Accolade 6020-0537Recalls & reported problems
Howmedica Osteonics Corp. · 132 Degree Neck Angle Hip Stem · Class 2 · Implant
Quick summary
This Howmedica Accolade hip stem is an implanted part used in hip replacement. Reports mostly describe wear where the head connects to the stem, with the head coming loose, metal debris around the joint, pain and need for revision surgery. There are 23 reports that mention 23 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 23
- Deaths reported
- 0
- Injuries reported
- 23
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 04546540510570), received between Jun 20, 2018 and Mar 19, 2026 — classified as 23 injuries.
The most frequently reported specific device problem is “Degraded” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Degraded9
- Detachment of Device or Device Component5
- Insufficient Device Problem Information5
- Material Erosion4
- Corroded3
- Device Dislodged or Dislocated3
- Break2
- Fracture2
Patient problems
- Injury13
- Pain6
- Ambulation or Postural Difficulties4
- Inflammation4
- Reaction4
- Metal Related Pathology3
- Foreign Body Reaction2
- Insufficient Information2
Recent adverse event reports
The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.
InjuryMar 19, 2026Detachment of Device or Device Component
PATIENT HAD EMERGENCY REVISION SURGERY DUE TO A HEAD DISASSOCIATION IN HIS RIGHT HIP. ON (B)(6) 2025, THE PATIENT REPORTED STANDING UP FROM A COUCH WHEN HE EXPERIENCED SEVERE PAIN AND WAS UNABLE TO WALK. PATIENT IS SEEKING COMPENSATION.
Patient problems: Ambulation or Postural Difficulties; Joint Dislocation; Pain
MAUDE report 24636623 ↗InjuryOct 28, 2025Degraded; Degraded
PATIENT HAD HIP SURGERY 10 YEARS AGO. HE CONTACTED STRYKER TO INQUIRE WHO IS GOING TO PAY HIS HOSPITAL BILLS FOR HIS MRI & BLOODWORK OF WHICH HE HAS METALLOSIS. HE IS DISABLED & HAS MEDICARE. PATIENT STATED HE WAS WALKING PRIOR TO SURGERY AND NOW IS A QUADRAPLEGIC (BROKE HIS NECK MANY YEARS PRIOR). HE WAS TOLD HE NEED ANOTHER MRI AND REVISION SURGERY TO REPLACE CERAMIC HEAD & CUP. UNABLE TO OBTAIN AN ANSWER FROM THE SURGEON REGARDING BILL PAYMENT. HE WOULD LIKE TO KNOW WHAT ARE HIS ALTERNATIVES TO THIS SITUATION.
Patient problems: Metal Related Pathology; Pain
MAUDE report 23404112 ↗InjuryApr 8, 2025Corroded; Degraded; Detachment of Device or Device Component; Corroded; Degraded; Detachment of Device or Device Component
NO NEW INFORMATION.
Patient problems: Insufficient Information; Metal Related Pathology
MAUDE report 21795505 ↗InjuryFeb 6, 2025Degraded; Material Erosion; Detachment of Device or Device Component; Degraded; Material Erosion; Detachment of Device or Device Component
IT WAS REPORTED BY COUNSEL THAT CLAIMANT UNDERWENT RIGHT THA ON (B)(6) 2013 AND WAS IMPLANTED WITH AN LFIT ANATOMIC V40 FEMORAL HEAD AND #5 ACCOLADE TMZF STEM. ALLEGEDLY SHE DEVELOPED METALLOSIS ABOUT THE HIP AND HAS FRACTURED HER FEMORAL HEAD OFF OF THE TRUNNION. SHE WAS REVISED ON (B)(6) 2024.
Patient problems: Foreign Body Reaction; Metal Related Pathology
MAUDE report 21321413 ↗InjuryApr 16, 2024Fracture; Fracture
IT WAS REPORTED BY CLAIMANT'S COUNSEL THAT THE PATIENT UNDERWENT RIGHT THA ON (B)(6) 2009 AND WAS IMPLANTED WITH AN LFIT V40 COCR FEMORAL HEAD AND AN ACCOLADE TMZF HIP STEM. HE WAS REVISED ON (B)(6) 2021 DUE TO ALLEGED FRACTURED TRUNNION AND PARTIAL DISLOCATION.
Patient problems: Ambulation or Postural Difficulties
MAUDE report 19122656 ↗InjuryJul 18, 2022Degraded; Degraded
IT WAS REPORTED THROUGH CLAIMANT'S COUNSEL THAT ALLEGEDLY THE PATIENT WAS IMPLANTED WITH AN LFIT ANATOMIC COCR V40 FEMORAL HEAD ON HIS RIGHT HIP ON OR ABOUT (B)(6) 2011 AND WAS REVISED ON (B)(6) 2022. DURING THE REVISION THE SURGEON NOTED "¿ THERE WAS A SIGNIFICANT AMOUNT OF BLACK METALLOSIS FLUID ENCOUNTERED. AN EXTENSIVE DEBRIDEMENT OF ALL METAL DEBRIS WAS PERFORMED AND IT WAS FOUND THAT HIS TRUNNION WAS COMPLETELY DESTROYED."
Patient problems: Foreign Body Reaction; Insufficient Information
MAUDE report 15046515 ↗InjurySep 30, 2021Detachment of Device or Device Component; Device Dislodged or Dislocated; Detachment of Device or Device Component; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT'S RIGHT HIP WAS REVISED. THE PATIENT HAD FALLEN AND THE HEAD DISASSOCIATED FROM THE STEM. INTRAOPERATIVELY, THE SURGEON NOTICED THAT THERE WAS A LOT OF WEAR AROUND THE TRUNNION. REP NOTED THAT "THE TIP OF THE TRUNNION WAS GONE." REP HAS THE IMPLANTS FOR RETURN.
Patient problems: Ambulation or Postural Difficulties; Fall
MAUDE report 12554364 ↗InjuryMay 19, 2021Material Erosion; Detachment of Device or Device Component; Insufficient Device Problem Information; Material Erosion; Detachment of Device or Device Component; Insufficient Device Problem Information
PATIENT HAD HIP REPLACEMENT INSERTED IN 2006. WAS DOING FINE UNTIL 2 WEEKS AGO AND CONSULTED WITH THE DR. WHO DID REVISION HIP SURGERY YESTERDAY UPDATE PER PER FORM: " PATIENT SAID HE FELT ACUTE CHANGE ON HIS HIP PER DESCRIPTION FROM OP-NOTES." *UPDATE ON (B)(6) 2021 PER CLINICIAN REVIEW: "[...] THE EVENT A REVISION SURGERY WAS CONFIRMED VIA THE OP NOTE. THE REVISION WAS FOR HEAD-STEM DISENGAGEMENT. TRUNNION WEAR AND METALLOSIS NOTED AT REVISION. [...]
Patient problems: Failure of Implant; Pain
MAUDE report 11847864 ↗InjuryJan 5, 2021Material Erosion; Material Erosion
IT WAS REPORTED THAT THE PATIENT'S RIGHT HIP WAS REVISED DUE TO TRUNNIONS.
Patient problems: Inflammation; No Information
MAUDE report 11118306 ↗InjuryNov 10, 2020Material Erosion; Material Erosion
IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED DUE TO SEVERE WEAR OF THE TRUNNION AND NOTCHING ON THE MEDIAL SIDE OF THE STEM CAUSED BY THE SHELL. INTRA-OPERATIVELY, METALLOSIS WAS NOTED. THE STEM, HEAD AND LINER WERE REVISED. REP CONFIRMED THERE ARE NO ALLEGATIONS AGAINST THE REVISED LINER.
Patient problems: Inflammation
MAUDE report 10819615 ↗InjuryMay 18, 2020Fracture; Device Dislodged or Dislocated; Fracture; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT'S RIGHT HIP WAS REVISED. INTRA-OPERATIVE OBSERVATIONS REPORTED TO THE REP: A MODERATE AMOUNT OF BLACK TISSUE, THE TRUNNION WAS WORN AND CRACKED OFF. THE LINER WAS DISCOLORED. THE PORTION OF THE STEM INSIDE THE HEAD WAS REMOVED WITH DIFFICULTY.
Patient problems: Inflammation; Injury
MAUDE report 10065859 ↗InjuryFeb 6, 2020Corroded; Insufficient Device Problem Information; Corroded; Insufficient Device Problem Information
InjuryOct 16, 2019Material Deformation; Material Deformation
THE CUSTOMER REPORTED THAT THE TMZF STEM HAD MAJOR TRUNNIONOSIS AND THE PATIENT REQUIRED REVISION SURGERY.
Patient problems: Inflammation; Injury; Reaction
MAUDE report 9197366 ↗InjurySep 3, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED DUE TO DISASSOCIATION OF THE HEAD FROM THE STEM. INTRA-OPERATIVELY, AN EXTREME AMOUNT OF TRUNNION WEAR WAS NOTED, AND DAMAGE TO THE RIM OF THE LINER FROM THE STEM. THE SHELL WAS WELL-FIXED AND WELL- POSITIONED AND WAS NOT REVISED. THE PATIENT WAS REVISED TO A RESTORATION MODULAR STEM CONSTRUCT, CERAMIC HEAD, AND MDM/ADM LINER CONSTRUCT. PRIMARY IMPLANTATION OCCURRED WITHIN THE REVISION FACILITY'S HOSPITAL NETWORK BUT THE EXACT HOSPITAL IS UNKNOWN.
Patient problems: Injury
MAUDE report 8955492 ↗InjuryAug 23, 2019Break; Break
InjuryJun 20, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT PATIENT'S RIGHT HIP WAS REVISED DUE TO DISASSOCIATION OF THE HEAD FROM THE STEM. INTRA-OPERATIVELY, AN EXTREME AMOUNT OF BLACK TISSUE AND DISCOLORATION OF THE LINER WERE NOTED. PATIENT WAS REVISED TO A COMPETITOR STEM AND HEAD WITH A STRYKER LINER. THE SHELL WAS NOT REVISED.
Patient problems: Injury; Reaction
MAUDE report 8715883 ↗InjuryApr 16, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED AFTER PATIENT COMPLAINT OF PAIN. INTRA-OPERATIVELY, EXCESSIVE BLACK TISSUE, SIGNIFICANT TRUNNION WEAR, AND CORROSION WERE NOTED. PATIENT'S ENTIRE HIP CONSTRUCT WAS REMOVED AND A SPACER WAS PLACED AS THE SURGEON ALSO SUSPECTED INFECTION (INFECTING MICROORGANISM NOT REPORTED TO REP).
Patient problems: Injury; Pain; Unspecified Infection
MAUDE report 8517932 ↗InjuryFeb 21, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
HIP REVISION OF ACCOLADE TMZF STEM WITH DISASSOCIATED HEAD.
Patient problems: Injury; Joint Dislocation
MAUDE report 8358185 ↗InjuryOct 22, 2018Degraded; Degraded
IT WAS REPORTED THAT PATIENT'S LEFT HIP WAS REVISED. AS REPORTED BY REP: "THE FEMORAL HEAD AND FEMORAL STEM NECK TRUNION IS WORN. THE FEMORAL HEAD DISASSOCIATED".
Patient problems: Injury; Reaction
MAUDE report 7991578 ↗InjuryOct 11, 2018Degraded; Degraded
IT WAS REPORTED THAT PATIENT'S HIP WAS REVISED DUE TO SUSPECTED TRUNNIONOSIS. REP WAS NOT PRESENT FOR PROCEDURE. NO PATIENT COMPLAINTS, PRE-OPERATIVE DIAGNOSES, REASON FOR SUSPICION OF TRUNNIONOSIS, OR INTRA-OPERATIVE FINDINGS WERE REPORTED TO THE REP. PRIMARY AND REVISION OPERATIVE NOTES WERE PROVIDED. PATIENT'S STEM, FEMORAL HEAD, LINER, AND SHELL WERE REVISED. REASON FOR SHELL REVISION WAS NOT REPORTED TO THE REP.
Patient problems: Injury
MAUDE report 7958740 ↗InjuryOct 8, 2018Degraded; Degraded
IT WAS REPORTED THAT THE PATIENT REQUIRED A HIP REVISION. UPDATE; PER RECEIVED IMAGES, IT APPEARS THE TRUNNION IS WORN AND HEAD DISLOCATED.
Patient problems: Injury
MAUDE report 7944551 ↗InjurySep 11, 2018Break; Degraded; Material Deformation; Break; Degraded; Material Deformation
AS REPORTED: "PATIENT HAS PAIN AND SHORTENING OF HIP. COMPLAINS HIP GAVE OUT". PRE-REVISION X-RAYS REVEAL THE HEAD DISASSOCIATED FROM THE STEM AND SHOW AT LEAST THE HEAD ROTATED WITHIN THE SHELL OR LINER. EXPLANT PICTURES SHOW THAT THE TRUNNION WAS WORN AT THE TIP, AND BROKEN OFF OF THE STEM. REVISION IMPLANTS ARE A 44MM G POLY LINER, 44 +0 LFIT HEAD, AND A SIZE 6 35 X 111MM STEM.
Patient problems: Ambulation or Postural Difficulties; Injury; Pain
MAUDE report 7864592 ↗InjuryJun 20, 2018Corroded; Degraded; Disassembly; Corroded; Degraded; Disassembly
IT WAS REPORTED THAT ON (B)(6) 2018, DR. (B)(6) PERFORMED A REVISION OF AN ACCOLADE STEM WITH A DISLOCATED FEMORAL HEAD COMPONENT AS VISIBLE BY PREOP X-RAY. UPDATE PER MED REVIEW: CUP MALPOSITION IN NEAR ABSENT ANTEVERSION HAS CONTRIBUTED TO AN OVERLOAD CONDITION IN THE ARTHROPLASTY, ESPECIALLY NEAR THE TRUNNION SECTION CAUSING SECONDARY CORROSION BETWEEN FEMORAL HEAD AND STEM TAPER LEADING TO PROGRESSIVE METAL SUBSTANCE LOSS WITH METALLOSIS IN THE JOINT SPACE, LOSS OF TAPER LOCK AND ULTIMATELY DISASSOCIATION BETWEEN STEM TRUNNION AND FEMORAL HEAD REQUIRING REVISION. ADDITIONALLY, THERE IS OSTEOLYSIS AROUND THE PROXIMAL STEM SECTION ALTHOUGH THE STEM REMAINS WELL FIXED IN THE MORE DISTAL PARTS. ALSO AROUND THE CUP IS OSTEOLYSIS EVIDENT, ESPECIALLY AROUND THE SUPPLEMENTAL SCREW FIXATION ALTHOUGH ALSO THE CUP STILL APPEARS STABLE IN THE BONE NEAR THE PERIPHERY.
Patient problems: Injury; Osteolysis; Reaction
MAUDE report 7620182 ↗
Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices
- NA 1365-51-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented753 reports
- NA 1365-50-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented412 reports
- ARTICUL/EZE 1365-21-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall280 reports
- ARTICUL/EZE 1365-11-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death262 reports
- LFIT 6260-9-236Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented256 reports
- PINNACLE GRIPTION 1217-32-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 deaths239 reports
- ELEOS™ 25002111EOnkos Surgical, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 deaths230 reports
- Novation 130-36-52Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented2 recalls124 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.