EIT TLIF Trial TFT81030Recalls & reported problems
EIT Emerging Implant Technologies GmbH · Trial, H 10mm, L 32mm, W 10mm, 8° · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04260557775202), received between Jul 21, 2022 and Jul 21, 2022 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K170503); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 21, 2022Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN AUSTRALIA AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2022, THE SURGEON WAS TRIALING EIT CONDUIT CAGE WITH THE APPROPRIATE TRIAL INSERTER (TFT20101) WITH A 10MM (TFT81030) TRIAL. UPON TIGHTENING THE TRAIL AS PER THE SURGICAL TECHNIQUE, THE INNER SHAFT OF THE TRIAL SHEARED, LEAVING THE TRIAL AND THE BROKEN INSERT INSIDE. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THIS REPORT IS FOR ONE (1) TRIAL H 10MM 8DEG 32/10. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15075690 ↗
FDA clearance and classification
Cleared through 510(k) K170503 on Jul 11, 2017 (decision: Substantially Equivalent). Submitted by Eit Emerging Implant Technologies GmbH as “EIT Cellular Titanium Cervical Cage, EIT Cellular Titanium PLIF Cages, EIT Cellular Titanium TLIF Cages, and EIT Cellular Titanium ALIF Cages”. The FDA review took 140 days. FDA record ↗
The FDA regulates product code MAX (intervertebral fusion device with bone graft, lumbar) under 21 CFR 888.3080 as a Class II device. 21 CFR 888.3080 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
FDA definition of this device type: Intended to stabilize spinal segment to promote fusion in order to restrict motion and decrease pain using bone graft.
Recalls under the same FDA submission
These recalls involve devices cleared under K170503. One submission often covers many models, so check the product and lot details in each record.
- Z-1324-2022 · May 20, 2022 · Open, ClassifiedEIT CIF cage, H 7mm, 8degree, S- an intervertebral body fusion devices indicated for use with autograft when used as an adjunct to fusion in patients with cervical disc disease (DD
- Z-1323-2022 · May 20, 2022 · Open, ClassifiedEIT CIF cage, H 6mm, 8degree, L- an intervertebral body fusion devices indicated for use with autograft when used as an adjunct to fusion in patients with cervical disc disease (DD
Common questions about the EIT TLIF Trial TFT81030
Has the EIT TLIF Trial TFT81030 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04260557775202), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the EIT TLIF Trial TFT81030?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Jul 21, 2022. That report lists device problem “Break” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the EIT TLIF Trial TFT81030?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the EIT TLIF Trial TFT81030 FDA approved?
It was cleared through 510(k) K170503 on Jul 11, 2017. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the EIT TLIF Trial TFT81030, and is it still sold?
The FDA's device database lists EIT Emerging Implant Technologies GmbH as the company responsible for it (brand EIT TLIF Trial, model TFT81030). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Intervertebral Fusion Device With Bone Graft, Lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page