Skip to content

EIT PLIF cage PUI41006Recalls & reported problems

EIT Emerging Implant Technologies GmbH · EIT PLIF, H 10mm, 4°, 26/9 · Class 2 · Implant

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04260557778517), received between Jul 28, 2022 and Oct 6, 2023 — classified as 2 injuries.

The most frequently reported specific device problem is “Migration” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration2

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Insufficient Information1
  • Pain1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 6, 2023Migration; Migration

    THE DATE OF THE REVISION WAS (B)(6) 2023. THE ORIGINAL SURGERY WAS (B)(6) 2023. THE SCREWS WERE FINE AND NOT REVISED. PATIENT IS DOING BETTER. THE HOSPITALS LEGAL TEAM WILL NOT RELEASE THE IMPLANT YET DUE TO THE WIFE OF THE PATIENT WAS WANTING TO IMPLANT AS WELL.

    Patient problems: Insufficient Information

    MAUDE report 17887075 ↗
  • InjuryJul 28, 2022Migration; Migration

    IT WAS REPORTED THAT ON (B)(6) 2022, PATIENT UNDERWENT A TLIF SPINAL PROCEDURE (DATE UNKNOWN) AND SOME POINT AFTERWARD THE TLIF SPACER EXTRUDED POSTERIORLY INTO THE SPINAL CANAL. PATIENT RETURNED TO THE SURGEON COMPLAINING OF LEFT LEG PAIN. UPON X-RAY, THE MAL-POSITIONED SPACER WAS DISCOVERED AND A REVISION SURGERY WAS SCHEDULED. ON (B)(6) 2022 PATIENT UNDERWENT AN ALIF PROCEDURE TO REMOVE THE TLIF SPACER AND REPLACE WITH A SYNFIX ALIF STAND-ALONE SPACER. THIS REPORT IS FOR ONE (1) POST IBF UI H 10 MM 4DEG 26/9. THIS IS REPORT 1 OF 1 FOR COMPLAINT: (B)(4).

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Pain

    MAUDE report 15121201 ↗

Other Intervertebral Fusion Device With Bone Graft, Lumbar devices

All Intervertebral Fusion Device With Bone Graft, Lumbar devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.