EIT PLIF cage PUI81006Recalls & reported problems
EIT Emerging Implant Technologies GmbH · EIT PLIF, H 10mm, 8°, 26/9 · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04260557779972), received between Jul 9, 2020 and Nov 7, 2021 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Migration1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 7, 2021Break; Break
IT WAS REPORTED THAT ON (B)(6) 2021, THE PATIENT UNDERWENT FOR A SPINAL FUSION SURGERY. DURING THE SURGERY, THE SURGEON WAS DOING A TLIF AT L5-S1 WITH SEVERE DEGENERATIVE DISEASE AND PREVIOUS DECOMPRESSION OF L2-S1. ALSO USED CONDUIT T/PLIF INTERBODY SYSTEM FOR IMPLANT. SURGEON USED PADDLE DISTRACTORS. THE CAGE LOOKED DIFFERENT, MAYBE LONGER. AFTER TAMPING, IT WAS NOTICED THAT SOMETHING LOOKED WRONG WITH IMAGE AND CAGE. ATTEMPTED TO REMOVE THE CAGE BUT ONLY HALF OF THE CAGE CAME OUT BECAUSE IT HAD SHEARED IN HALF. THERE WAS A SURGICAL DELAY OF FORTY-FIVE (45) MINUTES. THERE WERE NO PATIENT CONSEQUENCES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. CONCOMITANT DEVICE REPORTED: T-HANDLE INSERTER (PART# PET30101A, LOT# E20DI0683, QUANTITY 1) UNKNOWN MALLET (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN) UNKNOWN DISTRACTOR (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN). THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR ONE (1) POST IBF UI H 10MM 8DEG 26/9. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12760840 ↗InjuryJul 9, 2020Migration; Migration
IT WAS REPORTED ON AN UNKNOWN DATE THAT ON (B)(6) 2020, THE SURGEON PERFORMED AN EXTENSION OF A BILATERAL PEDICLE SCREW CONSTRUCT FROM L2-4 WHERE HE EXTENDED THE CONSTRUCT DOWN TO L5 AND INSERTED A CONDUIT STRAIGHT TLIF CAGE AT THE L4/5 DISC SPACE. THIS CASE WAS DONE BECAUSE OF A GRADE 1/GRADE 2 SPONDYLOLISTHESIS. ON (B)(6) THE SURGEON INFORMED THAT HE NEEDED TO REMOVE THE CAGE BECAUSE OF NONUNION AND HAD MIGRATED POSTERIOR TOWARDS THE SPINAL CANAL. THE SURGEON PERFORMED THE SURGERY AND WAS ABLE TO REMOVE THE CAGE WITHOUT ANY ISSUES. HE THEN USED ANOTHER COMPANY TO INSERT A LATERAL CAGE. THERE WERE NO PATIENT CONSEQUENCES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. CONCOMITANT DEVICE REPORTED: UNKNOWN SCREWS (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN); UNKNOWN ROD (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN); UNKNOWN SETSCREWS (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN). THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR (1) POST IBF UI H 10MM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Code Available
MAUDE report 10254956 ↗
Other Intervertebral Fusion Device With Bone Graft, Lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page