Skip to content

MAGNETOM Sola 11410231Recalls & reported problems

Siemens Healthcare GmbH · MRI Scanner · Class 2

All models of MAGNETOM Sola (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04056869217864), received between Jan 24, 2024 and Aug 26, 2026 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 1, 2023Class IIOpen, Classified

    In some cases, when lubricating grease is present within the headband of the over-ear headphones, it may cause visible, dot- or lineshaped, fat-isointense artifacts during head examinations only. These potential artifacts depend on the position of the headband, the amount of lubricating grease, as well as the sequence parameters used

    Z-2648-2023 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Insufficient Device Problem Information2
  • Wireless Communication Problem1

Patient problems

  • Burn(s)2
  • Acoustic Shock1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 26, 2026Wireless Communication Problem

    Siemens Healthineers (Siemens) was notified of an operator injury associated with the MAGNETOM Sola system. The reported event indicated that the operator went into the examination room to test how loud the voice instructions were through the headphones. The operator stated that the volume was so loud that it caused acoustic shock and the feeling of a blocked ear. After the system was rebooted, the sound volume returned to normal. Based on additional information received by Siemens on August 21, 2026, the operator was treated with steroids. This indicates tinnitus therapy and is considered medical intervention to prevent potential permanent hearing damage or persistent tinnitus.

    Patient problems: Acoustic Shock

    MAUDE report 50454977 ↗
  • InjuryJul 24, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    SIEMENS HEALTHINEERS WAS NOTIFIED OF A POTENTIAL PATIENT INJURY AFTER A FEMORAL MRI EXAMINATION USING THE MAGNETOM SOLA SYSTEM. THE PATIENT REPORTED WHOLE BODY WARMTH DURING THEIR SCAN WHEN A TECHNICIAN WENT IN TO ADMINISTER CONTRAST AGENT. LATER THAT DAY, SKIN REDNESS WAS NOTED IN THE GROIN AREA AND A 2 MM BLACK SPOT ON THE OPPOSITE THIGH WAS FOUND. THE INJURIES WERE NOTICED IN THE WARD WHERE THE PATIENT WAS TREATED FOR VARIOUS PRE-EXISTING CONDITIONS AS AN INPATIENT. THE PATIENT'S PRE-EXISTING CONDITIONS WERE DESCRIBED AS THE FOLLOWING: EXTENSIVE PELVIC FIXATION FROM REMOVAL OF OSTEOSARCOMA, PREVIOUS HEMIPELVECTOMY AND SACRUM REMOVAL AND EXTENSIVE FIXATIONS IN THE LUMBAR SPINE. FURTHERMORE, THE PATIENT HAD A RECENT TRAUMATIC FALL RESULTING IN MULTIPLE PROCEDURES AND DEBRIDEMENT. THE AREA OF OR NEAR THE REDNESS WAS COVERED IN A BANDAGE FROM A PREVIOUS DEBRIDEMENT DURING THE MRI EXAMINATION. THE PATIENT STATED THAT THE BLACK SPOT WAS NOT THERE PRIOR TO THE MRI. THE REDNESS WAS TREATED WITH A COLD COMPRESS AND FLAMAZINE CREAM. THERE IS NO ADDITIONAL INFORMATION AVAILABLE, AND IF ANY TREATMENT WAS NECESSARY FOR THE BLACK SPOT. THE INJURY IS STILL RESOLVING. THE PATIENT DECLINED PICTU

    Patient problems: Burn(s)

    MAUDE report 22586184 ↗
  • MalfunctionJan 24, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    SIEMENS HEALTHINEERS WAS NOTIFIED THAT A PATIENT SUSTAINED AN INJURY OF UNKNOWN SEVERITY DURING A HAMSTRING ATTACHMENT BILATERAL MRI EXAMINATION WITH THE MAGENETOM SOLA SYSTEM. SIEMENS HEALTHINEERS WAS INFORMED THAT THE PATIENT HAD REDNESS ON HIS LOWER BACK AND AN APPROXIMATELY 2CM BURN ON HIS PENIS. THE SEVERITY OF THE BURN IS UNKNOWN TO SIEMENS HEALTHINEERS AS OF THE DATE OF THIS REPORT. ADDITIONAL INFORMATION HAS BEEN REQUESTED FROM THE CUSTOMER FACILITY REGARDING THE EVENT, EXTENT OF THE PATIENT¿S INJURY, AND ANY MEDICAL TREATMENT PROVIDED.

    Patient problems: Burn(s)

    MAUDE report 18570131 ↗

Other System, nuclear magnetic resonance spectroscopic devices

All System, nuclear magnetic resonance spectroscopic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page