NA 5100120450Recalls & reported problems
Stryker Corporation · SD/PD Series Medium Straight Attachment · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Stryker SD/PD series straight surgical drill attachment is a handpiece used in ENT surgery. Of 54 adverse event reports, most describe the attachment overheating, often during testing or demonstration before a procedure or while in use; a few mention parts coming loose, material breaking off, or fluid leaking. There are no reported deaths and no reported injuries, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 54
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 54 adverse event reports that list this exact device (DI 04546540369956), received between Jul 24, 2018 and Oct 2, 2023 — classified as 54 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (50 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device50
- Detachment of Device or Device Component2
- Fluid/Blood Leak2
- Material Fragmentation2
Patient problems
- No Patient Involvement29
- No Clinical Signs, Symptoms or Conditions26
- No Consequences Or Impact To Patient7
- No Known Impact Or Consequence To Patient2
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 54 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 2, 2023Overheating of Device; Overheating of Device
THE USER FACILITY REPORTED THAT THE DEVICE OVERHEATED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED. PATIENT INVOLVEMENT WAS NOT REPORTED AT THIS TIME.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 17855556 ↗MalfunctionSep 25, 2023Overheating of Device; Overheating of Device
THE USER FACILITY REPORTED THAT THE DEVICE OVERHEATED AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17813722 ↗MalfunctionJun 6, 2022Overheating of Device; Overheating of Device
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14604712 ↗MalfunctionApr 21, 2022Overheating of Device; Overheating of Device
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14169951 ↗MalfunctionApr 13, 2022Overheating of Device; Overheating of Device
THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING DEMONSTRATED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14106528 ↗MalfunctionOct 6, 2021Overheating of Device; Overheating of Device
THE USER FACILITY REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12586543 ↗MalfunctionJun 16, 2021Overheating of Device; Overheating of Device
THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING DEMONSTRATED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12012527 ↗MalfunctionJun 16, 2021Overheating of Device; Overheating of Device
THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING DEMONSTRATED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12012480 ↗MalfunctionJun 8, 2021Overheating of Device; Overheating of Device
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11961030 ↗MalfunctionApr 29, 2021Overheating of Device; Overheating of Device
THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING DEMONSTRATED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11747098 ↗MalfunctionApr 29, 2021Overheating of Device; Overheating of Device
THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING DEMONSTRATED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11747092 ↗MalfunctionApr 7, 2021Overheating of Device; Overheating of Device
THE MANUFACTURER REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING DEMONSTRATED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11628816 ↗MalfunctionFeb 25, 2021Overheating of Device; Overheating of Device
THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11379873 ↗MalfunctionNov 17, 2020Overheating of Device; Overheating of Device
THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED DURING A DEMO VISIT AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10854293 ↗MalfunctionNov 13, 2020Overheating of Device; Overheating of Device
THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10836881 ↗MalfunctionOct 16, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
THE MANUFACTURER SERVICE TECHNICIAN REPORTED THAT THE DEVICE HAD A FRACTURED FOOT. THE EVENT OCCURRED DURING FUNCTIONAL TESTING WHILE IT WAS BEING SERVICED AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10693219 ↗MalfunctionOct 1, 2020Overheating of Device; Overheating of Device
IT WAS REPORTED THE DEVICE WAS OVERHEATING DURING TESTING CONDUCTED AT THE USER FACILITY. THE USER FACILITY REPORTED THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10614026 ↗MalfunctionAug 20, 2020Overheating of Device; Overheating of Device
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED AT THE USER FACILITY. THERE WAS NO KNOWN PATIENT IMPACT, NO KNOWN DELAY, NO KNOWN MEDICAL INTERVENTION, AND NO KNOWN ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10434250 ↗MalfunctionJul 31, 2020Overheating of Device; Overheating of Device
THE MANUFACTURER SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED. THE EVENT OCCURRED DURING FUNCTIONAL TESTING WHILE IT WAS BEING SERVICED AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 10350104 ↗MalfunctionApr 14, 2020Overheating of Device; Overheating of Device
THE MANUFACTURER REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 9957180 ↗MalfunctionMar 31, 2020Overheating of Device; Overheating of Device
THE MANUFACTURER REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 9907831 ↗MalfunctionMar 31, 2020Overheating of Device; Overheating of Device
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 9907826 ↗MalfunctionMar 31, 2020Overheating of Device; Overheating of Device
THE MANUFACTURER REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 9907818 ↗MalfunctionMar 23, 2020Overheating of Device; Overheating of Device
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Patient Involvement
MAUDE report 9867898 ↗MalfunctionMar 23, 2020Overheating of Device; Overheating of Device
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Patient Involvement
MAUDE report 9867886 ↗
Other DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices
- RemB 6400034000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1,618 reports
- NA XMAXAnspach Effort, The · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1,225 reports
- NA EG1ASynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE859 reports
- NA CRANI-ASynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1 recall772 reports
- NA 5450400000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE775 reports
- NA 05.001.010Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE770 reports
- RemB 6400015000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE603 reports
- NA QD11Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE580 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page