Skip to content

Navilas® Laser System 577sRecalls & reported problems

OD-OS GmbH · The Navilas® Laser System 577s is a laser photocoagulator with an integrated digital fundus camera. The Navilas® Laser System 577s combines imaging technologies (fundus live imaging, and infra-red imaging) with established laser photocoagulation treatment methods by providing the doctor a system for imaging and treatment planning prior to the photocoagulation. · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04260284390013), received between Apr 1, 2026 and Apr 1, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information1

Patient problems

  • Retinal Injury1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 1, 2026Insufficient Device Problem Information

    ON (B)(6) 2026, THE CUSTOMER REPORTED HEARING A STRANGE "STATIC NOISE" COMING FROM THE LASER OPTICAL HEAD OF A NAVILAS 577S LASER SYSTEM DEVICE (SN (B)(6)) LASTING APPROXIMATELY 5 MINUTES BEFORE DISAPPEARING. SUBSEQUENTLY, THE CUSTOMER PERFORMED A TREATMENT SESSION, AND UPON REVIEW OF THE TREATMENT DOCUMENTATION, THE PHYSICIAN OBSERVED THE LASER FIRING WITHIN THE SET EXCLUSION BOUNDARY RATHER THAN THE INTENDED TREATMENT ZONE. THE SYSTEM'S LOG FILES FROM THE DATE OF THE INCIDENT SHOWED MULTIPLE WARNINGS RELATED TO THE AIMING BEAM POSITION NOT BEING VERIFIED, INCLUDING: · "AIMING BEAM POSITION COULD NOT BE VERIFIED. IMPROVE AIMING BEAM VISIBILITY." "SYSTEM IS UNABLE TO CONFIRM AIMING BEAM POSITION." · "REJECTED SPOT APPLICATION DUE TO MISSING AIMING BEAM POSITION." THE USER ELECTED TO PROCEED WITH THE LASER TREATMENT DESPITE THE WARNING MESSAGES. THE TREATMENT IMAGES LATER SHOWED SPOT OFFSET IN TWO TREATMENT PATTERNS, INCLUDING SOME SPOTS WITHIN THE EXCLUSION AREA. NO BURNS WERE VISIBLE IN THE CENTRAL MACULA, AND THE FOVEA WAS NOT AFFECTED. THE PHYSICIAN REPORTED THAT NO SIGNIFICANT VISION LOSS WAS OBSERVED NOR IS EXPECTED, AND THE PATIENT'S PROGNOSIS IS GOOD. SERVICE INSPECTIONS WE

    Patient problems: Retinal Injury

    MAUDE report 24754074 ↗

Other Powered Laser Surgical Instrument devices

All Powered Laser Surgical Instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page