Skip to content

HOFFMANN 5079-6-040Recalls & reported problems

Stryker GmbH · Connecting Rod Micro · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Stryker Hoffmann connecting rod is named in 40 adverse event reports, with no deaths or injuries; the reports mostly describe cracks in the rod, sometimes after cleaning, sterilization, reprocessing, or reuse. One open recall starting in July 2026 concerns potential cracking of Hoffmann II Carbon Connecting Rods after cleaning and sterilization. There were 40 reports in the last 12 months and none in the prior 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
21
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 21 adverse event reports that list this exact device (DI 04546540389909), received between May 28, 2026 and Aug 27, 2026 — classified as 21 malfunctions.

The most frequently reported specific device problem is “Crack” (21 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 11 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 10 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202611
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 1, 2026Open, Classified

    Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.

    Z-2828-2026 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Crack21

Patient problems

  • No Clinical Signs, Symptoms or Conditions21

Recent adverse event reports

The 21 most recent of 21 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 27, 2026Crack

    It was reported that, cracks were seen on the product after inspection.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50478958 ↗
  • MalfunctionAug 27, 2026Crack

    It was reported that, cracks were seen on the product after inspection.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50478947 ↗
  • MalfunctionAug 27, 2026Crack

    It was reported that, cracks were seen on the product after inspection.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50478692 ↗
  • MalfunctionAug 26, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50456982 ↗
  • MalfunctionAug 24, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50418737 ↗
  • MalfunctionAug 24, 2026Crack

    Cracks identified in Hoffmann Micro carbon rods.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50418734 ↗
  • MalfunctionAug 24, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50418717 ↗
  • MalfunctionAug 24, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50410157 ↗
  • MalfunctionAug 20, 2026Crack

    It was reported that, there were cracks on the rods.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50347175 ↗
  • MalfunctionAug 20, 2026Crack

    It was reported that, there were cracks on the rods.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50347174 ↗
  • MalfunctionMay 28, 2026Crack

    AS REPORTED: "CARBON RODS SHOW VISIBLE CRACKS AFTER STERILIZATION/CLEANING (REPROCESSING)'.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25322155 ↗
  • MalfunctionMay 28, 2026Crack

    AS REPORTED: "CARBON RODS SHOW VISIBLE CRACKS AFTER STERILIZATION/CLEANING (REPROCESSING)".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25322143 ↗
  • MalfunctionMay 28, 2026Crack

    AS REPORTED: "CARBON RODS SHOW VISIBLE CRACKS AFTER STERILIZATION/CLEANING (REPROCESSING)".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25321949 ↗
  • MalfunctionMay 28, 2026Crack

    AS REPORTED: "CARBON RODS SHOW VISIBLE CRACKS AFTER STERILIZATION/CLEANING (REPROCESSING)".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25321927 ↗
  • MalfunctionMay 28, 2026Crack

    AS REPORTED: "CARBON RODS SHOW SMALL CRACKS AND SWELLING AFTER THE FIRST CLEANING/STERILIZATION CYCLE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25321716 ↗
  • MalfunctionMay 28, 2026Crack

    IT WAS REPORTED THAT THE REPLACEMENT DELIVERY OF THE CARBON RODS WERE RECEIVED. THEY WERE PROCESSED ONCE IN THE REPROCESSING/STERILIZATION UNIT FOR MEDIZINPRODUKTE (AEMP (AUFBEREITUNGSEINHEIT FOR MEDIZINPRODUKTE)), UNFORTUNATELY THE CARBON RODS BROKE AGAIN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25321555 ↗
  • MalfunctionMay 28, 2026Crack

    IT WAS REPORTED THAT THE REPLACEMENT DELIVERY OF THE CARBON RODS WERE RECEIVED. THEY WERE PROCESSED ONCE IN THE REPROCESSING/STERILIZATION UNIT FOR MEDIZINPRODUKTE (AEMP (AUFBEREITUNGSEINHEIT FOR MEDIZINPRODUKTE)), UNFORTUNATELY THE CARBON RODS BROKE AGAIN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25321537 ↗
  • MalfunctionMay 28, 2026Crack

    IT WAS REPORTED THAT THE REPLACEMENT DELIVERY OF THE CARBON RODS WERE RECEIVED. THEY WERE PROCESSED ONCE IN THE REPROCESSING/STERILIZATION UNIT FOR MEDIZINPRODUKTE (AEMP (AUFBEREITUNGSEINHEIT FOR MEDIZINPRODUKTE)), UNFORTUNATELY THE CARBON RODS BROKE AGAIN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25320943 ↗
  • MalfunctionMay 28, 2026Crack

    IT WAS REPORTED THAT THE REPLACEMENT DELIVERY OF THE CARBON RODS WERE RECEIVED. THEY WERE PROCESSED ONCE IN THE REPROCESSING/STERILIZATION UNIT FOR MEDIZINPRODUKTE (AEMP (AUFBEREITUNGSEINHEIT FOR MEDIZINPRODUKTE)), UNFORTUNATELY THE CARBON RODS BROKE AGAIN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25320923 ↗
  • MalfunctionMay 28, 2026Crack

    AS REPORTED: ''CARBON RODS SHOW SMALL CRACKS AND SWELLING AFTER THE FIRST CLEANING/STERILIZATION CYCLE.''

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25318334 ↗
  • MalfunctionMay 28, 2026Crack

    AS REPORTED: "CARBON RODS SHOW SMALL CRACKS AND SWELLING AFTER THE FIRST CLEANING/STERILIZATION CYCLE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25318330 ↗

Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices

All APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.