Skip to content

NA 8564851Recalls & reported problems

Richard Wolf GmbH · CONNECTION CABLE WL 3M reusable · Class 1

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04055207066676), received between Oct 4, 2024 and May 26, 2026 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Disconnection” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Disconnection3
  • Connection Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 26, 2026Disconnection

    IT WAS REPORTED THAT THE PLUGGED CORD ATTACHED TO THE MOTOR AND THE CONSOLE WOULD NOT REGISTER THE PRESENCE OF THE MOTOR DURING A PROCEDURE. NO INJURIES OR UNACCEPTABLE RISKS WERE REPORTED. THERE WAS A SIMILAR BACK UP INSTRUMENT AVAILABLE. NO ADDITIONAL TREATMENT WAS REQUIRED AND THE SCHEDULED PROCEDURE WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25284387 ↗
  • MalfunctionAug 6, 2025Connection Problem; Connection Problem

    THE USER REPORTED THAT THE CONNECTION CABLE DID NOT WORK DURING AN ARTHROSCOPY PROCEDURE. THERE WAS NO REPORT OF ANY HARM TO THE PATIENT, USERS, OR THIRD PARTIES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22730420 ↗
  • InjuryOct 4, 2024Disconnection; Disconnection

    THE USER FACILITY HAS INFORMED RICHARD WOLF MEDICAL CORPORATION (RWMIC) OF AN ISSUE REGARDING A CONNECTION CABLE WL 3M, PART ID: 8564.851, BATCH# 230372. ACCORDING TO THE RECEIVED INFORMATION, THE USER INDICATED THAT THE CONNECTION BETWEEN THE MOTOR CONTROL UNIT, CONNECTION CABLE, AND THE HANDPIECE HAD BEEN INTERRUPTED. DUE TO IT, THE MOTOR CONTROL UNIT FAILED TO RECOGNIZE THE HANDPIECE. THE CONNECTION CABLE WAS CHECKED, AND NO BENT PINS WERE FOUND. THE REPORTED EVENT HAPPENED ON A THIRD CASE, AT THE BEGINNING OF A HOLMIUM LASER ENUCLEATION OF THE PROSTATE (HOLEP) PROCEDURE BUT PRIOR TO BEING USED ON THE PATIENT. HOWEVER, THERE WAS A REPORTED 15-MINUTE DELAY TROUBLESHOOTING THE DEVICE. THE PROCEDURE WAS NOT COMPLETED, AND THE PATIENT WAS BROUGHT BACK 2 DAYS LATER FOR MORCELLATION. ANOTHER CONNECTION CABLE WAS USED, AND IT WORKED FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20369024 ↗
  • InjuryOct 4, 2024Disconnection; Disconnection

    THE USER FACILITY HAS INFORMED RICHARD WOLF MEDICAL CORPORATION (RWMIC) OF AN ISSUE REGARDING A CONNECTION CABLE WL 3M, PART ID: 8564.851, BATCH# 230372. ACCORDING TO THE RECEIVED INFORMATION, THE USER INDICATED THAT THE CONNECTION BETWEEN THE MOTOR CONTROL UNIT, CONNECTION CABLE, AND THE HANDPIECE HAD BEEN INTERRUPTED. DUE TO IT, THE MOTOR CONTROL UNIT FAILED TO RECOGNIZE THE HANDPIECE. THE CONNECTION CABLE WAS CHECKED, AND NO BENT PINS WERE FOUND. THE REPORTED EVENT HAPPENED ON A THIRD CASE, AT THE BEGINNING OF A HOLMIUM LASER ENUCLEATION OF THE PROSTATE (HOLEP) PROCEDURE BUT PRIOR TO BEING USED ON THE PATIENT. HOWEVER, THERE WAS A REPORTED 15-MINUTE DELAY TROUBLESHOOTING THE DEVICE. THE PROCEDURE WAS NOT COMPLETED, AND THE PATIENT WAS BROUGHT BACK 2 DAYS LATER FOR MORCELLATION. ANOTHER CONNECTION CABLE WAS USED, AND IT WORKED FINE.