NA 0277096275Recalls & reported problems
Stryker Corporation · Reciprocating Blade, Double Sided, Offset (70.0 x 0.64 x 12.6mm) · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Stryker reciprocating saw blade is used for cutting bone and tissue during surgery. The 72 reports mostly describe the blade snapping or breaking mid-use, with a few mentioning blades arriving non-sterile or breaking down into pieces. One report involved an injury; no deaths were reported. Most reports note no effect on the patient, though a few mention imaging being needed. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 72
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 72 adverse event reports that list this exact device (DI 04546540051769), received between Aug 27, 2018 and Jul 22, 2026 — classified as 1 injury, 71 malfunctions.
The most frequently reported specific device problem is “Fracture” (61 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture61
- Break7
- Delivered as Unsterile Product3
- Material Disintegration2
- Packaging Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions57
- No Consequences Or Impact To Patient16
- Device Embedded In Tissue or Plaque2
- No Known Impact Or Consequence To Patient2
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 72 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 22, 2026Fracture
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25908435 ↗MalfunctionApr 22, 2026Fracture
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1 ¿ MARCH 31 2026. THIS REPORT SUMMARIZES 4 EVENTS FOR THE FAILURE: FRACTURE. - 2 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT. - 2 EVENTS HAD IMAGING REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24954382 ↗MalfunctionJan 23, 2026Fracture
MalfunctionJul 29, 2025Fracture; Fracture
THIS REPORT SUMMARIZES 4 EVENTS FOR THE FAILURE: FRACTURE. 4 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: Device Embedded In Tissue or Plaque; No Clinical Signs, Symptoms or Conditions
MAUDE report 22647045 ↗InjuryJun 24, 2025Fracture; Fracture
MalfunctionApr 21, 2025Fracture; Fracture
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. - 2 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21876038 ↗MalfunctionApr 21, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS IN WHICH THE DEVICE WAS REPORTEDLY RECEIVED IN SUCH A MANNER TO INDICATE THAT ITS STERILITY HAD BEEN COMPROMISED. - 3 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21875882 ↗MalfunctionJan 21, 2025Fracture; Fracture
THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 4 EVENTS HAD PATIENT INVOLVEMENT: NO PATIENT IMPACT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 21200287 ↗MalfunctionJul 24, 2024Delivered as Unsterile Product; Delivered as Unsterile Product
THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD DEBRIS IN STERILE PACKAGE. - 5 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19818492 ↗MalfunctionOct 23, 2023Fracture; Fracture
THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 5 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17992798 ↗MalfunctionOct 23, 2023Delivered as Unsterile Product; Delivered as Unsterile Product
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD DEBRIS IN STERILE PACKAGE. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17987932 ↗MalfunctionJul 26, 2023Fracture; Fracture
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17398844 ↗MalfunctionMay 19, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A TOTAL KNEE REVISION, THE BLADE BROKE AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16962741 ↗MalfunctionMay 19, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A TOTAL KNEE REVISION, THE BLADE BROKE AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16962737 ↗MalfunctionMay 19, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A TOTAL KNEE REVISION, THE BLADE BROKE AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16962736 ↗MalfunctionMar 2, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A PROCEDURE, THE BLADE BROKE AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION AND NO DELAYS WERE REPORTED AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16467045 ↗MalfunctionJan 6, 2023Material Disintegration; Material Disintegration
IT WAS REPORTED THAT DURING A PROCEDURE, THE BLADE WAS SHREDDING AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16111053 ↗MalfunctionJan 6, 2023Material Disintegration; Material Disintegration
IT WAS REPORTED THAT DURING A PROCEDURE, THE BLADE WAS SHREDDING AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16111050 ↗MalfunctionDec 15, 2022Fracture; Fracture
IT WAS REPORTED VIA MEDWATCH MW5113268: THAT DURING A REVISION TKA, THE BLADE BROKE AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15985680 ↗MalfunctionDec 14, 2022Fracture; Fracture
IT WAS REPORTED VIA MEDWATCH MW5113263: THAT DURING A PRIMARY TKA, THE BLADE BROKE AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15980404 ↗MalfunctionDec 14, 2022Fracture; Fracture
IT WAS REPORTED VIA MEDWATCH MW5113267: THAT DURING A PRIMARY TKA, THE BLADE BROKE AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15980266 ↗MalfunctionNov 29, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A KNEE REVISION PROCEDURE, THE BLADE BROKE AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15880661 ↗MalfunctionNov 16, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A FEMORAL CUT IN AN ANTERIOR HIP CASE, THE BLADE BROKE AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15810755 ↗MalfunctionNov 10, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A HTO PROCEDURE, THE BLADE BROKE. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15769278 ↗MalfunctionNov 9, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A REVISION OF A TOTAL KNEE PROCEDURE, THE BLADE BROKE AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15762800 ↗
FDA clearance and classification
The FDA regulates product code GFA (blade, saw, general & plastic surgery, surgical) under 21 CFR 878.4820 as a Class I device. 21 CFR 878.4820 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 0277096275
Has the NA 0277096275 been recalled?
No FDA recall record in our data names this exact device identifier (DI 04546540051769), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 0277096275?
The FDA's MAUDE database lists 72 adverse event reports for this device, received between Aug 27, 2018 and Jul 22, 2026. In those reports, the most common specific device problems are “Fracture” (61), “Break” (7) and “Delivered as Unsterile Product” (3); the most common patient problem is “Device Embedded In Tissue or Plaque” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 0277096275?
None of the reports are classified as deaths, 1 as an injury and 71 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 0277096275 FDA approved?
Devices of this type (product code GFA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Stryker Corporation.
Who makes the NA 0277096275, and is it still sold?
The FDA's device database lists Stryker Corporation as the company responsible for it (brand NA, model 0277096275). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL devices
- NA 511.801Synthes GmbH · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL223 reports
- Formula 0375544000Stryker Corporation · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL98 reports
- NA 0277096325Stryker Corporation · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL95 reports
- NA 310.25Synthes GmbH · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL91 reports
- Arthrex® AR-8400TDArthrex, Inc. · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL86 reports
- Arthrex® AR-7300DSArthrex, Inc. · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL71 reports
- NA 0277096277Stryker Corporation · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL69 reports
- NA 0277096326Stryker Corporation · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL60 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page