Skip to content

NA 6000-999-002Recalls & reported problems

Stryker Leibinger GmbH & Co. KG · NAVLOCK 2-7MM · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 04546540391902), received between Aug 2, 2018 and Dec 12, 2023 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Break” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break7
  • Fracture5
  • Detachment of Device or Device Component2

Patient problems

  • No Consequences Or Impact To Patient8
  • No Patient Involvement3
  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 12, 2023Fracture; Fracture

    PER THE CUSTOMER, THE METAL IS DAMAGED, AND A PIECE OF METAL BROKE OFF IT INTRAOPERATIVELY. ADDITIONAL INFORMATION HAS BEEN REQUESTED FROM THE USER FACILITY.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 18310005 ↗
  • MalfunctionOct 22, 2019Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING SET UP IT WAS OBSERVED THAT SCREW BROKE OFF; THIS POSES THE RISK OF A SMALL COMPONENT BEING LOST IN THE SURGICAL SITE. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING SETUP AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement

    MAUDE report 9221167 ↗
  • MalfunctionJun 27, 2019Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component

    IT WAS REPORTED THAT, DURING TESTING, IT WAS OBSERVED THAT THE LOCKING MECHANISM IS MISSING; THIS POSES THE RISK OF A SMALL COMPONENT BEING LOST IN THE SURGICAL SITE. . NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement

    MAUDE report 8741571 ↗
  • MalfunctionMay 9, 2019Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE USER FACILITY PRIOR TO THE START OF A SURGICAL PROCEDURE THE DEVICE WAS FOUND TO HAVE A BROKEN TIP.   NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient; No Patient Involvement

    MAUDE report 8596341 ↗
  • MalfunctionApr 29, 2019Break; Break

    IT WAS REPORTED THAT DURING THE SERVICE EVALUATION PROCESS CONDUCTED AT THE MANUFACTURER FACILITY IT WAS OBSERVED THAT THE HOOK OF THE DEVICE WAS BROKEN OFF. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8560140 ↗
  • MalfunctionDec 18, 2018Break; Break

    IT WAS REPORTED THAT DURING SET UP AT THE USER FACILITY THE UNIT WAS FOUND TO BE BROKEN. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8175858 ↗
  • MalfunctionDec 13, 2018Break; Break

    IT WAS REPORTED THAT DURING SET UP AT THE USER FACILITY THE SMALL DISK WAS DISCOVERED TO BE BROKEN OFF. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE SAME DEVICE WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8159967 ↗
  • MalfunctionNov 13, 2018Break; Break

    IT WAS REPORTED THAT DURING SET UP AT THE USER FACILITY THE SMALL DISK WAS DISCOVERED TO BE BROKEN OFF. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE SAME DEVICE WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8064447 ↗
  • MalfunctionNov 7, 2018Break; Break

    IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY A PIECE BROKE OFF THE DEVICE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE SAME DEVICE WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8047767 ↗
  • MalfunctionOct 2, 2018Fracture; Fracture

    IT WAS REPORTED THAT DURING SET UP AT THE USER FACILITY THE SMALL DISK WAS DISCOVERED TO BE BROKEN OFF. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE SAME DEVICE WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7925514 ↗
  • MalfunctionAug 2, 2018Break; Break

    IT WAS REPORTED THAT PRIOR TO A CASE AT THE USER FACILITY THE UNIT FELL APART. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7742006 ↗

Other Neurological stereotaxic Instrument devices

All Neurological stereotaxic Instrument devices →

Sources and updates