EXETER 0580-1-352Recalls & reported problems
Howmedica Osteonics Corp. · V40 STEM · Class 2 · Implant
All models of EXETER (36)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 25
- Deaths reported
- 0
- Injuries reported
- 23
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 04546540153241), received between Jul 31, 2017 and Sep 24, 2026 — classified as 23 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Fracture” (13 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Aug 28, 2025Class IIOpen, Classified
a potential product mix where the package labeled as Catalog Number 0580-1-442, lot A00976, may contain Catalog Number 0580-1-352, lot G8754849, and vice versa. Patient labels may also be impacted.
Z-2640-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Fracture13
- Break4
- Loss of Osseointegration3
- Adverse Event Without Identified Device or Use Problem2
- Crack1
- Degraded1
- Detachment of Device or Device Component1
- Failure to Advance1
Patient problems
- Injury7
- Ambulation or Postural Difficulties6
- Insufficient Information6
- Pain5
- Fall2
- Inadequate Osseointegration2
- No Consequences Or Impact To Patient2
- Failure of Implant1
Recent adverse event reports
The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.
InjurySep 24, 2026Loss of Osseointegration
The following information was provided: Revision surgery was performed due to loosening of stem.
Patient problems: Inadequate Osseointegration
MAUDE report 50892361 ↗InjuryAug 25, 2026Loss of Osseointegration
As per MedWatch reports MW5190227 MW5190228 MW5190229: it was reported that the hip revision procedure is being undertaken due to a loose femoral stem related to the cement mantle. The surgeon plans to retain the existing pinnacle acetabular shell, remove the intact ceramic liner, and replace it with a pinnacle dm liner to improve stability.
Patient problems: Inadequate Osseointegration
MAUDE report 50442519 ↗InjuryAug 19, 2026Fracture
It was reported that "patient fell out on hip. Then arrived in hospital with Broken implant on Xray (b)(6) 2026."
Patient problems: Fall
MAUDE report 50330735 ↗InjuryMar 17, 2026Migration
THE FOLLOWING INFORMATION WAS PROVIDED: A TOTAL HIP REPLACEMENT WAS DONE 0N (B)(6) 2023. THE STEM HAS SUBSIDED. REVISED.
Patient problems: Insufficient Information
MAUDE report 24616800 ↗InjurySep 15, 2025Loss of Osseointegration; Failure to Advance; Loss of Osseointegration; Failure to Advance
NO NEW INFORMATION.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 23059245 ↗InjuryMar 24, 2025Mechanical Problem; Adverse Event Without Identified Device or Use Problem; Mechanical Problem; Adverse Event Without Identified Device or Use Problem
SURGEON ADVISED PT EXPERIENCING LEG LENGTH DISCREPANCY, NEEDS REVISION OF EXETER STEM. PRIMARY SURGERY PERFORMED; REVISION PLANNED. 1 MONTH TILL PLANNED SURGERY DATE.
Patient problems: Unequal Limb Length
MAUDE report 21676630 ↗InjuryJul 31, 2024Fracture; Fracture
IT WAS REPORTED THAT THE PATIENTS X-RAY SHOWED A FRACTURED STEM. PATIENT PRESENTED TO EMERGENCY (B)(6)2024. IT WAS FURTHER REPORTED THAT "PATIENT REQUIRED REVISION HIP ARTHROPLASTY WITH FEMORAL OSTEOTOMY, CEMENT REMOVAL. POLY LINER CHANGED AND NEW RESTORATION MODULAR STEM IMPLANTATION WITH 5 DALL MILES CABLES."
Patient problems: Ambulation or Postural Difficulties; Insufficient Information
MAUDE report 19871952 ↗InjuryJun 6, 2023Fracture; Fracture
THE PRIMARY SURGERY DATE FOR BOTH WAS (B)(6) 2011. THE PATIENT WENT TO THE HOSPITAL ON (B)(6) 2023 WITH PAIN ON THE RIGHT THIGH. THE BROKEN STEM WAS FOUND ON X-RAY. THE REVISION SURGERY FOR THE RIGHT WAS ON (B)(6) 2023.
Patient problems: Ambulation or Postural Difficulties; Pain
MAUDE report 17069110 ↗InjuryMay 23, 2023Detachment of Device or Device Component; Insufficient Device Problem Information; Detachment of Device or Device Component; Insufficient Device Problem Information
AS PER REP: "IT WAS A REVISION HIP. WE REMOVED AN EXETER STEM WHICH WAS A 37.5 NO.0" UPDATE: PATIENT WAS REVISED DUE TO DISLOCATION BETWEEN STEM AND HEAD.
Patient problems: Insufficient Information; Joint Dislocation
MAUDE report 16982493 ↗InjuryJan 21, 2023Fracture; Fracture
EXETER STEM FRACTURED. REVISION SURGERY NEEDED TO REPLACE FRACTURED STEM.
Patient problems: Ambulation or Postural Difficulties; Insufficient Information
MAUDE report 16221506 ↗InjuryNov 1, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED, "REVISED AN EXETER/MITCH COMBINATION THIS MORNING". THIS PI IS FOR THE RIGHT HIP.
Patient problems: Insufficient Information
MAUDE report 15707304 ↗InjurySep 22, 2022Fracture; Fracture
EXETER STEM TRUNNION BREAKAGE - PATIENT WALKING AT WORK AND INSTANTLY COLLAPSED. PATIENT PRESENTED TO HOSPITAL. REPORTED BY REGISTRAR/SURGEON TO TERRITORY MANAGER. REVISION SURGERY ON (B)(6) 2022. NO SIGNS OF METALOSIS IN TISSUES UPON REVISION. (INITIAL PROCEDURE ON (B)(6) 2016) RIGHT SIDE.
Patient problems: Ambulation or Postural Difficulties
MAUDE report 15462528 ↗InjuryApr 5, 2022Fracture; Fracture
IT HAS BEEN REPORTED "PATIENT OPERATED WITH CEMENTED BIPOLAR/HEMIARTHROPLASTY DONE IN (B)(6) 2015 HAS COME BACK TODAY WITH BROKEN NECK OF THE CEMENTED FEMORAL STEM. EXETER STEM LIKELY AS PER THE SURGEON. BROKEN NECK OF FEMORAL CEMENTED STEM CAUSING HIP PAIN, LEADING TO IMPLANT FAILURE AND THEREBY NOW THE PATIENT IS BEEN POSTED FOR REVISION THR FOR EXPLANTING THE BROKEN STEM AND REVISING THE FEMUR. INDEX SURGERY DONE IN 2015. THE STEM IS LIKELY TO BE EXETER CEMENTED FEMORAL STEM SYSTEM AS PER THE OPERATING SURGEON. MORE DETAILS LIKE IMPLANT LABELS, DISCHARGE SUMMARY AND PATIENT DETAILS FROM INDEX SURGERY HAVE BEEN REQUESTED FOR."
Patient problems: Pain
MAUDE report 14018060 ↗InjuryDec 10, 2021Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT PRESENTED WITH STEM FX AND PAIN IN THE RIGHT HIP. THE XRAYS SHOWED FX AND THE PATIENT REQUIRED A REVISION SURGERY TO A RESTORATION MODULAR.
Patient problems: Pain
MAUDE report 12974859 ↗InjuryOct 12, 2021Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT PRESENTED WITH STEM FX AND PAIN. XRAY SHOWS FX. PATIENT IS MALE 77YEARS OLD LEFT SIDE WEIGHT 108KG. IT WAS REPORTED THAT A REVISION WAS REQUIRED.
Patient problems: Pain
MAUDE report 12618838 ↗InjuryAug 31, 2020Fracture; Fracture
IT WAS REPORTED THAT THE IMPLANTED EXETER STEM HAD BROKEN. REVISION SURGERY WAS COMPLETED AND THE EXETER WAS REMOVED AND REPLACED WITH A REST MOD. IT WAS FURTHER REPORTED THAT THE BROKEN STEM WAS DISCOVERED FOLLOWING X RAYS AND THAT THE INITIAL SIGNS OF THE BROKEN STEM WERE NOTICED WHEN THE PATIENT FELT A CLICK WHILE WALKING THEN LATER HAD DIFFICULTY AMBULATING AND GETTING OUT OF BED AND REQUIRED USE OF WALKER.
Patient problems: Ambulation or Postural Difficulties; Failure of Implant; Injury
MAUDE report 10473522 ↗InjuryNov 21, 2019Break; Fracture; Break; Fracture
BROKEN EXTER STEM. REVISED STEM TO A CEMENT IN CEMENT STEM.
Patient problems: Ambulation or Postural Difficulties; Injury
MAUDE report 9353392 ↗InjurySep 27, 2019Break; Break
THE PATIENT UNDERWENT A PRIMARY HIP SURGERY ON (B)(6) 2017. AFTER PERFORMING A NORMAL RECOVERY, ON (B)(6) HE WENT TO THE DOCTOR BECAUSE HE HAD PAIN. THE DOCTOR WHEN CHECKING AND AFTER SEEING THE PLAQUE, CONFIRMS THAT THE PROSTHESIS PLACED IS BROKEN. AFTER A TELEPHONE COMMUNICATION WITH DR, HE SAYS THAT THE PATIENT SUFFERED A FALL, IT IS NOT KNOWN IF THIS IS THE ROOT CAUSE. THE SURGEON OF THE SECOND CONSULTATION IS NOT THE SAME SURGEON OF THE PRIMARY SURGERY.
Patient problems: Fall; Pain
MAUDE report 9128931 ↗InjuryMay 15, 2019Break; Break
EXETER STEM 37.5 N0 WAS IMPLANTED IN (B)(6) 2012 INTO PATIENT . IT WAS DISCOVERED AS BROKEN WITH THE X-RAY ATTACHED. REVISION SURGERY WILL BE SCHEDULED LATER ON. UPDATE: PATIENT DID NOT SUFFER ANY TRAUMA. LEFT SIDE.
Patient problems: Injury
MAUDE report 8612155 ↗InjuryApr 16, 2019Fracture; Fracture
REVISION RIGHT TOTAL HIP REPLACEMENT. REMOVAL OF BROKEN EXETER STEM. REVISED TO RESTORATION MODULAR. "SUDDEN INCREASE IN PAIN 3 WEEKS AGO SIMPLY WALKING, NO FALL."
Patient problems: Injury
MAUDE report 8518203 ↗InjuryJan 28, 2019Crack; Fracture; Crack; Fracture
A DR REVISED AN EXETER HIP STEM AS IT APPEARED TO HAVE FRACTURED WHEN VIEWED ON X-RAY. THE PATIENT SAID SHE WAS WALKING DOWNSTAIRS A FEW DAYS PRIOR TO (B)(6) AND SHE MISSED A STEP AND LANDED HEAVILY AND FELT SOMETHING CRACK IN HER LEFT LEG. THE SURGEON REMOVED THE EXETER STEM AND HEAD AND REPLACED IT WITH A RESTORATION MODULAR PROSTHESIS. THE SURGEON COMMENTED THAT THE PATIENT HAD A TRIDENT CUP INSITU WITH A CERAMIC LINER.
Patient problems: Injury
MAUDE report 8281204 ↗InjuryAug 14, 2018Break; Fracture; Break; Fracture
PATIENT HAD A TOTAL BILATERAL HIP ARTHROPLASTY (SAME LOT BOTH SIDES) IN 2013. THE LEFT SIDE FRACTURED IN DECEMBER 2016 AND A REVISION SURGERY WAS REQUIRED. DOCTOR REPORTS THAT PATIENT IS OBESE AND HAS A HISTORY OF "STUMBLING". THE CONTRALATERAL STEM REMAINS ASYMPTOMATIC UNTIL THE LAST FOLLOW-UP. PATIENT SUBMITTED BROKEN PIECE FOR ANALYSIS IN THIRD-PARTY LABORATORY .
Patient problems: Injury
MAUDE report 7780592 ↗InjuryJun 19, 2018Degraded; Material Deformation; Degraded; Material Deformation
THE SURGEON REVISED A FAILED EXETER / CONTEMPORARY CUP. THE SURGEON REPORTED A SHATTERED FEMORAL HEAD AND A VERY WORN FEMORAL TRUNNION. UPDATE- ONE YEAR FOLLOW UP X-RAY SHOWED A CERAMIC HEAD FRACTURE.
Patient problems: Injury
MAUDE report 7615712 ↗MalfunctionNov 2, 2017Fitting Problem; Fitting Problem
THE CUSTOMER REPORTED THAT THE SPIGOT OF THE EXETER HIP WOULD NOT FIT ONTO THE EXETER INTRODUCER. THE SURGEON REMOVED THE PLASTIC PART AND PROCEEDED TO INTRODUCE THE EXETER STEM BY HAND. THERE WAS NO DELAY TO SURGERY BUT IT DID MEAN THAT THE STANDARD INSTRUMENTATION COULD NOT BE USED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6996187 ↗MalfunctionJul 31, 2017Sticking; Packaging Problem; Sticking; Packaging Problem
THE CUSTOMER REPORTED THAT, AS THE SCRUB NURSE PEELED BACK THE TOP FILM LAYER OF THE EXETER STEM PACKAGE, THE STERILE PACKAGING LAYER UNDERNEATH WAS BONDED TO THE TOP LAYER. THE SURGEON CHOSE NOT TO USE THIS IMPLANT IN CASE STERILITY HAD BEEN BREACHED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6755404 ↗
FDA clearance and classification
The FDA regulates product code LPH (prosthesis, hip, semi-constrained, metal/polymer, porous uncemented) under 21 CFR 888.3358 as a Class II device. 21 CFR 888.3358 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the EXETER 0580-1-352
Has the EXETER 0580-1-352 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2640-2025 (Class II), began on Aug 28, 2025: a potential product mix where the package labeled as Catalog Number 0580-1-442, lot A00976, may contain Catalog Number 0580-1-352, lot G8754849, and vice versa. Patient labels may also be impacted. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the EXETER 0580-1-352?
The FDA's MAUDE database lists 25 adverse event reports for this device, received between Jul 31, 2017 and Sep 24, 2026. In those reports, the most common specific device problems are “Fracture” (13), “Break” (4) and “Loss of Osseointegration” (3); the most common patient problems are “Injury” (7), “Ambulation or Postural Difficulties” (6) and “Pain” (5). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the EXETER 0580-1-352?
None of the reports are classified as deaths, 23 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the EXETER 0580-1-352 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the EXETER 0580-1-352, and is it still sold?
The FDA's device database lists Howmedica Osteonics Corp. as the company responsible for it (brand EXETER, model 0580-1-352). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices
- PINNACLE 1217-25-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED734 reports
- NA 1365-52-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED683 reports
- APEX 1246-03-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 deaths610 reports
- PINNACLE 1217-20-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 deaths591 reports
- PINNACLE 1217-30-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death484 reports
- NA 1365-53-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED405 reports
- PINNACLE ALTRX 1221-36-054Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED385 reports
- PINNACLE ALTRX 1221-32-050Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED383 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page