Skip to content

XIA 4 5 48133230Recalls & reported problems

Stryker Corporation · 30-36 mm POLYAXIAL CROSS CONNECTOR · Class 2 · Implant

All models of XIA 4 5 (32)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04546540515087), received between Mar 17, 2016 and Aug 5, 2016 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1
  • Device Operates Differently Than Expected1
  • Material Deformation1
  • Material Integrity Problem1
  • Mechanical Jam1

Patient problems

  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 5, 2016Device Operates Differently Than Expected; Material Deformation; Material Integrity Problem; Material Deformation; Material Integrity Problem

    IT WAS REPORTED THAT; DURING SMALL SURGERY, THE SURGEON INSERTED THE SCREWS TO L5(CBT)-S. AND THE SURGEON TRIED TO USE THE CROSSLINK CONNECTOR(MEDIUM). BUT THE SURGEON COULD NOT ASSEMBLED THE CROSSLINK CONNECTOR. THEREFORE THE SURGEON TRIED TO REMOVE THE CROSSLINK CONNECTOR. WHEN THE SURGEON LOOSENED THE CENTER SCREW OF THE CONNECTOR, THE SCREW WAS DEFORMED. THEREFORE THE SURGEON REMOVED THE CROSSLINK CONNECTOR WITH THE ROD. AFTERWARDS, THE SURGEON USED THE LARGE CROSSLINK CONNECTOR. BUT, THE SURGEON COULD NOT ASSEMBLED THE CROSSLINK CONNECTOR. THEREFORE THE SURGEON NOT USED THE CROSSLINK CONNECTOR.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5851670 ↗
  • MalfunctionMar 17, 2016Mechanical Jam; Break; Mechanical Jam

    IT WAS REPORTED THAT DURING THE SURGERY, THE SURGEON USED A CROSS CONNECTOR. THE SURGEON FORCED A DRIVER TIP INTO THE SCREW OF THE CROSS CONNECTOR USING A DRIVER, THE SCREW OF THE CROSS CONNECTOR BROKE. THE SURGEON CHANGED THE CROSS CONNECTOR TO ANOTHER SAME SIZE PRODUCT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5509219 ↗

Other ORTHOSIS, SPINAL PEDICLE FIXATION devices

All ORTHOSIS, SPINAL PEDICLE FIXATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page