EIT LLIF Inserter LET30100Recalls & reported problems
EIT Emerging Implant Technologies GmbH · Implant Inserter · Class 2
All models of EIT LLIF Inserter (2)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This EIT inserter is a surgical tool used to place spinal fusion cages in the lower back. All 24 reports are malfunctions, mostly describing the inserter breaking or jamming during surgery — in one case a metal piece was lost in the disc space — with no deaths and no injuries reported. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 24
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 04260557837733), received between Nov 23, 2021 and Aug 6, 2024 — classified as 24 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (16 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Device-Device Incompatibility16
- Break8
- Mechanical Jam3
- Entrapment of Device1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions22
- Insufficient Information4
- Foreign Body In Patient1
Recent adverse event reports
The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 6, 2024Break; Break
IT WAS REPORTED THE CONDUIT LLIF INSERTER BROKE INTRAOPERATIVELY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY OR IMPACT TO THE PATIENT. THE PATIENT WAS REPORTED AS DOING REALLY WELL. THIS REPORT IS FOR A CONDUIT LLIF INSERTER. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19917880 ↗MalfunctionAug 1, 2024Device-Device Incompatibility; Device-Device Incompatibility
ADDITIONAL INFORMATION RECEIVED STATES THAT THERE HAS BEEN NO REMOVAL/ REVISION SURGERY AND NO OTHER MEDICAL INTERVENTION WAS REQUIRED.
Patient problems: Insufficient Information
MAUDE report 19882822 ↗MalfunctionFeb 15, 2024Break; Entrapment of Device; Break; Entrapment of Device
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN THE NETHERLANDS AS FOLLOWS: IT WAS REPORTED THAT ON AN ATP PROCEDURE WITH LLIF CONDUIT DURING THE INSERTION OF THE CAGE THE ANTI-ROTATION PART OF INNER SHAFT OF THE INSERTER WAS BROKEN. DURING THE IMPLANTATION OF THE CAGE THIS METAL PART OF THE INSTRUMENT DISAPPEARED BETWEEN THE ENDPLATE AND THE CAGE INTO THE DISC SPACE L3-4. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THERE WAS NO SURGICAL DELAY. THIS REPORT IS FOR ONE (1) IMPLANT INSERTER THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: Foreign Body In Patient; Insufficient Information
MAUDE report 18711219 ↗MalfunctionDec 4, 2023Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
DEVICE REPORT FROM THE UNITED KINGDOM REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6)-2023 THE PATIENT WAS UNDERGOING A LLIF PROCEDURE. DURING INSERTION OF THE IMPLANT AFTER HAMMERING IT INTO THE RIGHT POSITION, THE KNOB OF THE INSERTER GOT STUCK. THE KNOB WOULDN¿T ROTATE IN EITHER DIRECTION, MAKING IT IMPOSSIBLE TO DETACH IT FROM THE IMPLANT. WHEN THE SURGEON APPLIED FORCE ON THE KNOB, THE INNER SHAFT BROKE. THE SURGEON DECIDED TO KEEP THE IMPLANT IN THE PATIENT, BUT NO FURTHER IMPLANT INSERTERS WERE AVAILABLE, SO THE CAGE WAS NOT OPTIMALLY PLACED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO ADVERSE PATIENT IMPACT. NO FURTHER INFORMATION IS AVAILABLE TO REPORT. THIS REPORT IS FOR AN IMPLANT INSERTER. THIS IS REPORT 1 OF 5 FOR (B)(4).
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 18257514 ↗MalfunctionDec 4, 2023Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
DEVICE REPORT FROM THE UNITED KINGDOM REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023 THE PATIENT WAS UNDERGOING A LLIF PROCEDURE. DURING INSERTION OF THE IMPLANT AFTER HAMMERING IT INTO THE RIGHT POSITION, THE KNOB OF THE INSERTER GOT STUCK. THE KNOB WOULDN¿T ROTATE IN EITHER DIRECTION, MAKING IT IMPOSSIBLE TO DETACH IT FROM THE IMPLANT. WHEN THE SURGEON APPLIED FORCE ON THE KNOB, THE INNER SHAFT BROKE. THE SURGEON DECIDED TO KEEP THE IMPLANT IN THE PATIENT, BUT NO FURTHER IMPLANT INSERTERS WERE AVAILABLE, SO THE CAGE WAS NOT OPTIMALLY PLACED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO ADVERSE PATIENT IMPACT. NO FURTHER INFORMATION IS AVAILABLE TO REPORT. THIS REPORT IS FOR AN IMPLANT INSERTER. THIS IS REPORT 2 OF 5 FOR (B)(4).
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 18257506 ↗MalfunctionOct 18, 2023Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM THE UNITED KINGDOM REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2023, AN INSERTER BECAME JAMMED. THE TEAM WAS ABLE TO DETACH IT. PROCEDURE WAS COMPLETED SUCCESSFULLY WITHIN FOURTEEN MINUTES DELAY. THE PATIENT IS DOING WELL. THIS REPORT IS FOR AN IMPLANT INSERTER. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17961165 ↗MalfunctionJun 15, 2023Device-Device Incompatibility; Device-Device Incompatibility
THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17136240 ↗MalfunctionMay 2, 2023Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM DEPUY SYNTHESE REPORTS AN EVENT IN GERMANY AS FOLLOWS: IT WAS REPORTED THAT A CUSTOMER REPORTED, THE INSTRUMENTS CAN NO LONGER BE MOUNTED AND THUS CAN NO LONGER BE USED FOR SURGICAL PROCEDURES. THIS WAS DETERMINED OUTSIDE OF AN OPERATION, SO NO DAMAGE WAS CAUSED. NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR ONE (1) IMPLANT INSERTER. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16852423 ↗MalfunctionMar 22, 2023Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN SWITZERLAND AS FOLLOWS: IT WAS REPORTED THAT ON MARCH 6, 2023, INTRA OPERATIVELY WE COULD NOT LOOSEN THE IMPLANT INSERTER FROM THE CAGE ITSELF IN SITU. THE CAGE CONNECTED TO THE IMPLANT INSERTER WAS ALREADY IMPLANTED BUT WE COULD NOT DISCONNECT THE INSTRUMENT FROM THE IMPLANT AND HAD TO EXPLANT THE CAGE AGAIN. WE TOOK THE BACK UP INSTRUMENT FROM THE TRAY AND HAD TO OPEN ANOTHER STERILE IMPLANT. PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. NO PATIENT CONSEQUENCES. THIS REPORT IS FOR ONE (1) IMPLANT INSERTER. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16593069 ↗MalfunctionFeb 15, 2023Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN GERMANY AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, THAT INTRAOPERATIVELY IT WAS NO LONGER POSSIBLE TO REMOVE THE CAGE INSERTER FROM THE CONDUIT LLIF CAGE. AFTERWARDS, THE CAGE INCLUDING THE HOLDER WAS REMOVED AND REPLACED BY AN ALTERNATIVE (STROMBOLI-CAGE/FA. ULRICH). NO PATIENT HARM. SURGERY DELAY TIME CA. 20 MIN. THIS REPORT IS FOR ONE (1) IMPLANT INSERTER THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16378431 ↗MalfunctionFeb 7, 2023Mechanical Jam; Mechanical Jam
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN POLAND AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2023, THAT DURING THE PROCEDURE, THE IMPLANTATION TOOL JAMMED. WHEN THE IMPLANT WAS ALREADY IN THE PATIENT THERE WAS A PROBLEM WITH PULLING THE IMPLANT FROM THE FEEDER. THE DELAY IN THE PROCEDURE WAS LESS THAN 30 MINUTES. THE SURGERY WAS SUCCESSFULLY COMPLETED. THERE WERE NO CONSEQUENCES THIS REPORT IS FOR ONE (1) IMPLANT INSERTER THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16313439 ↗MalfunctionDec 23, 2022Break; Device-Device Incompatibility; Break; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN UNITED KINGDOM AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2022 THAT THE CONDUIT LLIF INSERTER BROKE AND WOULDN¿T RELEASE THE IMPLANT. THE IMPLANT HAD TO BE REMOVED AND ANOTHER ONE OPENED AS IT COULDN¿T BE TAKEN OFF THE INSERTER. THERE WAS MINIMAL DELAY AND THE CASE WENT WELL IN THE END WITH THE PATIENT DOING WELL NOW. THIS REPORT IS FOR ONE (1) IMPLANT INSERTER THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16042132 ↗MalfunctionDec 13, 2022Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN THE UNITED KINGDOM AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6)2022 , DURING IMPLANT INSERTION THE IMPLANT INSERTER HOLDER ROD (INTERNAL) SHEERED/SNAPPED INSIDE THE INSTRUMENT AT THE THREADED END RENDERING THE INSTRUMENT USELESS. FORTUNATELY THE SURGEON WAS ABLE TO SAFELY DETACH THE INSTRUMENT FROM THE IMPLANT WHICH WAS INSIDE THE PATIENT AND FINISH THE OPERATION WITHOUT COMPROMISING THE OUTCOME. ALL PIECES OF THE BROKEN INSTRUMENT WERE REMOVED FROM THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE IMPLANT INSERTER. THIS IS REPORT 1 OF 1 FOR PC-(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15969996 ↗MalfunctionNov 15, 2022Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN GERMANY AS FOLLOWS: IT WAS REPORTED THAT DURING A SPINAL PROCEDURE ON (B)(6) 2022, THE CONDUIT LLIF IMPLANT HOLDER COULD NOT BE REMOVED FROM IMPLANT. THE PROCEDURE WAS DELAYED FOR AN UNKNOWN AMOUNT OF TIME IN RELATION TO THIS EVENT. THE IMPLANT AND HOLDER WAS REMOVED. IT IS UNKNOWN IF THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE IMPLANT INSERTER. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15797606 ↗MalfunctionNov 10, 2022Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN GERMANY AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, ALL INSERTER DEVICES WERE DEFECTIVE. THE CUSTOMER WAS NOT ABLE TO ASSEMBLE THE INSTRUMENTS, AND THE HAMMER SHOWED A FRACTURE. NO PATIENT WAS HARMED, AND NO SURGICAL DELAY WAS REPORTED. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE TRIAL INSERTER. THIS IS REPORT 3 OF 4 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15770017 ↗MalfunctionOct 31, 2022Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN GERMANY AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, THE CAGE DID NOT MOVE FROM THE INSERTER AFTER IMPLANT PLACEMENT. SURGERY WAS DELAYED BY 30 MINUTES AS A RESULT OF THIS EVENT. THE SURGEON REMOVED THE CAGE WITH THE INSERTER AND USED A CAGE FROM A DIFFERENT COMPANY. THE PROCEDURE WAS THEN COMPLETED SUCCESSFULLY. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE IMPLANT INSERTER. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15698396 ↗MalfunctionAug 1, 2022Break; Break
DEVICE REPORT FROM GERMANY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2022, THE IMPLANT INSERTER BROKE. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR AN IMPLANT INSERTER. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15140737 ↗MalfunctionJul 26, 2022Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED ON (B)(6)2022, THE IMPLANT TRAY WAS BLOCKED, AND THE TOOL COULD NOT BE DETACHED FROM THE IMPLANT. THE GAUGE FEEDER WAS STUCK AND THE TOOL COULD NOT BE UNFOLDED. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THERE WAS A SURGICAL DELAY OF LESS THAN 30 MINUTES. THIS REPORT IS FOR ONE (1) IMPLANT INSERTER. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15103502 ↗MalfunctionJul 26, 2022Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM GERMANY REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022 DURING A MINI ALIF PROCEDURE (AS PER MAYER/ L4/5 / ACCESS FROM THE RIGHT, LEFT LATERAL POSITION), AFTER INSERTING THE CONDUIT LLIF CAGE, IT WAS NOT POSSIBLE TO DETACH THE INSERTION INSTRUMENT FROM THE CAGE. THE INSERTION INSTRUMENT BLOCKED THE CAGE AND IT COULD NOT BE RELEASED. THE CAGE WAS REMOVED WITH HELP OF THE IMPLANT-POSITIONING INSTRUMENT. ANOTHER FREESTYLE CAGE WAS INSERTED WITH THE HELP OF AN IMPLANT-PUSHER. THE TRIAL IMPLANT HOLDER ALSO JAMMED DURING DISASSEMBLY AND WAS BLOCKED COMPLETELY. THERE WAS NO ADVERSE PATIENT CONSEQUENCE, SURGERY WAS DELAYED FOR APPROXIMATELY 15 MINUTES. THERE IS NO FURTHER INFORMATION. THIS REPORT IS FOR AN IMPLANT INSERTER. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15102765 ↗MalfunctionJul 18, 2022Mechanical Jam; Mechanical Jam
DEVICE REPORT FROM THE UNITED KINGDOM REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, THE THREE SUB-COMPONENTS OF THE IMPLANT INSERTER COULD NOT BE REMOVED FROM ONE ANOTHER. THE INDIVIDUAL PARTS WERE STERILIZED WHILE ASSEMBLED AND CANNOT BE DISASSEMBLED. THERE IS NO PATIENT INVOLVEMENT. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15047751 ↗MalfunctionJun 15, 2022Break; Break
DEVICE REPORT FROM GERMANY REPORTS AN EVENT AS FOLLOWS: DURING AN INSPECTION AT THE LOANER SET DEPARTMENT ON AN UNKNOWN DATE, IT WAS NOTICED THAT THE IMPLANT HOLDER IS BROKEN. NO SURGERY IMPACT, NO PATIENT IMPACT. THIS REPORT IS FOR AN IMPLANT INSERTER. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14695124 ↗MalfunctionMar 31, 2022Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN NETHERLANDS AS FOLLOWS: IT WAS REPORTED THAT AN IMPLANT DID NOT COME OFF THE INSERTER DURING THE WORKSHOP. THIS REPORT IS FOR ONE (1) IMPLANT INSERTER. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13967728 ↗MalfunctionMar 11, 2022Mechanical Jam; Mechanical Jam
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN GERMANY AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, AFTER PLACING THE IMPLANT IN THE CORRECT POSITION IT WAS NOT POSSIBLE TO REMOVE THE IMPLANT FROM THE INSTRUMENT. THE INSTRUMENT JAMMED. SURGERY COMPLETED SUCCESSFULLY WITH ANOTHER AVAILABLE INSTRUMENT. NO ADVERSE PATIENT HARM OR IMPACT. THERE WAS A SURGICAL DELAY OF TEN (10) MINUTES. THIS REPORT INVOLVES ONE (1) IMPLANT INSERTER. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13739500 ↗MalfunctionNov 23, 2021Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING A PROCEDURE THE INSTRUMENT BECAME BLOCKED AND COULD NOT BE USED TO INSERT THE INSTRUMENT. THERE WAS NO SURGICAL DELAY. ANOTHER SYSTEM WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO CONSEQUENCE TO THE PATIENT. THIS REPORT INVOLVES ONE (1) IMPLANT INSERTER. THIS IS REPORT 1 OF 1 FOR PC-(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12861030 ↗
FDA clearance and classification
The FDA regulates product code MAX (intervertebral fusion device with bone graft, lumbar) under 21 CFR 888.3080 as a Class II device. 21 CFR 888.3080 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
FDA definition of this device type: Intended to stabilize spinal segment to promote fusion in order to restrict motion and decrease pain using bone graft.
Common questions about the EIT LLIF Inserter LET30100
Has the EIT LLIF Inserter LET30100 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04260557837733), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the EIT LLIF Inserter LET30100?
The FDA's MAUDE database lists 24 adverse event reports for this device, received between Nov 23, 2021 and Aug 6, 2024. In those reports, the most common specific device problems are “Device-Device Incompatibility” (16), “Break” (8) and “Mechanical Jam” (3); the most common patient problem is “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the EIT LLIF Inserter LET30100?
None of the reports are classified as deaths, 0 as injuries and 24 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the EIT LLIF Inserter LET30100, and is it still sold?
The FDA's device database lists EIT Emerging Implant Technologies GmbH as the company responsible for it (brand EIT LLIF Inserter, model LET30100). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Intervertebral Fusion Device With Bone Graft, Lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page