Quick summary

This Stryker Maestro straight "R" attachment is a surgical drill attachment used in ENT procedures. The 30 reports mostly describe the attachment overheating and parts coming loose or detaching, with many found disassembled or missing components during servicing. No deaths and no injuries are reported, and none of the reports describe harm to a patient. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
30
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 04546540469243), received between Jul 24, 2018 and Aug 28, 2025 — classified as 30 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (22 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device22
  • Detachment of Device or Device Component8

Patient problems

  • No Clinical Signs, Symptoms or Conditions22
  • No Patient Involvement6
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 28, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE MANUFACTURER SERVICE TECHNICIAN REPORTED THAT THE DEVICE WAS RECEIVED DISASSEMBLED AND MISSING COMPONENTS WHILE IT WAS BEING SERVICED AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22919274 ↗
  • MalfunctionAug 4, 2025Overheating of Device; Overheating of Device

    THE MANUFACTURER SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING SERVICED AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22706320 ↗
  • MalfunctionJun 4, 2025Overheating of Device; Overheating of Device

    THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING A PROCEDURE AT THE USER FACILITY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22140029 ↗
  • MalfunctionFeb 20, 2025Overheating of Device; Overheating of Device

    THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21425691 ↗
  • MalfunctionJan 17, 2025Overheating of Device; Overheating of Device

    THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21186129 ↗
  • MalfunctionDec 10, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE MANUFACTURER SERVICE TECHNICIAN REPORTED THAT THE ATTACHMENT HOUSING WAS UNTHREADED FROM THE FRONT HOUSING AND MISSING WHILE IT WAS BEING SERVICED AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20891834 ↗
  • MalfunctionOct 24, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING TESTING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20523522 ↗
  • MalfunctionJul 19, 2024Overheating of Device; Overheating of Device

    THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19780480 ↗
  • MalfunctionJul 19, 2024Overheating of Device; Overheating of Device

    THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19780478 ↗
  • MalfunctionMar 12, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE TIP BROKE OFF DURING A PROCEDURE AT THE USER FACILITY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18884457 ↗
  • MalfunctionApr 27, 2023Overheating of Device; Overheating of Device

    THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16828711 ↗
  • MalfunctionApr 27, 2023Overheating of Device; Overheating of Device

    THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16828710 ↗
  • MalfunctionApr 20, 2023Overheating of Device; Overheating of Device

    THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16780628 ↗
  • MalfunctionApr 20, 2023Overheating of Device; Overheating of Device

    THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16778025 ↗
  • MalfunctionApr 20, 2023Overheating of Device; Overheating of Device

    THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16778015 ↗
  • MalfunctionApr 13, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE DISASSEMBLED DURING A PROCEDURE AT THE USER FACILITY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16732948 ↗
  • MalfunctionApr 6, 2023Overheating of Device; Overheating of Device

    THE USER FACILITY REPORTED THAT THE DEVICE OVERHEATED DURING A PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16695085 ↗
  • MalfunctionJul 31, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE TIP BROKE OFF AT THE USER FACILITY. THERE WAS NO KNOWN PATIENT INVOLVEMENT, NO KNOWN DELAY, NO KNOWN MEDICAL INTERVENTION, AND NO KNOWN ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15136941 ↗
  • MalfunctionJul 1, 2021Overheating of Device; Overheating of Device

    THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12099620 ↗
  • MalfunctionJul 1, 2021Overheating of Device; Overheating of Device

    THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12099619 ↗
  • MalfunctionJul 1, 2021Overheating of Device; Overheating of Device

    THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12099509 ↗
  • MalfunctionFeb 9, 2021Overheating of Device; Overheating of Device

    THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11304650 ↗
  • MalfunctionJul 22, 2019Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE DEVICE OVERHEATED DURING A PROCEDURE. THERE WAS NO PATIENT IMPACT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8813008 ↗
  • MalfunctionJul 22, 2019Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE DEVICE OVERHEATED DURING A PROCEDURE.  THERE WAS NO PATIENT IMPACT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8813003 ↗
  • MalfunctionApr 10, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    THE MANUFACTURER SERVICE TECHNICIAN REPORTED THAT THE DEVICE NOSE IS BROKEN OFF AND MISSING WHILE IT WAS BEING SERVICED AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 8501658 ↗