OSARTIS GmbH · Simplex® HV with Gentamicin is a fast-setting acrylic resin with addition of gentamicin sulfate for use in bone surgery. Mixing the two separate sterile components produces a ductile bone cement which, after hardening, fixes the implant and transfers stresses produced during movement evenly to the bone. Simplex® HV with Gentamicinn cement powder also contains insoluble zirconium dioxide as an X-ray contrast medium. Simplex® HV with Gentamicin does not emit a signal and does not pose a safety risk in a magnetic resonance environment. · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Simplex HV with Gentamicin bone cement is a fast-setting acrylic cement with antibiotic, used in bone surgery to fix implants. Most of the 64 reports describe knee implants that later loosened or failed to grow into the bone, leading to pain, joint swelling, infection, and revision surgery; no deaths and 62 injuries are reported. One recall, started in October 2022 and still open, concerns packs distributed under a different part number than the one ordered.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
64
Deaths reported
0
Injuries reported
62
Recalls naming it
1

Summary

The FDA's MAUDE database contains 64 adverse event reports that list this exact device (DI 04260056885761), received between Aug 19, 2016 and Jun 26, 2026 — classified as 62 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (17 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 21, 2022Class IIOpen, Classified

    Simplex HV With Gentamicin CE, 10 PACKS (P/N 6193-1-010) were distributed to customers who had ordered Simplex HV With Gentamicin in US, 10 PACKS (P/N 6195-1-010)

    Z-0240-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem27
  • Insufficient Device Problem Information17
  • Loosening of Implant Not Related to Bone-Ingrowth15
  • Patient Device Interaction Problem13
  • Loss of Osseointegration11
  • Appropriate Device Problem Term/Code Not Available7
  • Contamination3
  • Missing Value Reason3

Patient problems

  • Arthralgia17
  • Failure of Implant16
  • Unspecified Infection15
  • Inadequate Osseointegration12
  • Injury9
  • Joint Laxity9
  • Implant Pain8
  • Pain3

Recent adverse event reports

The 25 most recent of 64 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 26, 2026Patient Device Interaction Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A RIGHT TOTAL KNEE REPLACEMENT ON (B)(6) 2024. IN LATE (B)(6), THE PATIENT ALLEGEDLY DEVELOPED INCREASING PAIN AND FUNCTIONAL IMPAIRMENT. THE PATIENT WAS REVISED ON (B)(6) 2025 DUE TO THE ALLEGED "FAILURE OF THE POLYETHYLENE LINER, WHICH PERMITTED ABNORMAL MOVEMENT AND MISALIGNMENT OF THE FEMORAL AND TIBIAL COMPONENTS".

    Patient problems: Arthralgia; Loss of Range of Motion

    MAUDE report 25634353 ↗
  • InjuryDec 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO INFECTION. AN I&D WITH WASHOUT AND POLY SWAP OF A 4X9 CS INSERT WAS PERFORMED.

    Patient problems: Implant Pain; Unspecified Infection

    MAUDE report 23741362 ↗
  • InjuryOct 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT REPORTED; HE HAD A PARTIAL KNEE REPLACEMENT WITH STRYKER ORTHOPEDICS KNEE AND HAVING PROBLEMS WITH IT. SWELLING WITH FLUIDS AND PAIN. FLUIDS WERE DRAINED BUT RETURNED. PATIENT HAS RECEIVED SHOTS OF CORTISONE.

    Patient problems: Arthralgia; Arthritis; Implant Pain; Joint effusion

    MAUDE report 23356123 ↗
  • InjuryMay 14, 2025Expiration Date Error; Expiration Date Error

    A SALES GENERAL MANAGER REPORTED THAT, ACCORDING TO A SALES REP, A SURGEON HAD INADVERTENTLY USED SIMPLEX HV WITH GENTAMICIN DURING A TOTAL KNEE REPLACEMENT THAT WAS EXPIRED SINCE TWO YEARS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22017970 ↗
  • InjuryJun 26, 2024Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth; Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth

    PATIENT INFORMATION PROVIDED BY THE US DISTRIBUTION PARTNER: "MY KNEE NEVER FULLY WORKED AFTER THE SURGERY AND THE IMPLANT BECAME LOOSE, MY KNEE WOULD BEND ALL THE WAY BACKWARDS , I'VE FALLEN A FEW TIMES AND SO MUCH PAIN. I ALSO HAVE TO GET KNEE REPLACEMENT TO REPLACE THE IMPLANT AND CURRENTLY WEARING A BRACE. DID THERAPY AT HOME AND WITH THERAPIST. SCHEDULED SURGERY IN 2024 TO REPLACE THE ORIGINAL IMPLANT." NO INFORMATION REGARDING THE REVISION SURGERY HAS BEEN PROVIDED.

    Patient problems: Arthralgia; Implant Pain; Joint Laxity

    MAUDE report 19612001 ↗
  • InjuryJun 14, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED: "PATIENT WAS REVISED DUE TO POSSIBLE INFECTION. NO OTHER INFORMATION AVAILABLE DUE TO HOSPITAL POLICY." RIGHT KNEE. A 3X13 TS INSERT WAS EXCHANGED FOR ANOTHER.

    Patient problems: Unspecified Infection

    MAUDE report 19535036 ↗
  • InjuryJan 25, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH DISTRIBUTORS FACEBOOK PAGE: "MY MAKO KNEE LASTED 8 WEEKS ! I HAD A REVISION A FEW YEARS LATER!" UPDATE PER CONVERSATION WITH PATIENT: "PATIENT HAD A PRIMARY RIGHT MAKO PARTIAL KNEE AND WAS REVISED TO CONVERT TO A FULL KNEE. PATIENT REPORTED THAT SURGEON CHOSE TO CONVERT TO A FULL KNEE AS THE "OUTER PORTION OF THE HIS NATIVE KNEE BECAME BONE ON BONE". UPDATE: PER CLINICIAN REVIEW OF PROVIDED MEDICAL RECORDS: X-RAYS IN 2016 AND 2017 SHOWED ADVANCED OA OF THE LATERAL COMPARTMENT AND WEAR OF THE POLYETHYLENE WITH SOME LIKELY SUBSIDENCE.

    Patient problems: Arthralgia; Implant Pain

    MAUDE report 18578203 ↗
  • InjuryMay 17, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED AFTER PATIENT COMPLAINT OF PAIN FOLLOWING 2 FALLS. PRIOR TO THE FALLS THE KNEE WAS GREAT. PRE-OPERATIVELY, SURGEON REPORTED TIBIAL FRACTURE AND BASEPLATE SUBSIDENCE. THE BASEPLATE, INSERT, AND FEMORAL COMPONENT WERE REVISED. THE SALES REP CONFIRMED FEMUR WAS REVISED TO CONVERT TO A PS KNEE, AND THAT THERE WERE NO ALLEGATIONS AGAINST THE REVISED INSERT. THE PATELLAR COMPONENT WAS RETAINED. THE REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Arthralgia; Implant Pain

    MAUDE report 16944038 ↗
  • InjuryDec 27, 2022Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO SWELLING, CLANKING, AND JOINT LAXITY. A CR CEMENTED FEMUR AND CS INSERT WERE REVISED TO PS IMPLANTS. THE REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Arthralgia; Implant Pain; Joint Laxity

    MAUDE report 16049971 ↗
  • InjuryDec 16, 2022Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem

    A PATIENT REPORTED PRIMARY RIGHT KNEE IMPLANT ON (B)(6) 2021 AND REPORTED ONGOING PAIN AND INSTABILITY.

    Patient problems: Arthralgia; Implant Pain; Joint Laxity

    MAUDE report 15993941 ↗
  • InjuryOct 29, 2022Loss of Osseointegration; Loss of Osseointegration

    AS REPORTED VIA MW5112078: THIS PATIENT HAD A TOTAL KNEE REPLACEMENT PERFORMED ON THE DATE INDICATED. THE KNEE FAILED MUCH EARLIER THAN EXPECTED. WORK UP WAS NEGATIVE FOR INFECTION. THE RECORDS INDICATE THE ORIGINAL SURGERY WAS DONE USING STRYKER HIGH VISCOSITY BONE CEMENT FOR THE KNEE REPLACEMENT SURGERY. THE PATIENT HAD TO HAVE REVISION SURGERY AS INDICATED. THE FINDINGS AT THE TIME OF SURGERY WERE CLASSIC FOR A CEMENT FAILURE IN THAT THE CEMENT WAS NOT BONDED OR ADHERENT TO THE IMPLANTS BUT WAS AFFIXED TO THE BONE. THIS CASE IS ONE OF MANY CASES OF FAILURE AFTER 1.5-2.5 YEARS WHEN USING THIS PRODUCT.

    Patient problems: Inadequate Osseointegration

    MAUDE report 15696889 ↗
  • InjuryOct 21, 2022Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth; Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED IN MEDWATCH MW5112078: THE PATIENT HAD A TOTAL KNEE REPLACEMENT PERFORMED ON (B)(6) 2015. THE KNEE FAILED MUCH EARLIER THAN EXPECTED. WORK UP WAS NEGATIVE FOR INFECTION. THE RECORDS INDICATE THE ORIGINAL SURGERY WAS DONE USING STRYKER HIGH VISCOSITY BONE CEMENT FOR THE KNEE REPLACEMENT SURGERY. THE PATIENT HAD TO HAVE REVISION SURGERY AS INDICATED ON (B)(6) 2020. THE FINDINGS AT THE TIME OF SURGERY WERE THAT THE CEMENT WAS NOT BONDED OR ADHERENT TO THE IMPLANTS BUT WAS AFFIXED TO THE BONE.

    Patient problems: Failure of Implant; Joint Laxity

    MAUDE report 15645922 ↗
  • InjurySep 20, 2022Loss of Osseointegration; Loss of Osseointegration

    AS REPORTED IN MW5111445: "THE PATIENT HAD A TOTAL KNEE REPLACEMENT PERFORMED ON THE DATE INDICATED. THE KNEE FAILED MUCH EARLIER THAN EXPECTED. WORK UP WAS NEGATIVE FOR INFECTION. THE RECORDS INDICATE THE ORIGINAL SURGERY WAS DONE USING STRYKER HIGH VISCOSITY BONE CEMENT FOR THE KNEE REPLACEMENT SURGERY. THE PATIENT HAD TO HAVE REVISION SURGERY AS INDICATED. THE FINDINGS AT THE TIME OF SURGERY WERE CLASSIC FOR A CEMENT FAILURE IN THAT THE CEMENT WAS NOT BONDED OR ADHERENT TO THE IMPLANTS BUT WAS AFFIXED TO THE BONE. THIS CASE IS ONE OF MANY CASES OF FAILURE AFTER 1.5-2.5 YEARS WHEN USING THIS PRODUCT.

    Patient problems: Inadequate Osseointegration

    MAUDE report 15453199 ↗
  • InjurySep 16, 2022Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth; Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED IN MEDWATCH MW5111445: THE PATIENT HAD A TOTAL KNEE REPLACEMENT PERFORMED ON (B)(6) 2016. THE KNEE FAILED MUCH EARLIER THAN EXPECTED. WORK UP WAS NEGATIVE FOR INFECTION. THE RECORDS INDICATE THE ORIGINAL SURGERY WAS DONE USING STRYKER HIGH VISCOSITY BONE CEMENT FOR THE KNEE REPLACEMENT SURGERY. THE PATIENT HAD TO HAVE REVISION SURGERY AS INDICATED ON (B)(6) 2020. THE FINDINGS AT THE TIME OF SURGERY WERE THAT THE CEMENT WAS NOT BONDED OR ADHERENT TO THE IMPLANTS BUT WAS AFFIXED TO THE BONE.

    Patient problems: Failure of Implant; Joint Laxity

    MAUDE report 15435658 ↗
  • InjuryJul 12, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    PATIENT REPORTED UPCOMING LEFT KNEE REVISION (SCHEDULED ON (B)(6) 2022) DUE TO LOOSENING. PATIENT REPORTED A SUCCESSFUL PRIMARY SURGERY UNTIL (B)(6) 2022 (IMPLANTED ON (B)(6) 2017) WHERE SHE NOTICED A "LUMP" BEHIND HER LEFT KNEE. PATIENT ALLEGES SURGEON NOTED LOOSENING AND "LEAKING" OF CEMENT. PATIENT DID NOT REPORT MANUFACTURER OF THE KNEE IMPLANTS ASIDE FROM THE CEMENT. PATIENT WANTS COMPENSATION AND IS IN THE PROCESS OF RETAINING AN ATTORNEY.

    Patient problems: Arthralgia; Joint Laxity

    MAUDE report 14985768 ↗
  • InjuryJul 7, 2022Loss of Osseointegration; Loss of Osseointegration

    PATIENT REPORTED UPCOMING LEFT KNEE REVISION (SCHEDULED (B)(6) 2022) DUE TO LOOSENING. PATIENT REPORTED A SUCCESSFUL PRIMARY SURGERY UNTIL APRIL 2022 WHERE SHE NOTICED A "LUMP" BEHIND HER LEFT KNEE. PATIENT ALLEGES SURGEON NOTED LOOSENING AND "LEAKING" OF CEMENT. PATIENT DID NOT REPORT MANUFACTURER OF THE KNEE IMPLANTS ASIDE FROM THE CEMENT. PATIENT NOTED SURGEON ADVISED OF A POTENTIAL STRYKER RECALL ON THE CEMENT AND PATIENT WOULD LIKE TO KNOW IF PART OF A RECALL. PATIENT WANTS COMPENSATION AND IS IN THE PROCESS OF RETAINING AN ATTORNEY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 14951979 ↗
  • InjuryMay 10, 2022Unstable; Unstable

    PATIENT CALLED AND IS INQUIRING ABOUT RECALL FOR HER RIGHT KNEE SURGERY ON (B)(6) 2016. UPDATE 13-APR-2022: PATIENT IS EXPERIENCING INSTABILITY. SHE HAS SEEN THREE DIFFERENT SURGEONS AND HAS UPCOMING REVISION SURGERY IN MID JUNE. PATIENT WOULD LIKE TO KNOW WHAT IMPLANTS SHE CAN USE IN HER REVISION FOR A BETTER OUTCOME.

    Patient problems: Ambulation or Postural Difficulties

    MAUDE report 14344862 ↗
  • InjuryMay 10, 2022Unstable; Unstable

    PATIENT CALLED AND IS INQUIRING ABOUT RECALL FOR HER RIGHT KNEE SURGERY ON (B)(6) 2016. UPDATE ON (B)(6) 2022: PATIENT IS EXPERIENCING INSTABILITY. SHE HAS SEEN THREE DIFFERENT SURGEONS AND HAS UPCOMING REVISION SURGERY IN MID (B)(6). PATIENT WOULD LIKE TO KNOW WHAT IMPLANTS SHE CAN USE IN HER REVISION FOR A BETTER OUTCOME.

    Patient problems: Ambulation or Postural Difficulties

    MAUDE report 14344861 ↗
  • InjuryNov 12, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED DUE TO INFECTION (ORIGINALLY IMPLANTED ON (B)(6) 2018). ALL IMPLANTS, A 6X9 MM CR INSERT, SIZE 5 CR FEMORAL COMPONENT, SIZE 6 PRIMARY BASEPLATE, AND 39 MM SYMMETRIC PATELLA WERE REMOVED. A WASHOUT WAS DONE AND A 6X13 MM PS INSERT, SIZE 5 PS FEMORAL COMPONENT, SIZE 6 UNIVERSAL BASEPLATE, AND NEW 39 MM PATELLA WERE IMPLANTED. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Unspecified Infection

    MAUDE report 12801705 ↗
  • InjuryOct 4, 2021Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth; Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THROUGH THE FILING OF A LAWSUIT THAT ALLEGEDLY THE PATIENT UNDERWENT A TKA ON (B)(6) 2015 AND DURING THE PROCEDURE SIMPLEX HV BONE CEMENT WAS UTILIZED TO CEMENT ALL COMPONENTS IN PATIENT'S KNEE. IT IS FURTHER ALLEGED THAT ON (B)(6) 2019 THE PATIENT UNDERWENT REVISION SURGERY DUE TO TIBIAL COMPONENT LOOSENING.

    Patient problems: Implant Pain; Joint Laxity

    MAUDE report 12567059 ↗
  • InjuryJan 13, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION OF A RIGHT KNEE ON (B)(6) 2020. THE ORIGINAL IMPLANT DATE OF TS SURGERY WAS ON (B)(6) 2017. REASON FOR REVISION IS INFECTION. NO MORE INFORMATION IS AVAILABLE PER DOCTOR.

    Patient problems: Unspecified Infection

    MAUDE report 11160744 ↗
  • InjuryDec 8, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED. AFTER GETTING A FLU SHOT, PATIENT FELT UNWELL AND AN ACHE DEVELOPED IN HIS KNEE SHORTLY AFTER. DURING REVISION AN INFECTION WAS CONFIRMED AND A POLY SWAP WITH WASHOUT WAS PERFORMED.

    Patient problems: Arthralgia; Unspecified Infection

    MAUDE report 10967319 ↗
  • InjuryNov 10, 2020Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Loosening of Implant Not Related to Bone-Ingrowth; Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THROUGH THE FILING OF A LAWSUIT THAT ALLEGEDLY THE PATIENT UNDERWENT A RIGHT SIDED TKA ON (B)(6) 2017 AND DURING THIS PROCEDURE THE SURGEON UTILIZED SIMPLEX G-HV BONE CEMENT TO CEMENT ALL THE COMPONENTS. IT IS FURTHER ALLEGED THAT ON (B)(6) 2019 SHE UNDERWENT A RIGHT SIDED REVISION AND THE OPERATIVE NOTES SHOW THAT THE TIBIAL COMPONENT WAS LOOSE. ADDITIONAL INFORMATION HAVE NOT BEEN PROVIDED.

    Patient problems: Arthralgia; Failure of Implant

    MAUDE report 10816022 ↗
  • InjuryNov 10, 2020Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Loosening of Implant Not Related to Bone-Ingrowth; Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED BY THE PATIENT'S ATTORNEY THROUGH A LEGAL CLAIM THAT ALLEGEDLY ON (B)(6) 2016 THE PATIENT UNDERWENT A LEFT KNEE MEDIAL COMPARTMENT MAKOPLASTY AND DURING THAT PROCEDURE SIMPLEX HV BONE CEMENT WITH GENTAMICIN WAS UTILIZED TO CEMENT ALL COMPONENTS. IT IS FURTHER ALLEGED THAT ON (B)(6) 2019 THE PATIENT UNDERWENT REVISION SURGERY DUE TO LOOSENING OF HER TIBIAL COMPONENT. NO ADDITIONAL INFORMATION HAVE BEEN PROVIDED.

    Patient problems: Arthralgia; Failure of Implant

    MAUDE report 10815561 ↗
  • InjuryNov 10, 2020Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Loosening of Implant Not Related to Bone-Ingrowth; Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED BY THE PATIENT'S ATTORNEY THROUGH A LEGAL CLAIM THAT ALLEGEDLY ON (B)(6) 2018 THE PATIENT UNDERWENT A LEFT TOTAL KNEE ARTHROPLASTY WITH SIMPLEX HV BONE CEMENT WITH GENTAMICIN. IT IS FURTHER ALLEGED THAT ON (B)(6) 2019 THE PATIENT UNDERWENT REVISION SURGERY DUE TO ALLEGED LOOSENING. NO ADDITIONAL INFORMATION HAVE BEEN PROVIDED.

    Patient problems: Arthralgia; Failure of Implant

    MAUDE report 10815560 ↗

Other Bone Cement, Antibiotic devices