HOFFMANN 5079-6-150Recalls & reported problems
Stryker GmbH · Connecting Rod Micro · Class 2
All models of HOFFMANN (45)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04546540389954), received between Aug 24, 2026 and Aug 28, 2026 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Crack” (10 reports).
1 FDA recall record name this device's identifier, 1 still open.
6 other recalls cover devices cleared under the same FDA submission (K033476); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jul 1, 2026Class IIOpen, Classified
Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.
Z-2833-2026 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Crack10
Patient problems
- No Clinical Signs, Symptoms or Conditions10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 28, 2026Crack
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50501591 ↗MalfunctionAug 28, 2026Crack
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50501588 ↗MalfunctionAug 27, 2026Crack
It was reported that, cracks were seen on the product after inspection.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50478247 ↗MalfunctionAug 27, 2026Crack
As reported: "Cracks identified in Hoffmann Micro carbon rods."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50476264 ↗MalfunctionAug 27, 2026Crack
As reported: "Cracks identified in Hoffmann Micro carbon rods."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50475928 ↗MalfunctionAug 26, 2026Crack
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50459570 ↗MalfunctionAug 24, 2026Crack
Cracks identified in Hoffmann Micro carbon rods.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50418746 ↗MalfunctionAug 24, 2026Crack
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50418727 ↗MalfunctionAug 24, 2026Crack
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50418725 ↗MalfunctionAug 24, 2026Crack
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50418715 ↗
FDA clearance and classification
Cleared through Special 510(k) K033476 on Jan 20, 2004 (decision: Substantially Equivalent). Submitted by Howmedica Osteonics Corp. as “Modification To Hoffmann Ii Micro External Fixation System”. The FDA review took 78 days. FDA record ↗
The FDA regulates product code KTT (appliance, fixation, nail/blade/plate combination, multiple component) under 21 CFR 888.3030 as a Class II device. 21 CFR 888.3030 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K033476. One submission often covers many models, so check the product and lot details in each record.
- Z-2830-2026 · Jul 1, 2026 · Open, ClassifiedStryker Hoffmann II Connecting Rod Micro 3 x 60mm. Catalog Number: 50796060.
- Z-2831-2026 · Jul 1, 2026 · Open, ClassifiedStryker Hoffmann II Connecting Rod Micro 3 x 90mm. Catalog Number: 50796090.
- Z-2829-2026 · Jul 1, 2026 · Open, ClassifiedStryker Hoffmann II Connecting Rod Micro 3 x 50mm. Catalog Number: 50796050.
- Z-2828-2026 · Jul 1, 2026 · Open, ClassifiedStryker Hoffmann II Connecting Rod Micro 3 x 40mm. Catalog Number: 50796040.
- Z-2827-2026 · Jul 1, 2026 · Open, ClassifiedStryker Hoffmann II Connecting Rod Micro 3 x 30mm. Catalog Number: 50796030.
- Z-2832-2026 · Jul 1, 2026 · Open, ClassifiedStryker Hoffmann II Connecting Rod Micro 3 x 120mm. Catalog Number: 50796120.
Common questions about the HOFFMANN 5079-6-150
Has the HOFFMANN 5079-6-150 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2833-2026 (Class II), began on Jul 1, 2026: Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the HOFFMANN 5079-6-150?
The FDA's MAUDE database lists 10 adverse event reports for this device, received between Aug 24, 2026 and Aug 28, 2026. In those reports, the most common specific device problem is “Crack” (10). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the HOFFMANN 5079-6-150?
None of the reports are classified as deaths, 0 as injuries and 10 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the HOFFMANN 5079-6-150 FDA approved?
It was cleared through Special 510(k) K033476 on Jan 20, 2004. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the HOFFMANN 5079-6-150, and is it still sold?
The FDA's device database lists Stryker GmbH as the company responsible for it (brand HOFFMANN, model 5079-6-150). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices
- PRECICE® Plating System LP4.5-DLF31Nuvasive Specialized Orthopedics, Inc. · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component3 recalls0 reports
- R3 71364451Smith & Nephew, Inc. · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT145 reports
- DFS HA BONE SCREWS HAA60-15050Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-11030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15020Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-18030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15060Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page