Alumina 6565-0-036Recalls & reported problems
Howmedica Osteonics Corp. · Alumina Ceramic V40 Femoral Head · Class 2 · Implant
All models of Alumina (16)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04546540402141), received between Feb 10, 2020 and Apr 3, 2026 — classified as 4 injuries, 1 malfunction.
The most frequently reported specific device problem is “Fracture” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K023901, K173499, K243784); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture2
- Insufficient Device Problem Information2
- Noise, Audible2
- Degraded1
Patient problems
- Pain3
- Metal Related Pathology2
- Failure of Implant1
- Inflammation1
- Injury1
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- Reaction1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryApr 3, 2026Noise, Audible
THE FOLLOWING INFORMATION WAS PROVIDED: HIP REVISION OF CERAMIC ON CERAMIC HIP DUE TO PATIENT COMPLAINT OF "SQUEAKING." LEFT HIP.
Patient problems: Insufficient Information
MAUDE report 24779753 ↗MalfunctionFeb 26, 2024Fracture; Fracture
ON JANUARY 29, 2024, AN ADVERSE EVENT OCCURRED DURING THE IMPLANTATION OF THE CERAMIC ALUMINA HEAD REF: 6565-0-036, SN:(B)(6). WHILE ADJUSTING THE HIP JOINT AND CHECKING THE PROPER ROTATION OF THE JOINT ( ABDUCTION, ADDUCTION OF THE HIP JOINT), THE ALUMINA HEAD THAT HAD ALREADY BEEN IMPLANTED ON THE ACCOLADE II STEM BROKE. SINCE THE SAME HEAD ( ALUMINA IN SIZE 36) WAS AVAILABLE AT THE STRYKER IMPLANT COMMISSARY IN THE OPERATING THEATER, THE PROCEDURE WAS COMPLETED WITHOUT A PROBLEM AND AT THE RIGHT TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18785513 ↗InjuryOct 5, 2020Insufficient Device Problem Information; Noise, Audible; Insufficient Device Problem Information; Noise, Audible
REVISION SURGERY ON STRYKER IMPLANTS. REPLACED WITH STRYKER STEM AND FEMORAL HEAD AND DEPUY CUP AND INSERT ON (B)(6) 2020. PATIENT COMPLAINT OF PAIN AND SQUEEK / CLUNKINESS. SIGNS OF TRUNIONOSIS AND IMPINGEMENT NOTED BY SURGEON. STRIPE WEAR ON CERAMIC HEAD. LINER RIM DAMAGED BY WEAR FROM STEM. REVISION ON PREVIOUS PRIMARY SURGERY.
Patient problems: Metal Related Pathology; Pain
MAUDE report 10631086 ↗InjurySep 22, 2020Degraded; Degraded
IT HAS BEEN REPORTED BY THE CUSTOMER THAT THE IMPLANTS WERE REMOVED AFTER 1 YEAR AND 4 MONTHS IMPLANTED IN THE PATIENT DUE TO WEAR. UPDATE: ACCORDING TO DR, "PRIOR TO THE INCIDENT THERE WERE NO ACCIDENTS WHAT'S OVER OR OTHER SURGERIES. THE PATIENT PERFORMED REGULAR DAILY ACTIVITY (NO SPORTS) INCLUDING WORK IN HIS RESTAURANT. THE PATIENT FIRST CAME TO SEE ME ON (B)(6) IN 2019 FOR THE NOISE HE NOTICED WHILE WALKING...I SAW THE PATIENT AGAIN ON (B)(6) IN 2019 FOR NOW PAINFUL MOVEMENT OFT HE LEFT HIP AND AN X-RAY WAS PERFORMED...THE REVISION SURGERY FORT HE LEFT HIP COULD BE PERFORMED AS PLANNED ON (B)(6) IN 2020...((B)(6) 2020) THERE WAS, AS EXPECTED, A STRONG METALLOSIS OFT HE SURROUNDING SOFT TISSUE AT THE LEFT HIP. CUP, PE AND FEMORAL HEAD WERE REMOVED ENTIRELY FOR MATERIAL TESTING. AFTER TAKING SEVERAL BONE AND SOFT TISSUE SAMPLES OF THE SUPERFICIAL AND DEEP LAYERS OF TISSUE, AN EXTENSIVE BONE AND SOFT TISSUE DEBRIDEMENT AND JET-LAVAGE WERE PERFORMED. THE REPLACEMENT CUP COULD BE POSITIONED WITHOUT PROBLEMS AND WAS FIXED WITH ADDITIONAL CANCELLOUS BONE SCREWS. AFTER INSERTING THE PE INSERT AND TESTING FOR CORRECT PLACEMENT AND STABILITY, A CERAMIC V40 HEAD 36MM WAS USED AND JOINT
Patient problems: Failure of Implant; Pain
MAUDE report 10566697 ↗InjuryFeb 10, 2020Fracture; Insufficient Device Problem Information; Fracture; Insufficient Device Problem Information
IT WAS REPORTED THROUGH THE FILING OF A LAWSUIT THAT ALLEGEDLY THE PATIENT WAS IMPLANTED WITH AN LFIT ANATOMIC COCR V40 FEMORAL HEAD ON HIS LEFT HIP ON OR ABOUT (B)(6) 2005 AND THE FEMORAL HEAD WAS EXPLANTED ON (B)(6) 2013. IT IS FURTHER ALLEGED THAT HE SUFFERED INJURIES AS A RESULT OF IMPLANTATION AND EXPLANTATION OF THE DEVICE(S) AT ISSUE AND EXCESSIVE LEVELS OF CHROMIUM AND COBALT IN HIS BLOOD.
Patient problems: Inflammation; Injury; Metal Related Pathology; Pain; Reaction
MAUDE report 9689077 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K023901, K173499, K243784. One submission often covers many models, so check the product and lot details in each record.
- Z-0314-2022 · Oct 15, 2021 · Open, Classified28MM -4 LFIT V40 HEAD, Catalog Number 6260-9-122
- Z-0313-2022 · Oct 15, 2021 · Open, Classified28MM -4 LFIT V40 HEAD, Catalog Number 6260-9-028
Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices
- BIOLOX DELTA 1365-32-310Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented551 reports
- PINNACLE ALTRX 1221-36-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented481 reports
- BIOLOX DELTA 1365-32-320Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death314 reports
- BIOLOX DELTA 1365-36-720Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 death189 reports
- DJO SURGICAL 400-03-362Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented187 reports
- Anthology 71365705Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented176 reports
- BIOLOX DELTA 1365-40-710Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented170 reports
- PINNACLE GRIPTION 1217-32-054Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented169 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page