NA 6000-011-000Recalls & reported problems

Stryker Leibinger GmbH & Co. KG · LARGE POINTER · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Stryker LARGE POINTER is part of a neurological stereotaxic instrument set used to position needles and instruments during brain surgery. The 49 reports mostly describe the pointer tip bending or warping during a case, sometimes after calibration, which leads to inaccurate positioning; some also mention parts coming apart or breaking. No deaths and no injuries were reported, and most cases finished without harm to the patient. There are no recalls name this device, and no reports were filed in the last 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
49
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 49 adverse event reports that list this exact device (DI 04546540245106), received between Feb 3, 2017 and Oct 3, 2024 — classified as 49 malfunctions.

The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (21 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deformation Due to Compressive Stress21
  • Material Twisted/Bent19
  • Incorrect Measurement9
  • Disassembly8
  • Break6
  • Material Deformation4
  • Fracture3
  • Imprecision3

Patient problems

  • No Clinical Signs, Symptoms or Conditions25
  • No Patient Involvement16
  • No Consequences Or Impact To Patient13

Recent adverse event reports

The 25 most recent of 49 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 3, 2024Imprecision; Imprecision

    IT WAS REPORTED THAT AFTER CALIBRATION AND VALIDATION, THE LARGE POINTER WAS INACCURATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20363275 ↗
  • MalfunctionAug 7, 2024Material Deformation; Material Twisted/Bent; Material Deformation; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A CASE, IT WAS NOTICED THAT THE TIP OF THE LARGE POINTER WAS BENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19935603 ↗
  • MalfunctionJul 1, 2024Material Deformation; Material Deformation

    IT WAS REPORTED THAT DURING A CASE, IT WAS NOTICED THAT THE TIP OF THE LARGE POINTER WAS BENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19651425 ↗
  • MalfunctionMay 21, 2024Imprecision; Incorrect Measurement; Imprecision; Incorrect Measurement

    IT WAS REPORTED THAT DURING A CASE, THE LARGE POINTER WAS INACCURATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19367934 ↗
  • MalfunctionDec 28, 2023Deformation Due to Compressive Stress; Material Deformation; Deformation Due to Compressive Stress; Material Deformation

    PER THE CUSTOMER THE DEVICE WAS BENT AFTER VALIDATION WHICH CAN LEAD TO INACCURACIES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18409037 ↗
  • MalfunctionDec 12, 2023Incorrect Measurement; Deformation Due to Compressive Stress; Material Twisted/Bent; Incorrect Measurement; Deformation Due to Compressive Stress; Material Twisted/Bent

    PER THE CUSTOMER, THE DEVICE'S TIP WAS BENT DURING A PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18310364 ↗
  • MalfunctionNov 27, 2023Incorrect Measurement; Material Twisted/Bent; Incorrect Measurement; Material Twisted/Bent

    PER THE CUSTOMER, THE DEVICE WAS BENT AND INACCURATE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18208589 ↗
  • MalfunctionOct 23, 2023Deformation Due to Compressive Stress; Material Twisted/Bent; Deformation Due to Compressive Stress; Material Twisted/Bent

    PER THE CUSTOMER THE TIP WAS BENT DURING A SURGERY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17983930 ↗
  • MalfunctionOct 23, 2023Deformation Due to Compressive Stress; Material Twisted/Bent; Deformation Due to Compressive Stress; Material Twisted/Bent

    PER THE CUSTOMER THE TIP WAS BENT DURING A SURGERY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17983928 ↗
  • MalfunctionSep 25, 2023Deformation Due to Compressive Stress; Material Twisted/Bent; Deformation Due to Compressive Stress; Material Twisted/Bent

    PER THE CUSTOMER THE TIP WAS BENT DURING A SURGERY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17806573 ↗
  • MalfunctionSep 19, 2023Deformation Due to Compressive Stress; Material Twisted/Bent; Deformation Due to Compressive Stress; Material Twisted/Bent

    PER THE CUSTOMER THE TIP WAS BENT DURING A SURGERY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17773207 ↗
  • MalfunctionAug 28, 2023Incorrect Measurement; Deformation Due to Compressive Stress; Material Deformation; Incorrect Measurement; Deformation Due to Compressive Stress; Material Deformation

    PER THE CUSTOMER, THE TIP OF THE LONG POINTER WAS BENT, AND THE DEVICE COULD NOT RECOGNIZE IT ACCURATELY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17635615 ↗
  • MalfunctionJul 17, 2023Deformation Due to Compressive Stress; Material Twisted/Bent; Deformation Due to Compressive Stress; Material Twisted/Bent

    PER THE CUSTOMER, THE DEVICE WAS BENT DURING THE CASE WHICH CAN LEAD TO INACCURACIES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17331761 ↗
  • MalfunctionJun 28, 2023Deformation Due to Compressive Stress; Material Twisted/Bent; Deformation Due to Compressive Stress; Material Twisted/Bent

    PER THE CLINICAL SPECIALIST, THE DEVICE WAS BENT DURING A PROCEDURE WHICH CAN LEAD TO INACCURACIES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17222481 ↗
  • MalfunctionMay 30, 2023Deformation Due to Compressive Stress; Material Twisted/Bent; Deformation Due to Compressive Stress; Material Twisted/Bent

    PER THE CUSTOMER, THE TIP WAS BENT DURING A CASE, WHICH COULD POSSIBLY LEAD TO AN INCORRECT MEASUREMENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE SAME DEVICE WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17020544 ↗
  • MalfunctionApr 24, 2023Material Twisted/Bent; Material Twisted/Bent

    PER THE CUSTOMER, THE DEVICE BENT DURING THE CASE WHICH CAN LEAD TO IN ACCURACIES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16794439 ↗
  • MalfunctionJan 19, 2023Deformation Due to Compressive Stress; Material Twisted/Bent; Deformation Due to Compressive Stress; Material Twisted/Bent

    PER THE CUSTOMER, THE TIP WAS BENT DURING A PROCEDURE WHICH COULD CAUSE INACCURACIES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16202781 ↗
  • MalfunctionDec 12, 2022Incorrect Measurement; Material Twisted/Bent; Incorrect Measurement; Material Twisted/Bent

    PER THE CUSTOMER, THE DEVICE WAS FOUND TO BE BENT DURING THE PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15963461 ↗
  • MalfunctionDec 6, 2022Incorrect Measurement; Material Twisted/Bent; Incorrect Measurement; Material Twisted/Bent

    PER THE CUSTOMER, THE DEVICE WAS BENT DURING A PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15928339 ↗
  • MalfunctionNov 3, 2022Incorrect Measurement; Material Twisted/Bent; Incorrect Measurement; Material Twisted/Bent

    PER THE CUSTOMER, THE DEVICE WAS BENT DURING A PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15721985 ↗
  • MalfunctionOct 11, 2022Incorrect Measurement; Material Twisted/Bent; Incorrect Measurement; Material Twisted/Bent

    PER THE CUSTOMER, THE DEVICE WAS FOUND BENT AFTER A PROCEDURE WHICH COULD LEAD TO AN INACCURACY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15574627 ↗
  • MalfunctionJul 12, 2022Incorrect Measurement; Material Twisted/Bent; Incorrect Measurement; Material Twisted/Bent

    PER THE CUSTOMER, THE DEVICE'S TIP WAS BENT DURING A PROCEDURE WHICH COULD LEAD TO AN INACCURACY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14987950 ↗
  • MalfunctionMay 16, 2022Deformation Due to Compressive Stress; Material Twisted/Bent; Deformation Due to Compressive Stress; Material Twisted/Bent

    PER THE CUSTOMER, THE POINTER WAS BENT DURING THE CASE WHICH COULD LEAD TO AN INACCURACY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14409464 ↗
  • MalfunctionDec 21, 2021Imprecision; Deformation Due to Compressive Stress; Material Twisted/Bent; Imprecision; Deformation Due to Compressive Stress; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY THE HAND PIECE WAS BENT WHICH CAN LEAD TO AN INACCURACY.   THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13056258 ↗
  • MalfunctionMay 25, 2021Fracture; Fracture

    DURING THE SERVICE EVALUATION PROCESS CONDUCTED AT THE MANUFACTURER FACILITY IT WAS OBSERVED THAT THE TIP OF THE DEVICE WAS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11878181 ↗

Other Neurological stereotaxic Instrument devices

All Neurological stereotaxic Instrument devices →