Vyaire V-378540Recalls & reported problems
Jaeger Medical GmbH · Vyntus BODY · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04250892905693), received between Nov 21, 2024 and Nov 21, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Environmental Particulates1
Patient problems
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 21, 2024Detachment of Device or Device Component; Environmental Particulates; Detachment of Device or Device Component; Environmental Particulates
ON 22-OCT-2024 VYAIRE WAS INFORMED BY A CUSTOMER, THAT THE USS MODULE WAS DROPPED WHILE CONDUCTING THE TEST ON PATIENT AND THE USS CABLE DISLODGE FROM THE SENSOR LEAVE ONLY THE CABLE COVER ON IT. THE CUSTOMER RECONNECTED THE CABLE TO THE USS SENSOR WITHOUT TURNING OFF THE VYNTUS BODY. DUE TO THE MISSING CABLE COVER THIS RESULTED IN SMOKE COMING OUT OF THE FPV MODULE. THERE WAS NO PATIENT HARM REPORTED, FURTHER INFORMATION IS REQUESTED.
Patient problems: Insufficient Information
MAUDE report 20746540 ↗
Other Calculator, Pulmonary Function Data devices
- Vyaire V-707327Jaeger Medical GmbH · Calculator, Pulmonary Function Data1 recall0 reports
- Vyaire V-178500-13Jaeger Medical GmbH · Calculator, Pulmonary Function Data6 reports
- Vyaire V-178500Jaeger Medical GmbH · Calculator, Pulmonary Function Data4 reports
- Vyaire V-178603Jaeger Medical GmbH · Calculator, Pulmonary Function Data3 reports
- Jaeger 172520Jaeger Medical GmbH · Calculator, Pulmonary Function Data2 reports
- Vyaire V-178601Jaeger Medical GmbH · Calculator, Pulmonary Function Data2 reports
- Vyaire V-707960Jaeger Medical GmbH · Calculator, Pulmonary Function Data2 reports
- Jaeger V-175423Jaeger Medical GmbH · Calculator, Pulmonary Function Data2 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.