Vyntus V-378510Recalls & reported problems
Vyaire Medical, Inc. · Vyntus APS · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04250892904863), received between Aug 30, 2023 and Aug 30, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Computer Software Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Computer Software Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 30, 2023Computer Software Problem; Computer Software Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT ON THE METHACHOLINE CHALLENGE TEST REPORTS, THAT IF THE RESULT IS NEGATIVE BUT THE PD20 IS REACHED IN THE "NORMAL" RANGE, THE SENTRYSUITE 3.20.3 SOFTWARE OF THE VYNTUS APS INCORRECTLY INTERPRETS THIS AS "MILD" POSITIVE ON THE SEVERITY INDICATOR BAR FOR AIRWAY HYPERRESPONSIVENESS. DISPLAYED MEASUREMENT RESULTS ON THE REPORT ARE CORRECT. NO HARM TO THE PATIENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17650879 ↗
Other CALCULATOR, PULMONARY FUNCTION DATA devices
- Vyaire V-707327Jaeger Medical GmbH · Calculator, Pulmonary Function Data1 recall0 reports
- Vyaire V-178500-13Jaeger Medical GmbH · Calculator, Pulmonary Function Data6 reports
- Vyaire V-178500Jaeger Medical GmbH · Calculator, Pulmonary Function Data4 reports
- Vyaire V-178603Jaeger Medical GmbH · Calculator, Pulmonary Function Data3 reports
- Vyaire V-178601Jaeger Medical GmbH · Calculator, Pulmonary Function Data2 reports
- Jaeger 172520Jaeger Medical GmbH · Calculator, Pulmonary Function Data2 reports
- Jaeger V-175423Jaeger Medical GmbH · Calculator, Pulmonary Function Data2 reports
- Vyaire V-707960Jaeger Medical GmbH · Calculator, Pulmonary Function Data2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page