Skip to content

LAZIC 70.581.01Recalls & reported problems

Peter Lazic GmbH · CUSHING-LANGOLT SPECULA , 70 x 15MM FOR TRANS-SPHENOIDALE HYPOPHYSECTOMY BLACK COATED · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04250603738718), received between Dec 19, 2023 and Dec 19, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Mechanical Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 19, 2023Mechanical Problem; Mechanical Problem

    THE INCIDENT HAPPENED ON (B)(6) 2023. FEEDBACK THAT THE INCIDENT OCCURRED WAS RECEIVED BY PETER LAZIC GMBH ON (B)(6) 2023. THE INCIDENT HAPPENED IN A HOSPITAL IN GERMANY. THE HOSPITAL STATES THAT THE SPECULUM WAS INSERTED INTO THE RIGHT SIDE OF THE NOSE AND SPREAD OPEN. AFTER COMPLETION OF THE PROCEDURE, THE SPREADER COULD NOT BE TWISTED BACK TOGETHER, SO IT HAD TO BE REMOVED IN A WIDE OPEN STATE. THE HOSPITAL STATES THAT IT COULD HAVE LED TO A SERIOUS DETERIORATION OF THE PATIENT'S HEALTH. THE INVESTIGATION OF THE CAUSE IS ON-GOING. NO PATIENT DATA OR FURTHER DATA ON THE PROCEDURE ARE KNOWN. INTERNAL IDENTIFICATION NUMBER (B)(4) .

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18353295 ↗

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page