Dispos. Clip-On Remote Control 53151Recalls & reported problems

Brainlab SE · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04056481005658), received between Mar 5, 2024 and Mar 5, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Application Program Problem: Parameter Calculation Error” (1 report).

3 other recalls cover devices cleared under the same FDA submission (K092467, K102990, K252562); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Application Program Problem: Parameter Calculation Error1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 5, 2024Application Program Problem: Parameter Calculation Error; Application Program Problem: Parameter Calculation Error

    IT WAS REPORTED THAT DURING A THR SURGERY, WHEN REGISTERING THE LUNATE SURFACE OF THE ACETABULUM, THE COMPUTER IMMEDIATELY JUMPED FORWARD AND STARTED COLLECTING THE FOSSA. THE SURGEON WAS NOT CLICKING THE REMOTE AND THE POINT PROBE WAS STILL COLLECTING FOSSA POINTS IN FREE SPACE. IN ADDITION, WHEN TRIED TO RE-REGISTER THE FOSSA, AN ERROR MESSAGE WAS PROMPTED, STATING THAT THE PELVIS NEEDED TO BE RE-REGISTERED. AT THIS POINT ALL THE PREVIOUS REGISTRATION WAS DELETED AND SINCE THE NECK HAD ALREADY BEEN RESECTED, THE SURGEON RESORTED TO MANUAL INSTRUMENTATION AND THE PROCEDURE WAS RESUMED AFTER A NON-SIGNIFICANT DELAY. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18836674 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K092467, K102990, K252562. One submission often covers many models, so check the product and lot details in each record.

  • Z-0056-2021 · Aug 31, 2020 · Terminated
    Ultrasound Navigation Software 1.0 and Ultrasound Integration Software for Cranial/ENT Navigation 3.0.x and 3.1.x - Product Usage: intended to be an intra-operative image guided l
  • Z-0959-2017 · Oct 27, 2016 · Terminated
    Femoral and Tibial Cutting Block Adapter Base. Brainlab knee, BrainLAB Image Guided Surgery System / Instrument, Stereotaxic
  • Z-0598-2016 · Apr 22, 2013 · Terminated
    Brainlab Cranial Navigation System: An Image Guided Surgery System / Stereotactic. Radiology Departments. The BrainLAB Cranial IGS System is intended to be an intra-operative imag

Other Neurological Stereotaxic Instrument devices