Skip to content

XIA II 03821640Recalls & reported problems

Stryker Corporation · POLYAXIAL SCREW · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04546540149237), received between Oct 3, 2016 and Apr 14, 2021 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Break3
  • Fracture3
  • Material Integrity Problem2
  • Crack1
  • Improper or Incorrect Procedure or Method1
  • Material Deformation1
  • Material Separation1
  • Mechanical Jam1

Patient problems

  • Injury2
  • No Code Available2
  • No Clinical Signs, Symptoms or Conditions1
  • No Information1
  • No Known Impact Or Consequence To Patient1
  • Pain1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 14, 2021Material Separation; Material Deformation; Material Separation; Material Deformation

    IT WAS REPORTED THAT DURING ROTATION OF THE XIA LL POLYAXIAL SCREW THE TULIP HEAD OF THE SCREW CAME OUT OF THE SHAFT INTRA-OPERATIVELY. NO ADVERSE CONSEQUENCES WERE REPORTED. SURGERY WAS SUCCESSFULLY COMPLETED WITH A 10 MINUTE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11667374 ↗
  • MalfunctionApr 2, 2018Fracture; Break; Crack; Fracture

    IT WAS REPORTED THAT THE PATIENT NOTIFIED THE SURGEON OF SUDDEN ONSET OF AXIAL BACK PAIN. SURGEON TOOK X-RAYS, AND NOTICED SCREWS AT SR1 WERE BROKEN AND THE CAGE LOOKED CRACKED. REVISION SURGERY HAS NOT BEEN PLANNED.

    Patient problems: Pain

    MAUDE report 7390577 ↗
  • InjuryJan 18, 2018Improper or Incorrect Procedure or Method; Mechanical Jam; Positioning Problem; Mechanical Jam

    IT WAS REPORTED THAT; PATIENT WAS PUT IN PRONE POSITION. THIS WAS A REGULAR OPERATION STERILIZATION SPREAD TOWELS, FULL CLEAR SURGICAL FIELD EXPOSURE, CONVENTIONAL CONE, OPEN DEVICE AGENT TAKE NAIL, NEEDLE, THE ESTABLISHMENT OF FLUOROSCOPY LOCATION IS GOOD, CHOOSE SCREW IMPLANT, POOR POSITION AFTER PERSPECTIVE, REMOVE THE SCREW, THE BOOK FOUND THE SCREW ANGLE IS FIXED AND CANNOT BE RESTORED.

    Patient problems: No Information; No Known Impact Or Consequence To Patient

    MAUDE report 7203677 ↗
  • InjuryOct 3, 2016Fracture; Material Integrity Problem; Break; Fracture; Material Integrity Problem

    IT WAS REPORTED THAT; LUMBAR REVISION OF L4-S1. S1 SCREWS WERE BROKEN. PATIENT'S LAST SURGERY WAS DONE IN (B)(6) 2015. WE HAD TO REPLACE THE S1 SCREWS. THE 6.5X40MM WERE BROKEN ON THE SCREW SHAFT. WE HAD TO DRILL THEM OUT AND REPLACE WITH 7.5X 50MM SCREWS.

    Patient problems: Injury; No Code Available

    MAUDE report 5995382 ↗
  • MalfunctionOct 3, 2016Fracture; Material Integrity Problem; Break; Fracture; Material Integrity Problem

    IT WAS REPORTED THAT; LUMBAR REVISION OF L4-S1. S1 SCREWS WERE BROKEN. PATIENT'S LAST SURGERY WAS DONE IN (B)(6) 2015. WE HAD TO REPLACE THE S1 SCREWS. THE 6.5X40MM WERE BROKEN ON THE SCREW SHAFT. WE HAD TO DRILL THEM OUT AND REPLACE WITH 7.5X 50MM SCREWS.

    Patient problems: Injury; No Code Available

    MAUDE report 5995354 ↗

Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices

All APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices →

Sources and updates