EIT CIF Inserter CET30131Recalls & reported problems

EIT Emerging Implant Technologies GmbH · Slender Inserter · Class 2

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04260397082829), received between Jul 18, 2019 and Feb 12, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

2 other recalls cover devices cleared under the same FDA submission (K170503); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Device-Device Incompatibility1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Code Available1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 12, 2026Device-Device Incompatibility; Material Twisted/Bent

    DURING SET UP IN THE PREP ROOM, THE T-PIECE (PART C) THREAD IS DAMAGED AND NOT HOLDING THE PART TOGETHER. SLIM INSERTER MAD UP OF 5 PARTS CAME DISASSEMBLED IN TRAY, WHEN PUT TOGETHER IN THE PRE ROOM THE T PIECE OF PART C CAME OFF THE THREAD AND WAS THEN SLIPPING OFF DUE TO DAMAGED THREAD SO ITEM WAS UNUSABLE AND IN NEED OF REPLACEMENT. ITEM WAS RETURNED TO SDU FOR CLEANING AND PACKING SEPARATELY FOR POTENTIAL RETURN. THERE WAS NO PATIENT CONSEQUENCE. THERE WAS NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24322404 ↗
  • MalfunctionSep 5, 2024Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2024, DURING A 3 LEVEL ANTERIOR CERVICAL DISCECTOMY AND FUSION (ACDF) SURGERY, THE SURGEON SUCCESSFULLY INSERTED A CONDUIT ACIF CAGE ON THE SECOND LEVEL USING THE SLENDER INSERTER. WHEN THE INSERTER WAS HANDED BACK TO THE SCRUB, THE DEPTH STOP COMPONENT OF THE INSERTER BROKE OFF, AND IT APPEARS THAT THE THREADED PORTION HOLDING IT ONTO THE INSTRUMENT WAS BROKEN AND THEREFORE COULD NOT BE FIXED. NO FRAGMENTS WERE GENERATED. THERE WAS NO DELAY IN THE SURGICAL PROCEDURE BECAUSE THERE ARE 2 TYPES OF INSERTERS IN THE SET. THEREFORE, THE SURGEON USED THE OTHER AVAILABLE INSERTER FOR THE FINAL CAGE INSERTION. THERE WAS NO PATIENT CONSEQUENCE AND THE SURGERY WAS COMPLETED SUCCESSFULLY. THIS REPORT IS FOR ONE (1) SLENDER INSERTER. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20146649 ↗
  • MalfunctionAug 20, 2019Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING AN UNKNOWN PROCEDURE, THE EIT SLIM/SLENDER INSERTER BROKE. THE STOPPER PIECE ENDED UP SHEARING OFF AS THEY WERE INSERTING ONE (1) OF THE SPACERS. THIS COMPLAINT INVOLVES ONE (1) DEVICE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8909824 ↗
  • MalfunctionJul 18, 2019Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, UPON ARRIVAL AT THE SHIPPING PICKUP LOCATION, GARY OPENED THE SET TO DISCOVER THE DAMAGE TO THE SLENDER INSERTER AND UPON INSPECTION IT WAS NOTICED THAT THERE IS ALSO DISCOLORATION ON THE IMPACTOR. THERE WAS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES TWO (2) DEVICES.

    Patient problems: No Code Available

    MAUDE report 8804395 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K170503. One submission often covers many models, so check the product and lot details in each record.

  • Z-1324-2022 · May 20, 2022 · Open, Classified
    EIT CIF cage, H 7mm, 8degree, S- an intervertebral body fusion devices indicated for use with autograft when used as an adjunct to fusion in patients with cervical disc disease (DD
  • Z-1323-2022 · May 20, 2022 · Open, Classified
    EIT CIF cage, H 6mm, 8degree, L- an intervertebral body fusion devices indicated for use with autograft when used as an adjunct to fusion in patients with cervical disc disease (DD

Other Intervertebral Fusion Device With Bone Graft, Cervical devices

All Intervertebral Fusion Device With Bone Graft, Cervical devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.