Quick summary

This Exeter hip stem is part of a metal-and-polymer hip replacement system. The 31 reports mostly describe the stem fracturing or separating, hip pain, metal-related problems, and revision surgery; they mention 1 death and 28 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
31
Deaths reported
1
Injuries reported
28
Recalls naming it
0

Summary

The FDA's MAUDE database contains 31 adverse event reports that list this exact device (DI 04546540153302), received between Feb 7, 2014 and May 6, 2026 — classified as 1 death, 28 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (12 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture12
  • Adverse Event Without Identified Device or Use Problem7
  • Insufficient Device Problem Information5
  • Corroded4
  • Degraded3
  • Break2
  • Mechanical Problem2
  • Crack1

Patient problems

  • Ambulation or Postural Difficulties10
  • Injury7
  • Pain7
  • Insufficient Information5
  • Metal Related Pathology5
  • Foreign Body Reaction3
  • Bone Fracture(s)2
  • Fall2

Recent adverse event reports

The 25 most recent of 31 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 6, 2026Fracture

    THE FOLLOWING INFORMATION WAS PROVIDED: "PATIENT WITH AN EXETER STEM IMPLANTED IN LEFT HIP SINCE 10 YEARS SUDDENLY FELT PAIN WHEN WALKING INSIDE THE HOUSE AND WAS NOT ABLE TO LIFT HIS LEG. WENT TO THE ER AND WHERE THEY FIND A BROKEN STEM ON THE X-RAY AND HE WAS TAKEN TO OR THE NEXT DAY."

    Patient problems: Ambulation or Postural Difficulties; Pain

    MAUDE report 25091276 ↗
  • InjuryNov 21, 2025Fracture; Fracture

    THE FOLLOWING WAS REPORTED BY THE REP: "AN EXETER STEM WAS REVISED LAST WEEK AS IT HAD SNAPPED AT THE SHOULDER." PATIENT PRESENTED WITH BROKEN IMPLANT AFTER SLIPPING IN/ON THE WAY TO THE BATHROOM. PATIENT WAS REVISED ON (B)(6) 2025.

    Patient problems: Fall

    MAUDE report 23616274 ↗
  • InjuryJun 23, 2025Adverse Event Without Identified Device or Use Problem; Migration; Adverse Event Without Identified Device or Use Problem; Migration

    REVISION SURGERY FOR EXETER CEMENTED STEM, LINER POLY AND FEMORAL HEAD DUE TO SUBSIDENCE SECONDARY TO GT FRACTURE.

    Patient problems: Bone Fracture(s)

    MAUDE report 22305473 ↗
  • InjuryMar 4, 2025Fracture; Fracture

    NO NEW INFORMATION

    Patient problems: Ambulation or Postural Difficulties; Insufficient Information

    MAUDE report 21514213 ↗
  • InjuryJan 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT'S RIGHT HIP WAS REVISED DUE TO FEMORAL PAIN. INTRA-OPERATIVELY, NOTHING WAS OBSERVED TO BE LOOSE, BROKEN, OR WORN. AN EXETER STEM, HEAD, ADM/ MDM POLY INSERT AND MDM METAL LINER WERE REVISED TO A RESTORATION MODULAR STEM CONSTRUCT, HEAD, ADM/ MDM POLY INSERT AND MDM METAL LINER WITH CABLES. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON. CORRECTION: A TRIDENT X3 ELEVATED RIM INSERT WAS REVISED INSTEAD OF A ADM/ MDM POLY INSERT AND MDM METAL LINER.

    Patient problems: Pain

    MAUDE report 18451127 ↗
  • InjuryDec 7, 2023Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component

    CATASTROPHIC FAILURE OF EXETER STEM USED IN THA. PRIMARY OPERATION 2020 FOR OA. WHEN PATIENT WAS BENDING OVER IN CHAIR, HAD DISCOMFORT IN LEFT HIP. ABLE TO MOBILISE AFTER, WATERING GARDEN, WALKING, SUDDENLY FELT IT 'GIVE WAY'. WOUND HEALED. UPDATE: CLINICIAN REVIEW OF THE PROVIDED X-RAY AND CT SCAN IMAGES INDICATED THAT HEAD NECK DISASSOCIATION AND TRUNNION FRACTURE HAD OCCURRED.

    Patient problems: Ambulation or Postural Difficulties; Discomfort

    MAUDE report 18284672 ↗
  • InjuryApr 24, 2023Degraded; Degraded

    IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED DUE TO FEMORAL PAIN. A STEM (IMPLANTED (B)(6) 2012), HEAD AND LINER (BOTH IMPLANTED (B)(6) 2014) WERE REVISED TO A RESTORATION MODULAR STEM CONSTRUCT, FEMORAL HEAD, AND ADM/ MDM LINER CONSTRUCT. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON. UPDATE (B)(6) 2023 MF: PER OFFICE NOTES RECEIVED DATED (B)(6) 2022: PATIENT IS EXHIBITING "AREAS OF POSTERIOR PROXIMAL FEMORAL OSTEOLYSIS BETWEEN THE MID AND UPPER FEMORAL CEMENT MANTLE" AND AN "IMPINGEMENT PIT OR CYSTIFICATION ALSO NOTED IN THE SUPEROLATERAL ASPECT OF [THE] FEMORAL NECK." ADDITIONALLY, "THE RECENT CAT SCAN SHOWS DRAMATICALLY SIGNIFICANT CORTICAL BONE LOSS OR OSTEOLYSIS DEVELOPING AROUND THE FEMORAL CEMENT MANTLE" AND THE "CORTEX IS BECOMING REMARKABLY THIN IN SOME REGIONS." UPDATE (B)(6) 2023: THE OFFICE NOTES ALSO ALLEGED "UNDERLYING SOFT TISSUE DAMAGE FROM METALLOSIS."

    Patient problems: Metal Related Pathology; Osteolysis; Pain

    MAUDE report 16797225 ↗
  • InjuryJan 25, 2023Corroded; Adverse Event Without Identified Device or Use Problem; Corroded; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED, "THE SURGEON ADVISED THAT HE WAS REVISING AN EXETER/MITCH COMBINATION DUE TO ELEVATED COBALT/CHROMIUM LEVELS IN THE PATIENT."

    Patient problems: Foreign Body Reaction; Metal Related Pathology

    MAUDE report 16241035 ↗
  • InjuryDec 14, 2022Corroded; Material Erosion; Insufficient Device Problem Information; Corroded; Material Erosion; Insufficient Device Problem Information

    IT WAS REPORTED THAT A REVISION TOOK PLACE: REVISION OF AN EXETER/MITCH COMBINATION DUE TO ELEVATED COBALT/CHROMIUM LEVELS IN THE PATIENT.

    Patient problems: Foreign Body Reaction; Metal Related Pathology

    MAUDE report 15975819 ↗
  • InjuryOct 25, 2022Fracture; Fracture

    BROKEN EXETER STEM. CERAMIC HEAD WAS ALSO REVISED.

    Patient problems: Insufficient Information

    MAUDE report 15665070 ↗
  • InjuryOct 13, 2022Device Contaminated During Manufacture or Shipping; Adverse Event Without Identified Device or Use Problem; Device Contaminated During Manufacture or Shipping; Adverse Event Without Identified Device or Use Problem

    PATIENT ADMITTED FOR REVISION SURGERY, INDICATED DUE TO PERIPROSTHETIC INFECTION. STRYKER DEVICE EXPLANTED, AND COMPETITOR ITEMS IMPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 15593205 ↗
  • InjuryJun 6, 2022Fracture; Fracture

    IT HAS BEEN REPORTED BY THE DISTRIBUTOR THAT "THE PATIENT'S STEM BROKE AT TRUNNION. PATIENT DIDN'T FALL. PATIENT HAS REQUESTED AN INDEPENDENT ANALYSIS OF THE STEM. AS SUCH THE STEM WON'T BE AVAILABLE TO BE RETURNED".

    Patient problems: Ambulation or Postural Difficulties

    MAUDE report 14609130 ↗
  • InjuryMar 23, 2022Degraded; Adverse Event Without Identified Device or Use Problem; Degraded; Adverse Event Without Identified Device or Use Problem

    THE FOLLOWING WAS REPORTED: THE PATIENT HAD AN EXETER/ MITCH LEFT THR REVISION.

    Patient problems: Insufficient Information; Metal Related Pathology

    MAUDE report 13860835 ↗
  • InjurySep 24, 2021Corroded; Adverse Event Without Identified Device or Use Problem; Corroded; Adverse Event Without Identified Device or Use Problem

    REVISION OPERATION REQUIRED TO CORRECT STRYKER CONSTRAINED LINER IMPINGEMENT WHICH HAD LEAD TO EXCESSIVE PLASTIC WEAR AND THE LINER'S LOCKING RING BREAKING INSITU. UPDATE 23/SEPTEMBER/2021 WG: AS REPORTED IN THE (B)(6) 2021 OPERATION REPORT: "...EVIDENCE OF PROSTHETIC IMPINGEMENT WITH BURNISHING OF THE POSTERIOR SUPERIOR PORTION OF THE FEMORAL NECK (REFERRED TO IN THE OP REPORT AS AN 'EXISTING EXETER STEM")¿THE CONSTRAINED RING WAS BROKEN AND A PORTION OF THE ACETABULAR RIM WAS FRACTURED..." BASED ON OP REPORTS AND IMPLANT LOT CODE INFORMATION, THE PATIENT HAD A SPACER REVISION IN (B)(6) 2014 (PI (B)(4), CONFIRMED VIA OP REPORT). AT THAT TIME, AN EXETER CONTEMPORARY SHELL, STAINLESS STEEL HEAD, AND EXETER STEM WERE PLACED. SOME TIME BETWEEN (B)(6) 2014 AND (B)(6) 2017, SOME OR ALL OF THOSE IMPLANTS WERE REMOVED AND A CONSTRAINED LINER AND COCR FEMORAL HEAD WERE IMPLANTED (NO RECORD PROVIDED FOR THIS SURGERY, BUT THE REVISED HEAD AND LINER WERE NOT MENTIONED IN THE (B)(6) 2014 SURGERY, AND THE CONSTRAINED LINER WAS MANUFACTURED IN JULY 2014). IN AUGUST OF 2021, THOSE IMPLANTS WERE REVISED. THIS PI IS FOR THE REVISION OF THOSE IMPLANTS, IMPLANTED AT AN UNKNOWN DATE AT THE TIME OF REP

    Patient problems: Ambulation or Postural Difficulties; Insufficient Information

    MAUDE report 12530437 ↗
  • InjuryJul 30, 2021Corroded; Insufficient Device Problem Information; Corroded; Insufficient Device Problem Information

    THE CUSTOMER REPORTED THAT THE PATIENT WHO HAD PREVIOUSLY BEEN IMPLANTED WITH AN EXETER MITCH COMBINATION HAS BEEN REVISED. **UPDATE** REVISION LEFT HIP PATIENT HAD VERY HIGH METAL ION LEVELS.

    Patient problems: Metal Related Pathology

    MAUDE report 12252692 ↗
  • InjuryJul 27, 2021Fracture; Fracture

    THE CUSTOMER REPORTED THAT THE EXETER STEM WAS REVISED DUE TO A FRACTURED STEM.

    Patient problems: Ambulation or Postural Difficulties; Insufficient Information

    MAUDE report 12230030 ↗
  • InjuryJul 27, 2021Fracture; Fracture

    STRYKER REP REPORTED A BROKEN EXETER STEM THAT HAS NEEDED TO BE REVISED.

    Patient problems: Ambulation or Postural Difficulties

    MAUDE report 12229991 ↗
  • InjuryJun 29, 2021Fracture; Fracture

    IT HAS BEEN REPORTED "IMPLANT REVISED DUE TO IMPLANT BREAKAGE. NO TRAUMA, SUDDEN PAIN."

    Patient problems: Pain

    MAUDE report 12084212 ↗
  • InjuryDec 7, 2020Break; Break

    IT WAS REPORTED THAT WHILE THE PATIENT WAS GOING FROM A STANDING TO A SITTING POSITION THEY FELT SOMETHING SNAP. IT WAS FURTHER REPORTED THAT AN XRAY REVEALED A BROKEN EXETER STEM AT THE STEM'S NECK/HEAD JUNCTION AND A REVISION SURGERY WAS REQUIRED TO REPLACE THE STEM/HEAD/INSERT.

    Patient problems: Hip Fracture

    MAUDE report 10957875 ↗
  • InjuryNov 11, 2020Degraded; Degraded

    THA REVISION WAS PERFORMED DUE TO THE CUP LOOSENING. DURING THE REVISION A NECROTIC TISSUE-LIKE GRANULATION TISSUE FILLING AROUND THE TRUNNION OF THE STEM WAS FOUND, AND WHEN THE HEAD WAS REMOVED, THE ENTIRE TRUNNION LOST ITS LUSTER AND RUST-LIKE SUBSTANCE WAS ADHERED. PLEASE INVESTIGATE THE STATE THE TRUNNION.

    Patient problems: Granuloma; Inadequate Osseointegration

    MAUDE report 10823920 ↗
  • InjuryAug 3, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    DURING A THP REVISION FROM A PRIMARY ACCOLANDE TMZF/TRIDENT TITANIUM IN (B)(6) 2019, THE SURGEON DISCOVERED THAT THE POLYETHYLENE WAS DAMAGED/SCRATCHED WITHOUT NOTICING CORPORA LIBERA. NO DAMAGE NOTICEABLE AT THE OLD V40 HEAD. THE DAMAGED X3 POLYETHYLENE ON THIS SHORT TERM IS CONCERNING THE COSTUMER ABOUT THE QUALITY. REASON FOR REVISION - PERIPROSTHETIC FRACTURE.

    Patient problems: Bone Fracture(s); Injury

    MAUDE report 10355363 ↗
  • InjuryJun 10, 2020Fracture; Fracture

    HIP REVISION AFTER BROKEN STEM (EXETER) DIAGNOSIS.

    Patient problems: Ambulation or Postural Difficulties; Injury; Pain

    MAUDE report 10136870 ↗
  • InjuryJan 24, 2020Fracture; Fracture

    IT WAS REPORTED THAT THE NECK OF THE EXETER STEM WAS BROKEN IN SITU, THEREFORE THE IMPLANT WAS EXTRACTED AND REPLACED WITH A NEW EXETER STEM.

    Patient problems: Ambulation or Postural Difficulties; Injury

    MAUDE report 9627856 ↗
  • InjuryJan 13, 2020Break; Break

    THE SURGEON REPORTED THAT HE THE PATIENT WHO HAD BEEN IMPLANTED WITH AN EXETER STEM IS BEING REVISED. THE STEM BROKE FOLLOWING A LOW FALL FROM A BED.

    Patient problems: Fall; Injury

    MAUDE report 9583264 ↗
  • InjuryDec 26, 2018Crack; Fracture; Crack; Fracture

    STRYKER REP REPORTED THAT PATIENT UNDERWENT REVISION SURGERY ON AN EXETER STEM. REVISION DATE (B)(6) 2018.

    Patient problems: Ambulation or Postural Difficulties; Injury

    MAUDE report 8198027 ↗

Other Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices

All Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices →

Sources and updates