NA 6000-012-000Recalls & reported problems

Stryker Leibinger GmbH & Co. KG · SMALL POINTER · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Stryker Small Pointer is a surgical instrument used to help guide brain surgery. Most of the 33 reports describe the pointer tip or shaft breaking, snapping off, or coming apart, sometimes during calibration, a procedure, shipping, or service checks, plus a few cases of the pointer not measuring accurately. None of the reports describe a death or an injury, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
32
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 04546540245113), received between Feb 16, 2017 and Aug 8, 2024 — classified as 32 malfunctions.

The most frequently reported specific device problem is “Fracture” (19 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture19
  • Break9
  • Disassembly8
  • Imprecision1
  • Incorrect Measurement1
  • Mechanical Jam1

Patient problems

  • No Consequences Or Impact To Patient18
  • No Patient Involvement9
  • No Clinical Signs, Symptoms or Conditions4
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 8, 2024Imprecision; Imprecision

    IT WAS REPORTED THAT THERE WERE ACCURACY ISSUES DURING A CASE WHEN USING THE SMALL POINTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19938373 ↗
  • MalfunctionOct 2, 2023Fracture; Mechanical Jam; Fracture; Mechanical Jam

    PER THE CUSTOMER, THE SHORT POINTER WAS BROKEN DURING A PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17854070 ↗
  • MalfunctionMay 11, 2023Fracture; Fracture

    PER THE CUSTOMER, THE POINTER SNAPPED OFF DURING THE CASE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16912133 ↗
  • MalfunctionOct 18, 2022Incorrect Measurement; Incorrect Measurement

    PER THE CUSTOMER THE DEVICE WAS INACCURATE DURING REGISTRATION DURING A PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE SAME DEVICE WITHOUT CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15628726 ↗
  • MalfunctionOct 10, 2019Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT THE POINTER SHAFT WAS BROKEN OFF DURING SHIPPING; THIS POSES THE RISK OF A SMALL COMPONENT BEING LOST IN THE SURGICAL SITE. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED UPON RECEIPT AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Patient Involvement

    MAUDE report 9175660 ↗
  • MalfunctionSep 25, 2018Break; Break

    IT WAS REPORTED THAT DURING RECEIPT OF THE PRODUCT AT THE USER FACILITY IT WAS OBSERVED THAT THE BATTERY HOLDER IS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED WHEN THE PRODUCT WAS RECEIVED FROM TRANSIT, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7905782 ↗
  • MalfunctionSep 18, 2018Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT DURING THE SERVICE EVALUATION PROCESS CONDUCTED AT THE MANUFACTURER FACILITY IT WAS OBSERVED THAT THE TIP OF THE POINTER WAS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7885745 ↗
  • MalfunctionSep 18, 2018Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT DURING THE SERVICE EVALUATION PROCESS CONDUCTED AT THE MANUFACTURER FACILITY IT WAS OBSERVED THAT THE TIP OF THE POINTER WAS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7885742 ↗
  • MalfunctionAug 27, 2018Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT DURING THE SERVICE EVALUATION PROCESS CONDUCTED AT THE MANUFACTURER FACILITY IT WAS OBSERVED THAT THE TIP OF THE DEVICE IS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7818886 ↗
  • MalfunctionAug 24, 2018Fracture; Fracture

    IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE USER FACILITY IT WAS OBSERVED THAT THE TIP OF DEVICE WAS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7812658 ↗
  • MalfunctionAug 22, 2018Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE USER FACILITY IT WAS OBSERVED THAT THE TIP OF DEVICE IS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7805944 ↗
  • MalfunctionAug 20, 2018Fracture; Fracture

    IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE MANUFACTURER FACILITY IT WAS OBSERVED THAT THE TIP OF DEVICE IS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7797005 ↗
  • MalfunctionAug 20, 2018Fracture; Fracture

    IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE MANUFACTURER FACILITY IT WAS OBSERVED THAT THE TIP OF DEVICE IS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7797000 ↗
  • MalfunctionAug 20, 2018Fracture; Fracture

    IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE MANUFACTURER FACILITY IT WAS OBSERVED THAT THE TIP OF DEVICE IS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7796998 ↗
  • MalfunctionJul 26, 2018Fracture; Fracture

    IT WAS REPORTED THAT DURING SETUP PRIOR TO A SURGICAL PROCEDURE AT THE USER FACILITY THAT THE SHORT POINTER FELL OFF, AND THE TIP BROKE OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING SET UP AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7721901 ↗
  • MalfunctionJul 10, 2018Fracture; Fracture

    IT WAS REPORTED THAT DURING THE SERVICE EVALUATION PROCESS CONDUCTED AT THE MANUFACTURER FACILITY IT WAS OBSERVED THAT THE TIP OF THE DEVICE IS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7674211 ↗
  • MalfunctionJul 10, 2018Fracture; Fracture

    IT WAS REPORTED THAT DURING THE SERVICE EVALUATION PROCESS CONDUCTED AT THE MANUFACTURER FACILITY IT WAS OBSERVED THAT THE TIP OF THE DEVICE IS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7674203 ↗
  • MalfunctionJul 10, 2018Fracture; Fracture

    IT WAS REPORTED THAT DURING THE SERVICE EVALUATION PROCESS CONDUCTED AT THE MANUFACTURER FACILITY IT WAS OBSERVED THAT THE TIP OF THE DEVICE IS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7674201 ↗
  • MalfunctionJul 9, 2018Fracture; Fracture

    IT WAS REPORTED THAT DURING THE SERVICE EVALUATION PROCESS CONDUCTED AT THE MANUFACTURER FACILITY IT WAS OBSERVED THAT THE TIP OF THE DEVICE IS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7671074 ↗
  • MalfunctionJun 15, 2018Break; Disassembly; Break; Disassembly

    IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY THE TIP WAS FOUND BROKEN OFF. THERE WAS NOT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7607433 ↗
  • MalfunctionJun 14, 2018Disassembly; Disassembly

    IT WAS REPORTED THAT THE SHORT POINTER BROKE OFF AT THE USE FACILITY. THERE WAS NOT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7601008 ↗
  • MalfunctionJun 5, 2018Break; Break

    IT WAS REPORTED THAT DURING RECEIPT AT THE USER FACILITY, THE TIP WAS BROKEN OFF.  NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING RECEIPT OF THE PRODUCT AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7568966 ↗
  • MalfunctionJun 1, 2017Disassembly; Disassembly

    DURING DEVICE INSPECTION AT A CADAVER LAB THE TIP OF THE DEVICE WAS FOUND TO BE BROKEN OFF. NO PATIENT INVOLVEMENT OR PROCEDURAL DELAYS WERE REPORTED WITH THIS EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 6604304 ↗
  • MalfunctionMay 30, 2017Break; Break

    THE TIP OF THE DEVICE WAS FOUND TO BE BROKEN OFF DURING STERILE PROCESSING AT THE USER FACILITY. THE INITIAL REPORTER INDICATED THAT THE EVENT WAS NOT ASSOCIATED WITH ANY ADVERSE CONSEQUENCES OR PROCEDURAL DELAYS.

    Patient problems: No Patient Involvement

    MAUDE report 6598475 ↗
  • MalfunctionMay 19, 2017Fracture; Fracture

    IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE USER FACILITY THE TIP OF THE DEVICE BROKE OFF. NO PATIENT INVOLVEMENT OR PROCEDURAL DELAYS WERE REPORTED WITH THIS EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 6577742 ↗

Other Neurological stereotaxic Instrument devices

All Neurological stereotaxic Instrument devices →