HOFFMANN 5079-6-150Recalls & reported problems
Stryker GmbH · Connecting Rod Micro · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04546540389954), received between Aug 24, 2026 and Aug 28, 2026 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Crack” (10 reports).
1 FDA recall record name this device's identifier, 1 still open.
6 other recalls cover devices cleared under the same FDA submission (K033476); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jul 1, 2026Open, Classified
Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.
Z-2833-2026 · Root cause: Device Design · FDA record ↗
Most reported problems
Device problems
- Crack10
Patient problems
- No Clinical Signs, Symptoms or Conditions10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 28, 2026Crack
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50501591 ↗MalfunctionAug 28, 2026Crack
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50501588 ↗MalfunctionAug 27, 2026Crack
It was reported that, cracks were seen on the product after inspection.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50478247 ↗MalfunctionAug 27, 2026Crack
As reported: "Cracks identified in Hoffmann Micro carbon rods."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50476264 ↗MalfunctionAug 27, 2026Crack
As reported: "Cracks identified in Hoffmann Micro carbon rods."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50475928 ↗MalfunctionAug 26, 2026Crack
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50459570 ↗MalfunctionAug 24, 2026Crack
Cracks identified in Hoffmann Micro carbon rods.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50418746 ↗MalfunctionAug 24, 2026Crack
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50418727 ↗MalfunctionAug 24, 2026Crack
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50418725 ↗MalfunctionAug 24, 2026Crack
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50418715 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K033476. One submission often covers many models, so check the product and lot details in each record.
- Z-2832-2026 · Jul 1, 2026 · Open, ClassifiedStryker Hoffmann II Connecting Rod Micro 3 x 120mm. Catalog Number: 50796120.
- Z-2827-2026 · Jul 1, 2026 · Open, ClassifiedStryker Hoffmann II Connecting Rod Micro 3 x 30mm. Catalog Number: 50796030.
- Z-2831-2026 · Jul 1, 2026 · Open, ClassifiedStryker Hoffmann II Connecting Rod Micro 3 x 90mm. Catalog Number: 50796090.
- Z-2829-2026 · Jul 1, 2026 · Open, ClassifiedStryker Hoffmann II Connecting Rod Micro 3 x 50mm. Catalog Number: 50796050.
- Z-2828-2026 · Jul 1, 2026 · Open, ClassifiedStryker Hoffmann II Connecting Rod Micro 3 x 40mm. Catalog Number: 50796040.
- Z-2830-2026 · Jul 1, 2026 · Open, ClassifiedStryker Hoffmann II Connecting Rod Micro 3 x 60mm. Catalog Number: 50796060.
Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices
- PRECICE® Plating System LP4.5-DLF31Nuvasive Specialized Orthopedics, Inc. · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component3 recalls0 reports
- R3 71364451Smith & Nephew, Inc. · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT145 reports
- DFS HA BONE SCREWS HAA60-15040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15050Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAB60-12040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-13040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-18030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.