ARCADIS Varic 10143406Recalls & reported problems
Siemens Healthcare GmbH · mobile C-arm · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 04056869009018), received between Feb 22, 2021 and Apr 17, 2023 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Operating System Becomes Nonfunctional” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Operating System Becomes Nonfunctional2
- Radiation Output Failure2
- Computer Software Problem1
- No Device Output1
- Poor Quality Image1
- Radiation Output Problem1
- Unexpected/Unintended Radiation Output1
- Unintended Collision1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 17, 2023No Device Output; Radiation Output Problem; No Device Output; Radiation Output Problem
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE ARCADIS VARIC GEN 2 MOBILE X-RAY SYSTEM. THE USER REPORTED THAT THEY HAD TO REMOVE THE MAINS POWER CABLE IN ORDER TO STOP FLUOROSCOPY. AS A RESULT, PATIENT TREATMENT WAS TERMINATED. WE ARE UNAWARE OF ANY IMPACT TO THE STATE OF HEALTH OF THE PATIENT OR ANY USER INVOLVED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16750136 ↗MalfunctionNov 30, 2022Computer Software Problem; Computer Software Problem
ON (B)(6) 2022, SIEMENS BECAME AWARE OF A MALFUNCTION THAT OCCURRED WHILE OPERATING THE ARCADIS VARIS GEN 2 SYSTEM. ON NOVEMBER 16, 2022 INFORMATION WAS RECEIVED THAT THE ISSUE OCCURRED DURING AN EMERGENCY PROCEDURE WHEN THE USER STARTED FLUORO, HOWEVER, SEVERAL SECONDS LATER THE SOFTWARE FROZE. THE USER RESTARTED THE SYSTEM, BUT THE ISSUE RECURRED. THE PROCEDURE WAS COMPLETED ON AN ALTERNATE UNIT. WE ARE UNAWARE OF ANY IMPACT TO THE STATE OF HEALTH OF THE PATIENT INVOLVED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15885624 ↗MalfunctionAug 8, 2022Radiation Output Failure; Radiation Output Failure
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE ARCADIS VARIS GEN 2 SYSTEM. DURING SURGERY, THE USER REPORTED THAT NO X-RAY WAS POSSIBLE AND AN ERROR MESSAGE WAS SHOWN. WE ARE UNAWARE OF ANY IMPACT TO THE STATE OF HEALTH OF THE PATIENT INVOLVED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15182193 ↗MalfunctionFeb 14, 2022Radiation Output Failure; Radiation Output Failure
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE ARCADIS VARIC GEN 2 SYSTEM. THE USER REPORTED THAT THE INTENSIFIER FAILED DURING AN EXAM WHICH RESULTED IN OVERHEATING DESPITE NORMAL USE. THE PROCEDURE WAS CONTINUED AND COMPLETED ON AN ALTERNATE SYSTEM. WE ARE UNAWARE OF ANY IMPACT TO THE STATE OF HEALTH OF THE PATIENT INVOLVED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13521854 ↗MalfunctionSep 7, 2021Unexpected/Unintended Radiation Output; Unexpected/Unintended Radiation Output
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE ARCADIS VARIC GEN 2 SYSTEM. DURING A PROCEDURE, THE USER REPORTED THAT X-RAY DID NOT STOP AFTER THE FOOT SWITCH WAS RELEASED. WE ARE UNAWARE OF ANY IMPACT TO THE STATE OF HEALTH OF THE PATIENT INVOLVED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12432346 ↗MalfunctionSep 1, 2021Unintended Collision; Unintended Collision
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE ARCADIS VARIC GEN 2 SYSTEM. DURING A SURGICAL PROCEDURE, THE USER REPORTED THAT THE C-ARM HORIZONTAL COULD NOT BE LOCKED AND THERE WAS A COLLISION WITH THE PATIENT. THERE IS NO REPORT OF IMPACT TO THE STATE OF HEALTH OF ANY PATIENT OR USER INVOLVED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12406284 ↗MalfunctionApr 12, 2021Operating System Becomes Nonfunctional; Operating System Becomes Nonfunctional
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE ARCADIS VARIC GEN 2 SYSTEM. DURING A PROCEDURE, THE USER REPORTED THAT THE DEVICE COULD NOT DISPLAY THE IMAGE. WE ARE UNAWARE OF ANY IMPACT TO THE STATE OF HEALTH OF THE PATIENT INVOLVED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11648074 ↗MalfunctionMar 29, 2021Operating System Becomes Nonfunctional; Operating System Becomes Nonfunctional
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE ARCADIS VARIC GEN 2 SYSTEM. DURING OPERATION OF THE SYSTEM, THE USER REPORTED THAT THE SYSTEM HUNG UP. DURING THE ERROR IT APPEARED THAT THE SYSTEM WAS TOTALLY DOWN. WE ARE UNAWARE OF ANY IMPACT TO THE STATE OF HEALTH OF THE PATIENT INVOLVED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11581892 ↗MalfunctionFeb 22, 2021Poor Quality Image; Poor Quality Image
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE ARCADIS VARIC GEN 2 SYSTEM. DURING AN EMERGENCY PROCEDURE, THE USER REPORTED THAT THE SYSTEM DISPLAYED AN ERROR MESSAGE AND THE SYSTEM WAS NO LONGER OPERABLE. THE PROCEDURE WAS CONTINUED AND COMPLETED ON AN ALTERNATE SYSTEM. WE ARE UNAWARE OF ANY IMPACT TO THE STATE OF HEALTH OF THE PATIENT INVOLVED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11357508 ↗
Other Image-intensified fluoroscopic x-ray system, mobile devices
- O-ARM® BI70002000Medtronic Navigation, Inc. · Image-intensified fluoroscopic x-ray system, mobile2 recalls2,132 reports
- O-ARM® BI70000028120Medtronic Navigation, Inc. · Image-intensified fluoroscopic x-ray system, mobile1,343 reports
- O-ARM® BI70002000Medtronic Navigation, Inc. · Image-intensified fluoroscopic x-ray system, mobile2 recalls1 death811 reports
- O-ARM® BI70002000Medtronic Navigation, Inc. · Image-intensified fluoroscopic x-ray system, mobile259 reports
- Zenition Zenition 70Philips Medical Systems Nederland B.V. · Image-Intensified Fluoroscopic X-Ray System, Mobile2 recalls80 reports
- O-armTM O2 Imaging System BI70002000Medtronic Navigation, Inc. · Image-intensified fluoroscopic x-ray system, mobile2 recalls1 death72 reports
- Zenition Zenition 50Philips Medical Systems Nederland B.V. · Image-Intensified Fluoroscopic X-Ray System, Mobile3 recalls21 reports
- O-ARM® BI70002000Medtronic Navigation, Inc. · Image-intensified fluoroscopic x-ray system, mobile2 recalls65 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page