NA 6000-012-000Recalls & reported problems
Stryker Leibinger GmbH & Co. KG · SMALL POINTER · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Stryker Small Pointer is a surgical instrument used to help guide brain surgery. Most of the 33 reports describe the pointer tip or shaft breaking, snapping off, or coming apart, sometimes during calibration, a procedure, shipping, or service checks, plus a few cases of the pointer not measuring accurately. None of the reports describe a death or an injury, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 32
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 04546540245113), received between Feb 16, 2017 and Aug 8, 2024 — classified as 32 malfunctions.
The most frequently reported specific device problem is “Fracture” (19 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture19
- Break9
- Disassembly8
- Imprecision1
- Incorrect Measurement1
- Mechanical Jam1
Patient problems
- No Consequences Or Impact To Patient18
- No Patient Involvement9
- No Clinical Signs, Symptoms or Conditions4
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 8, 2024Imprecision; Imprecision
IT WAS REPORTED THAT THERE WERE ACCURACY ISSUES DURING A CASE WHEN USING THE SMALL POINTER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19938373 ↗MalfunctionOct 2, 2023Fracture; Mechanical Jam; Fracture; Mechanical Jam
PER THE CUSTOMER, THE SHORT POINTER WAS BROKEN DURING A PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17854070 ↗MalfunctionMay 11, 2023Fracture; Fracture
PER THE CUSTOMER, THE POINTER SNAPPED OFF DURING THE CASE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16912133 ↗MalfunctionOct 18, 2022Incorrect Measurement; Incorrect Measurement
PER THE CUSTOMER THE DEVICE WAS INACCURATE DURING REGISTRATION DURING A PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE SAME DEVICE WITHOUT CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15628726 ↗MalfunctionOct 10, 2019Break; Fracture; Break; Fracture
IT WAS REPORTED THAT THE POINTER SHAFT WAS BROKEN OFF DURING SHIPPING; THIS POSES THE RISK OF A SMALL COMPONENT BEING LOST IN THE SURGICAL SITE. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED UPON RECEIPT AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Patient problems: No Patient Involvement
MAUDE report 9175660 ↗MalfunctionSep 25, 2018Break; Break
IT WAS REPORTED THAT DURING RECEIPT OF THE PRODUCT AT THE USER FACILITY IT WAS OBSERVED THAT THE BATTERY HOLDER IS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED WHEN THE PRODUCT WAS RECEIVED FROM TRANSIT, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7905782 ↗MalfunctionSep 18, 2018Break; Fracture; Break; Fracture
IT WAS REPORTED THAT DURING THE SERVICE EVALUATION PROCESS CONDUCTED AT THE MANUFACTURER FACILITY IT WAS OBSERVED THAT THE TIP OF THE POINTER WAS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7885745 ↗MalfunctionSep 18, 2018Break; Fracture; Break; Fracture
IT WAS REPORTED THAT DURING THE SERVICE EVALUATION PROCESS CONDUCTED AT THE MANUFACTURER FACILITY IT WAS OBSERVED THAT THE TIP OF THE POINTER WAS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7885742 ↗MalfunctionAug 27, 2018Break; Fracture; Break; Fracture
IT WAS REPORTED THAT DURING THE SERVICE EVALUATION PROCESS CONDUCTED AT THE MANUFACTURER FACILITY IT WAS OBSERVED THAT THE TIP OF THE DEVICE IS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7818886 ↗MalfunctionAug 24, 2018Fracture; Fracture
IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE USER FACILITY IT WAS OBSERVED THAT THE TIP OF DEVICE WAS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7812658 ↗MalfunctionAug 22, 2018Break; Fracture; Break; Fracture
IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE USER FACILITY IT WAS OBSERVED THAT THE TIP OF DEVICE IS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7805944 ↗MalfunctionAug 20, 2018Fracture; Fracture
IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE MANUFACTURER FACILITY IT WAS OBSERVED THAT THE TIP OF DEVICE IS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7797005 ↗MalfunctionAug 20, 2018Fracture; Fracture
IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE MANUFACTURER FACILITY IT WAS OBSERVED THAT THE TIP OF DEVICE IS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7797000 ↗MalfunctionAug 20, 2018Fracture; Fracture
IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE MANUFACTURER FACILITY IT WAS OBSERVED THAT THE TIP OF DEVICE IS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7796998 ↗MalfunctionJul 26, 2018Fracture; Fracture
IT WAS REPORTED THAT DURING SETUP PRIOR TO A SURGICAL PROCEDURE AT THE USER FACILITY THAT THE SHORT POINTER FELL OFF, AND THE TIP BROKE OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING SET UP AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7721901 ↗MalfunctionJul 10, 2018Fracture; Fracture
IT WAS REPORTED THAT DURING THE SERVICE EVALUATION PROCESS CONDUCTED AT THE MANUFACTURER FACILITY IT WAS OBSERVED THAT THE TIP OF THE DEVICE IS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7674211 ↗MalfunctionJul 10, 2018Fracture; Fracture
IT WAS REPORTED THAT DURING THE SERVICE EVALUATION PROCESS CONDUCTED AT THE MANUFACTURER FACILITY IT WAS OBSERVED THAT THE TIP OF THE DEVICE IS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7674203 ↗MalfunctionJul 10, 2018Fracture; Fracture
IT WAS REPORTED THAT DURING THE SERVICE EVALUATION PROCESS CONDUCTED AT THE MANUFACTURER FACILITY IT WAS OBSERVED THAT THE TIP OF THE DEVICE IS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7674201 ↗MalfunctionJul 9, 2018Fracture; Fracture
IT WAS REPORTED THAT DURING THE SERVICE EVALUATION PROCESS CONDUCTED AT THE MANUFACTURER FACILITY IT WAS OBSERVED THAT THE TIP OF THE DEVICE IS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7671074 ↗MalfunctionJun 15, 2018Break; Disassembly; Break; Disassembly
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY THE TIP WAS FOUND BROKEN OFF. THERE WAS NOT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7607433 ↗MalfunctionJun 14, 2018Disassembly; Disassembly
IT WAS REPORTED THAT THE SHORT POINTER BROKE OFF AT THE USE FACILITY. THERE WAS NOT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7601008 ↗MalfunctionJun 5, 2018Break; Break
IT WAS REPORTED THAT DURING RECEIPT AT THE USER FACILITY, THE TIP WAS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING RECEIPT OF THE PRODUCT AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7568966 ↗MalfunctionJun 1, 2017Disassembly; Disassembly
DURING DEVICE INSPECTION AT A CADAVER LAB THE TIP OF THE DEVICE WAS FOUND TO BE BROKEN OFF. NO PATIENT INVOLVEMENT OR PROCEDURAL DELAYS WERE REPORTED WITH THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 6604304 ↗MalfunctionMay 30, 2017Break; Break
THE TIP OF THE DEVICE WAS FOUND TO BE BROKEN OFF DURING STERILE PROCESSING AT THE USER FACILITY. THE INITIAL REPORTER INDICATED THAT THE EVENT WAS NOT ASSOCIATED WITH ANY ADVERSE CONSEQUENCES OR PROCEDURAL DELAYS.
Patient problems: No Patient Involvement
MAUDE report 6598475 ↗MalfunctionMay 19, 2017Fracture; Fracture
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE USER FACILITY THE TIP OF THE DEVICE BROKE OFF. NO PATIENT INVOLVEMENT OR PROCEDURAL DELAYS WERE REPORTED WITH THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 6577742 ↗
Other Neurological stereotaxic Instrument devices
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument4 deaths4,668 reports
- Stealthstation® S7™ 9733856Medtronic Navigation, Inc. · Neurological stereotaxic Instrument6 deaths3,908 reports
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 death1,896 reports
- StealthStation S8 9735669Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1,831 reports
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument2 deaths1,542 reports
- Fusion™ ENT Image Guidance System 9733560XOMMedtronic Navigation, Inc. · Neurological stereotaxic Instrument1,316 reports
- Stealthstation® S7™ 9733858Medtronic Navigation, Inc. · Neurological stereotaxic Instrument8 deaths984 reports
- Stealthstation® S7™ 9733858Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 death695 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page