APEX 5038-2-110SRecalls & reported problems
Stryker GmbH · Self-drilling Half Pin · Class 2 · Implant
All models of APEX (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04546540435408), received between Jul 30, 2018 and Sep 25, 2025 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break5
- Fracture4
Patient problems
- Foreign Body In Patient5
- No Known Impact Or Consequence To Patient2
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjurySep 25, 2025Break; Fracture; Break; Fracture
AS REPORTED: "DURING A PRECISION NAVIGATION TKA CASE. THE APEX PIN SNAPPED INSIDE THE PATIENTS TIBIA DURING INSERTION AT THE START OF THE CASE. THE BROKEN END COULD NOT BE RETRIEVED AND A NEW PIN WAS OPENED AND USED. SURGICAL DELAY WHILE TRYING TO RETRIEVE THE BROKEN END."
Patient problems: Foreign Body In Patient
MAUDE report 23144772 ↗MalfunctionNov 13, 2020Fracture; Fracture
AS REPORTED: " 3 X STERILE APEX PINS REF NUMBER, (B)(4) WERE USED IN A NAVIGATED HIGH TIBIAL OSTEOTOMY CASE, INSERTED WITH A WIRED DRIVER, UPON REMOVAL WITH THE WIRE DRIVER POST FINISHING THE CASE IT WAS NOTED THAT THE TIPS OF THE PINS HAD BROKEN OFF. NOTICED AFTER SURGERY WAS COMPLETED AND REMOVING THE PINS."
Patient problems: Foreign Body In Patient
MAUDE report 10840129 ↗MalfunctionNov 13, 2020Fracture; Fracture
AS REPORTED: " 3 X STERILE APEX PINS REF NUMBER , (B)(4) WERE USED IN A NAVIGATED HIGH TIBIAL OSTEOTOMY CASE, INSERTED WITH A WIRED DRIVER, UPON REMOVAL WITH THE WIRE DRIVER POST FINISHING THE CASE IT WAS NOTED THAT THE TIPS OF THE PINS HAD BROKEN OFF. NOTICED AFTER SURGERY WAS COMPLETED AND REMOVING THE PINS."
Patient problems: Foreign Body In Patient
MAUDE report 10839995 ↗MalfunctionNov 13, 2020Fracture; Fracture
AS REPORTED: " 3 X STERILE APEX PINS REF NUMBER , (B)(4) WERE USED IN A NAVIGATED HIGH TIBIAL OSTEOTOMY CASE, INSERTED WITH A WIRED DRIVER, UPON REMOVAL WITH THE WIRE DRIVER POST FINISHING THE CASE IT WAS NOTED THAT THE TIPS OF THE PINS HAD BROKEN OFF. NOTICED AFTER SURGERY WAS COMPLETED AND REMOVING THE PINS."
Patient problems: Foreign Body In Patient
MAUDE report 10839954 ↗MalfunctionAug 29, 2019Break
APEX PINS WERE USED TO ATTACH AN ORTHOLOCK TO THE TIBIA DURING A NAVIGATED TOTAL KNEE. THE TIP OF THE APEX PIN DRILL WAS BROKEN AT THE TIME OF REMOVAL OF THE PIN FROM THE TIBIAL SHAFT AND RETAINED IN THE BONE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 8946783 ↗MalfunctionAug 29, 2019Break
APEX PINS WERE USED TO ATTACH AN ORTHOLOCK TO THE TIBIA DURING A NAVIGATED TOTAL KNEE. THE TIP OF THE APEX PIN DRILL WAS BROKEN AT THE TIME OF REMOVAL OF THE PIN FROM THE TIBIAL SHAFT AND RETAINED IN THE BONE. SURGICAL DELAY. THE SURGEON DECIDED TO X-RAY THE PATIENTS TIBIA TO CONFIRM THE TIP OF THE APEX PIN WAS STILL LODGED IN THE CORTEX.
Patient problems: Foreign Body In Patient
MAUDE report 8946675 ↗MalfunctionJul 30, 2018Break
BROKEN 3 MM APEX PINS DURING REMOVAL FROM BONE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7732920 ↗MalfunctionJul 30, 2018Break
BROKEN 3MM APEX PINS DURING REMOVAL FROM BONE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7732855 ↗
Other PIN, FIXATION, THREADED devices
- ORTHOFIX MODULSYSTEM 1Orthofix Srl · Pin, Fixation, Threaded1 recall2 reports
- ORTHOFIX MODULSYSTEM 1Orthofix Srl · Pin, Fixation, Threaded1 recall0 reports
- ORTHOFIX MODULSYSTEM 1Orthofix Srl · Pin, Fixation, Threaded1 recall0 reports
- ORTHOFIX MODULSYSTEM 1Orthofix Srl · Pin, Fixation, Threaded1 recall0 reports
- APEX 5038-1-080Stryker GmbH · PIN, FIXATION, THREADED48 reports
- Agee Digit Widget® DWD-232Hand Biomechanics Lab Inc · Pin, Fixation, Threaded32 reports
- SPATIAL FRAME 71070720Smith & Nephew, Inc. · PIN, FIXATION, THREADED15 reports
- APEX 5038-2-110Stryker GmbH · PIN, FIXATION, THREADED11 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page