Skip to content

REFLEX HYBRID 48651472Recalls & reported problems

Stryker Corporation · FOUR-LEVEL PLATE · Class 2 · Implant

All models of REFLEX HYBRID (34)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04546540535016), received between Dec 13, 2017 and May 7, 2021 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Device Operates Differently Than Expected1
  • Fracture1
  • Material Deformation1
  • Material Protrusion/Extrusion1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 7, 2021Fracture; Material Deformation; Fracture; Material Deformation

    A COMPANY REPRESENTATIVE REPORTED THAT INSERTION OF A 4.0 MM REFLEX SCREW AT LEFT C4 WOULD NOT FINAL LOCK. WHILE REMOVING THE SCREW, THE SECURING MECHANISM OF A FOUR-LEVEL REFLEX HYBRID PLATE FRACTURED. ALL SCREWS AND THE PLATE WERE REMOVED AND REPLACED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO ADVERSE CONSEQUENCE TO THE PATIENT. THIS REPORT CAPTURES THE FOUR-LEVEL REFLEX-HYBRID PLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11791137 ↗
  • MalfunctionDec 13, 2017Device Operates Differently Than Expected; Material Protrusion/Extrusion; Device Operates Differently Than Expected

    IT WAS REPORTED THAT; AN ACDF C3-7 WAS PERFORMED. THE SURGEON FINAL TIGHTENING THE RIGHT C5 SCREW AND THE LOCKING RING PUSHED OUT AND STUCK OUT FROM THE PLATE. SURGEON REMOVED THE SCREW AND PULLED THE LOCKING RING FROM THE PLATE OUT OF THE SURGICAL SITE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7112611 ↗

Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices

All APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices →

Sources and updates