Skip to content

SIREMOBIL Compact L 08630506Recalls & reported problems

Siemens Healthcare GmbH · mobile C-arm · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04056869008998), received between Sep 5, 2023 and Mar 31, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Fire” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fire1
  • Radiation Leak1
  • Radiation Output Failure1
  • Radiation Overexposure1
  • Smoking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 31, 2026Radiation Overexposure

    SIEMENS BECAME AWARE OF A MALFUNCTION THAT OCCURRED WHILE OPERATING THE SIREMOMIL COMPACT L SYSTEM. DURING AN INTERVENTIONAL PROCEDURE, THE SYSTEM ACTIVATED ON ITS OWN AND POTENTIALLY RELEASED X-RAY. A CONTINUOUS BEEPING SOUND WAS REPORTED AT THE TIME OF THE EVENT. THE UNIT WAS SUBSEQUENTLY RESTARTED, AND THE PROCEDURE WAS COMPLETED USING THE SAME DEVICE WITHOUT FURTHER ISSUES. WE HAVE NO INDICATIONS OF ANY ADVERSE EFFECTS ON THE HEALTH STATUS OF THE INVOLVED PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24742981 ↗
  • MalfunctionNov 26, 2025Fire; Fire

    SIEMENS BECAME AWARE OF AN EVENT THAT OCCURRED WHILE OPERATING THE SUREMOBIL COMPACT L SYSTEM. DURING A SURGERY, DENSE BLACK SMOKE WAS OBSERVED EMANATING FROM THE MACHINE, ACCOMPANIED BY A STRONG BURNING SMELL. THE DEVICE WAS IMMEDIATELY REMOVED FROM SERVICE. ON NOVEMBER 11, 2025, SIEMENS WAS INFORMED THAT THE SMOKE CAUSED TEMPORARY DISCOMFORT TO THE PEOPLE PRESENT. DETAILED CLARIFICATIONS OF THIS EVENT ARE CURRENTLY ONGOING; THEREFORE, THE EVENT IS REPORTED IN DOUBT. WE HAVE NO INDICATIONS OF ANY SERIOUS ADVERSE EFFECTS ON THE HEALTH STATUS OF THE INVOLVED PERSONS.

    Patient problems: Insufficient Information

    MAUDE report 23658394 ↗
  • MalfunctionJul 7, 2025Smoking; Smoking

    SIEMENS BECAME AWARE OF AN INCIDENT THAT OCCURRED WHILE OPERATING THE SIREMOBIL COMPACT L SYSTEM. DURING AN EMERGENCY PATIENT PROCEDURE, THE USER OBSERVED SMOKE COMING FROM THE MONITOR TROLLEY. THE USER IMMEDIATELY TURNED OFF AND UNPLUGGED THE UNIT. THE PATIENT WAS TRANSFERRED TO A DIFFERENT MEDICAL FACILITY FOR FURTHER TREATMENT. WE HAVE NO INDICATIONS OF ANY ADVERSE EFFECTS ON THE HEALTH STATUS OF THE INVOLVED PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22411342 ↗
  • MalfunctionSep 5, 2023Radiation Leak; Radiation Output Failure; Radiation Leak; Radiation Output Failure

    SIEMENS BECAME AWARE OF AN INCIDENT THAT OCCURRED WHILE OPERATING THE SIREMOBIL COMPACT L (3) SYSTEM. THE FOOTSWITCH WAS STUCK CAUSING THE SYSTEM TO RELEASE RADIATION UNINTENTIONALLY FOR 5 MINUTES. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION TO CONTINUE WITH THE INVESTIGATION. WE HAVE NO INDICATIONS OF ANY ADVERSE EFFECTS ON THE HEALTH STATUS OF THE PERSONS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17682212 ↗

Other Image-intensified fluoroscopic x-ray system, mobile devices

All Image-intensified fluoroscopic x-ray system, mobile devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page