Skip to content

EXCOR® C60G-004mRecalls & reported problems

Berlin Heart GmbH · Arterial cannula for children Ø 12/9 mm; L 32 cm; angle 60° · Class 3

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04260090041338), received between Jan 2, 2025 and Jan 2, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Scratched Material” (1 report).

2 other recalls cover devices cleared under the same FDA submission (P160035); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Scratched Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 2, 2025Scratched Material; Scratched Material

    ON (B)(6) 2024 THE SITE CONTACTED BERLIN HEART INC. CLINICAL AFFAIRS (CA) TO REPORT A SCRATCH ON THE OUTFLOW CANNULA AT THE CANNULA-PUMP JUNCTION WITH BLOOD SURROUNDING THE METAL CONNECTOR OF THE PUMP. CA REQUESTED PICTURES OF THE REPORTED CONCERN. UPON REVIEWING THE PICTURES, CA CONFIRMED THE SCRATCH AND THE BLOOD AROUND THE CONNECTION OF THE PUMP. CA RECOMMENDED THAT THE SITE REMOVE THE AFFECTED SECTION OF CANNULA AND REPLACE THE BLOOD PUMP IF NEEDED. DESPITE THE ISSUE, THE EXCOR BLOOD PUMP MAINTAINED COMPLETE FILLING AND EJECTION, THE PATIENT REMAINED STABLE, AND NO SIGNIFICANT CLINICAL COMPLICATIONS WERE REPORTED. ON (B)(6) 2024, THE AFFECTED SECTION OF THE CANNULA WAS TRIMMED, AND THE PUMP WAS EXCHANGED DUE TO THE REPORTED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21056769 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P160035. One submission often covers many models, so check the product and lot details in each record.

  • Z-2320-2026 · Apr 24, 2026 · Open, Classified
    EXCOR Pediatric Ventricular Assist Device (VAD) System Blood Pump PU Valves: 10 ml, REF: P10P-001; 30 ml, REF: P30P-001x01
  • Z-1618-2023 · May 18, 2018 · Terminated
    EXCOR Pediatric Ventricular Assist Device Blood Pumps, Models: P10P-001 (10 ml blood pump); P15P-001 (10 ml blood pump); P25P-001x01 (25 ml blood pump); P30P-001x01 (30 ml blood

Other Ventricular (Assisst) Bypass devices

All Ventricular (Assisst) Bypass devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.