InterPulse 0210112000Recalls & reported problems
Stryker Corporation · Handpiece Set with Soft Tissue Tip and Suction Tube · Class 2
All models of InterPulse (18)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 20
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 04546540144232), received between Jan 28, 2019 and Apr 21, 2025 — classified as 20 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak6
- Detachment of Device or Device Component5
- Delivered as Unsterile Product3
- Fracture2
- Leak/Splash2
- Accessory Incompatible1
- Device Contaminated During Manufacture or Shipping1
Patient problems
- No Clinical Signs, Symptoms or Conditions12
- No Consequences Or Impact To Patient10
- No Patient Involvement2
Recent adverse event reports
The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 21, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE WAS REPORTEDLY CONTAMINATED DURING MANUFACTURE OR SHIPPING. - 1 EVENT HAD THE PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21878599 ↗MalfunctionOct 24, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A COMPONENT DETACH. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15663785 ↗MalfunctionJul 26, 2022Fracture; Fracture
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. - 3 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15102048 ↗MalfunctionApr 25, 2022Fluid/Blood Leak; Fluid/Blood Leak
THIS REPORT SUMMARIZES 10 MALFUNCTION EVENTS IN WHICH THE DEVICE WAS REPORTEDLY LEAKING. 10 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14193167 ↗MalfunctionJan 19, 2022Fluid/Blood Leak; Fluid/Blood Leak
THIS REPORT SUMMARIZES 6 MALFUNCTION EVENTS IN WHICH THE DEVICE WAS REPORTEDLY LEAKING. 6 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13290678 ↗MalfunctionJan 19, 2022Delivered as Unsterile Product; Delivered as Unsterile Product
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD DEBRIS IN STERILE PACKAGE. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13289668 ↗MalfunctionJul 21, 2021Delivered as Unsterile Product; Delivered as Unsterile Product
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD DEBRIS IN STERILE PACKAGE. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12203298 ↗MalfunctionApr 26, 2021Delivered as Unsterile Product; Delivered as Unsterile Product
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD DEBRIS IN STERILE PACKAGE. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11728424 ↗MalfunctionJan 22, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A COMPONENT DETACH. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11215119 ↗MalfunctionOct 28, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A COMPONENT DETACH. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10748406 ↗MalfunctionApr 24, 2020Fluid/Blood Leak; Fluid/Blood Leak
THIS REPORT SUMMARIZES <NOE> 4 </NOE> MALFUNCTION EVENTS IN WHICH THE DEVICE WAS REPORTEDLY LEAKING. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 3 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient; No Patient Involvement
MAUDE report 9997196 ↗MalfunctionApr 24, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES <NOE> 2 </NOE> MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. 2 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9996937 ↗MalfunctionJan 28, 2020Fluid/Blood Leak; Fluid/Blood Leak
THIS REPORT SUMMARIZES <NOE> 3 </NOE> MALFUNCTION EVENTS IN WHICH THE DEVICE WAS REPORTEDLY LEAKING. 3 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9636413 ↗MalfunctionJan 28, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. 2 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 9636410 ↗MalfunctionOct 28, 2019Fluid/Blood Leak; Fluid/Blood Leak
THIS REPORT SUMMARIZES <NOE> 4 </NOE> MALFUNCTION EVENTS IN WHICH THE DEVICE WAS REPORTEDLY LEAKING. 4 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Consequences Or Impact To Patient; No Patient Involvement
MAUDE report 9244998 ↗MalfunctionOct 2, 2019Fracture; Fracture
THE USER FACILITY REPORTED THAT THE DEVICE HAD A BROKEN BATTERY HOUSING PORT DURING A PROCEDURE. THERE WAS NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9145426 ↗MalfunctionJul 23, 2019Fluid/Blood Leak; Fluid/Blood Leak
THIS REPORT SUMMARIZES <NOE> 6 </NOE> MALFUNCTION EVENTS IN WHICH THE DEVICE WAS REPORTEDLY LEAKING. 6 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8816019 ↗MalfunctionJun 11, 2019Accessory Incompatible; Accessory Incompatible
THE USER FACILITY REPORTED THAT THE TIP COULD NOT BE CONNECTED TO THE HANDPIECE PRIOR TO A PROCEDURE. THERE WAS NO PATIENT IMPACT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8688337 ↗MalfunctionMar 20, 2019Leak/Splash; Leak/Splash
THE USER FACILITY REPORTED THAT SALINE WATER WAS LEAKING FROM THE BATTERY BOX, DURING A PROCEDURE. THERE WAS NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8436939 ↗MalfunctionJan 28, 2019Leak/Splash; Leak/Splash
THE USER FACILITY REPORTED THAT WATER LEAKED FROM THE BATTERY BOX DURING PREPARATION OF A PROCEDURE. THERE WAS NO PATIENT IMPACT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8282264 ↗
Other LAVAGE, JET devices
- Elefant ASP185Coloplast A/S · LAVAGE, JET260 reports
- PULSAVAC® 00-5150-475-01Zimmer Orthopaedic Surgical Products, Inc. · LAVAGE, JET255 reports
- PULSAVAC® 00-5150-482-00Zimmer Orthopaedic Surgical Products, Inc. · LAVAGE, JET197 reports
- PULSAVAC® 00-5150-475-00Zimmer Orthopaedic Surgical Products, Inc. · LAVAGE, JET164 reports
- Versajet 66800039Smith & Nephew Medical Limited · Lavage, Jet160 reports
- PULSAVAC® 00-5150-482-01Zimmer Orthopaedic Surgical Products, Inc. · LAVAGE, JET154 reports
- Surgiphor 910110Carefusion 2200, Inc · LAVAGE, JET140 reports
- Axius Blower/Mister CB-1000Maquet Cardiovascular LLC · Lavage, Jet1 recall43 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page