ORTHOLOCK 6007-003-000Recalls & reported problems
Stryker Leibinger GmbH & Co. KG · ORTHOLOCK · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04546540442895), received between Feb 10, 2023 and Sep 24, 2026 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Incorrect Measurement” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect Measurement4
- Improper or Incorrect Procedure or Method1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjurySep 24, 2026Incorrect Measurement
It was reported that during femoral impaction, the Ortholock became loose, resulting in loss of navigation functionality. Consequently, the procedure was completed using conventional manual instrumentation. Due to reduced navigational accuracy following device instability, bone cement was utilized to complete the procedure. No adverse patient outcome was reported.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 50882151 ↗MalfunctionFeb 12, 2026Incorrect Measurement
DURING A PROCEDURE AT THE USER FACILITY, THE HIP NAVI (DURING CUP PLACEMENT), THE ORTHOLOCK LOOSENED, CAUSING A SIGNIFICANT DROP IN ACCURACY, RENDERING THE NAVI UNUSABLE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE SAME DEVICE WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24323474 ↗MalfunctionFeb 12, 2026Incorrect Measurement
DURING A PROCEDURE AT THE USER FACILITY, THE HIP NAVI (DURING CUP PLACEMENT), THE ORTHOLOCK LOOSENED, CAUSING A SIGNIFICANT DROP IN ACCURACY, RENDERING THE NAVI UNUSABLE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE SAME DEVICE WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24323473 ↗MalfunctionFeb 10, 2023Incorrect Measurement; Improper or Incorrect Procedure or Method; Incorrect Measurement; Improper or Incorrect Procedure or Method
PER THE CUSTOMER, LICENSING FOR THE AUTOMATED FUNCTIONALITY PORTION OF THE Q GUIDANCE SYSTEM WAS NOT LOADED CORRECTLY. THE CUSTOMER INSTEAD USED THE NON-AUTOMATED SOFTWARE PACKAGE AVAILABLE WITH THE Q GUIDANCE SYSTEM. WHILE PLACING A HIP PIN, MOVEMENT WAS SHOWN ON THE SCREEN WITHOUT MOVING THE DEVICE. TWO OF THE EIGHT SCREWS WERE SHOWN MEDIAL AFTER FOLLOWING THE PLANNED PROJECTION. PER THE CUSTOMER, IT APPEARED THE SCREWS SHIFTED. PER ADDITIONAL CLARIFICATION, THE ORTHOLOCK PIN SYSTEM WAS USED DURING THE PROCEDURE, HOWEVER; ONLY TWO OF THE THREE SCREWS WERE USED TO FIXATE THE ORTHOLOCK ON THE PATIENT'S ANATOMY, WHICH CAN LEAD TO INACCURACY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16356317 ↗
Other Neurological stereotaxic Instrument devices
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument4 deaths4,668 reports
- Stealthstation® S7™ 9733856Medtronic Navigation, Inc. · Neurological stereotaxic Instrument6 deaths3,908 reports
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 death1,896 reports
- StealthStation S8 9735669Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1,831 reports
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument2 deaths1,542 reports
- Fusion™ ENT Image Guidance System 9733560XOMMedtronic Navigation, Inc. · Neurological stereotaxic Instrument1,316 reports
- Stealthstation® S7™ 9733858Medtronic Navigation, Inc. · Neurological stereotaxic Instrument8 deaths984 reports
- Stealthstation® S7™ 9733858Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 death695 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page