Skip to content

HOFFMANN 5079-6-150Recalls & reported problems

Stryker GmbH · Connecting Rod Micro · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04546540389954), received between Aug 24, 2026 and Aug 28, 2026 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Crack” (10 reports).

1 FDA recall record name this device's identifier, 1 still open.

6 other recalls cover devices cleared under the same FDA submission (K033476); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 10 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 1, 2026Open, Classified

    Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.

    Z-2833-2026 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Crack10

Patient problems

  • No Clinical Signs, Symptoms or Conditions10

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 28, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50501591 ↗
  • MalfunctionAug 28, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50501588 ↗
  • MalfunctionAug 27, 2026Crack

    It was reported that, cracks were seen on the product after inspection.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50478247 ↗
  • MalfunctionAug 27, 2026Crack

    As reported: "Cracks identified in Hoffmann Micro carbon rods."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50476264 ↗
  • MalfunctionAug 27, 2026Crack

    As reported: "Cracks identified in Hoffmann Micro carbon rods."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50475928 ↗
  • MalfunctionAug 26, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50459570 ↗
  • MalfunctionAug 24, 2026Crack

    Cracks identified in Hoffmann Micro carbon rods.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50418746 ↗
  • MalfunctionAug 24, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50418727 ↗
  • MalfunctionAug 24, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50418725 ↗
  • MalfunctionAug 24, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50418715 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K033476. One submission often covers many models, so check the product and lot details in each record.

  • Z-2832-2026 · Jul 1, 2026 · Open, Classified
    Stryker Hoffmann II Connecting Rod Micro 3 x 120mm. Catalog Number: 50796120.
  • Z-2827-2026 · Jul 1, 2026 · Open, Classified
    Stryker Hoffmann II Connecting Rod Micro 3 x 30mm. Catalog Number: 50796030.
  • Z-2831-2026 · Jul 1, 2026 · Open, Classified
    Stryker Hoffmann II Connecting Rod Micro 3 x 90mm. Catalog Number: 50796090.
  • Z-2829-2026 · Jul 1, 2026 · Open, Classified
    Stryker Hoffmann II Connecting Rod Micro 3 x 50mm. Catalog Number: 50796050.
  • Z-2828-2026 · Jul 1, 2026 · Open, Classified
    Stryker Hoffmann II Connecting Rod Micro 3 x 40mm. Catalog Number: 50796040.
  • Z-2830-2026 · Jul 1, 2026 · Open, Classified
    Stryker Hoffmann II Connecting Rod Micro 3 x 60mm. Catalog Number: 50796060.

Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices

All APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.