Skip to content

Vyntus V-378510Recalls & reported problems

Vyaire Medical, Inc. · Vyntus APS · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04250892904863), received between Aug 30, 2023 and Aug 30, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Computer Software Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Computer Software Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 30, 2023Computer Software Problem; Computer Software Problem

    IT WAS REPORTED TO VYAIRE MEDICAL THAT ON THE METHACHOLINE CHALLENGE TEST REPORTS, THAT IF THE RESULT IS NEGATIVE BUT THE PD20 IS REACHED IN THE "NORMAL" RANGE, THE SENTRYSUITE 3.20.3 SOFTWARE OF THE VYNTUS APS INCORRECTLY INTERPRETS THIS AS "MILD" POSITIVE ON THE SEVERITY INDICATOR BAR FOR AIRWAY HYPERRESPONSIVENESS. DISPLAYED MEASUREMENT RESULTS ON THE REPORT ARE CORRECT. NO HARM TO THE PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17650879 ↗

Other CALCULATOR, PULMONARY FUNCTION DATA devices

All CALCULATOR, PULMONARY FUNCTION DATA devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page