Skip to content

CH-200M 563-58100-47Recalls & reported problems

Shimadzu Corporation · Class 1

All models of CH-200M (2)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Apr 10, 2023Open, Classified

    Eight fixing bolts at the base of the X-ray ceiling tube column assembly may be loose or missing so service is required to install a cable, bracket, and label behind the bracket, indicating when service is required, to eliminate the risk of the device falling.

    Z-2270-2023 · Root cause: Device Design · Recall details · FDA record ↗

Other Tube Mount, X-Ray, Diagnostic devices

All Tube Mount, X-Ray, Diagnostic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page