Skip to content

XIA II 03821750Recalls & reported problems

Stryker Corporation · POLYAXIAL SCREW · Class 2 · Implant

All models of XIA II (34)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04546540149329), received between Sep 1, 2016 and May 8, 2018 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment Of Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Detachment Of Device Component1
  • Detachment of Device or Device Component1
  • Device Dislodged or Dislocated1
  • Material Separation1
  • Mechanical Problem1

Patient problems

  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 8, 2018Detachment of Device or Device Component; Material Separation; Detachment of Device or Device Component

    IT WAS REPORTED THAT WHEN A SCREW WAS IMPLANTED IN THE RIGHT S1 PEDICLE, AFTER ATTACHING ROD TO THE SADDLE WITH A BLOCKER - DURING MANIPULATION OF DISTRACTION TO COMPLETE DISCECTOMY AND INTERBODY DEVICE PLACEMENT, THE TULIP SEPARATED FROM THE SCREW SHANK. IT WAS DISCOVERED DURING X-RAY OF POST IMPLANTATION OF INTERBODY DEVICE. SCREW WAS THEN REMOVED WITH POLY HEAD ADJUSTER DRIVER, AND REPLACED WITH ANOTHER SCREW. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7493405 ↗
  • MalfunctionSep 1, 2016Detachment Of Device Component; Mechanical Problem; Device Dislodged or Dislocated; Mechanical Problem; Device Dislodged or Dislocated

    IT WAS REPORTED THAT; DURING A REVISION, THE PATIENT HAD HARDWARE IN L4-S1. WE WERE EXTENDING THE HARDWARE UP TO L3. WHEN THE DOCTOR TOOK OUT OLD ROD AND STARTED TURNING THE HEADS ON THE SCREWS THAT WERE ALREADY IN THE PATIENT THREE TULIP HEADS POPPED OFF. WE REPLACED THEM WITH OTHER SCREWS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5920662 ↗

Other ORTHOSIS, SPINAL PEDICLE FIXATION devices

All ORTHOSIS, SPINAL PEDICLE FIXATION devices →

Sources and updates