome, telescope Ø 4mm, massive, 12 O'clock cutting position, for slitting of urethral (transurethral) and ureteral (percutaneous) stenosis, reusable · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04055207019245), received between Aug 16, 2018 and Aug 16, 2018 — classified as 1 injury.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Break1
Patient problems
- Device Embedded In Tissue or Plaque1
- Necrosis1
- Swelling1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryAug 16, 2018Break; Break
ON 8/15/2018 RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION, HEREINAFTER REFERRED TO AS RWMIC, WAS MADE AWARE OF AN EVENT FROM RICHARD WOLF (B)(4) HEREINAFTER REFERRED TO AS (B)(4) OF AN INCIDENT THAT OCCURRED AT A HOSPITAL IN (B)(6). (B)(4) WAS NOTIFIED BY (B)(6) AUTHORITIES ON A DATE YET TO BE DETERMINED THAT A SIGNIFICANT INJURY OCCURRED DURING A PROCEDURE IN THE HOSPITAL INVOLVING A (B)(6) MALE PATIENT. (B)(4) ATTEMPTED TO CONTACT THE USER FACILITY TO OBTAIN INFORMATION ON THE DEVICE, PRODUCT CODE, LOT NUMBER BUT NO ADDITIONAL INFORMATION WAS AVAILABLE AT THAT TIME. AN OPTICAL URETHROTOMY WAS PERFORMED ON THE PATIENT ON (B)(6) 2018 AND AT THE VERY BEGINNING OF THE PROCEDURE, JUST AFTER THE URETHROTOMY HAD BEEN PERFORMED, IT WAS NOTED THAT THE BLADE HAD BECOME DISLODGED FROM THE END OF THE INSTRUMENT AND HAD BECOME EMBEDDED IN THE PERI-URETHRAL TISSUES. THE USER MADE REPEATED ATTEMPTS TO FIND AND REMOVE THIS BLADE BUT WAS UNABLE TO DO SO, DESPITE USING FLUOROSCOPIC SCREENING IN THE OPERATING THEATRE. VARIOUS DIFFERENT SCOPES WERE USED AS WELL AS DIFFERENT TYPES OF GRASPERS. THE PATIENT SUFFERED A VERY SIGNIFICANT INJURY AS A RESULT OF THE EQUIPMENT MALFUNCTION AND THE ATTEMPTS MAD
Patient problems: Device Embedded In Tissue or Plaque; Necrosis; Swelling
MAUDE report 7791287 ↗
Other Urethrotome devices
- Olympus A4674AOLYMPUS Winter & Ibe GmbH · Urethrotome1 recall169 reports
- Olympus A4673AOLYMPUS Winter & Ibe GmbH · Urethrotome6 reports
- n.a. 27033W-SKarl Storz GmbH & Co. KG · Urethrotome1 report
- n.a. 27068L-SKarl Storz GmbH & Co. KG · Urethrotome1 report
- USA Elite K-SBGyrus ACMI, LLC · Urethrotome1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.