This Stryker XIA 3 spinal fixation blocker has 193 reports, 100 injuries and no deaths. Many reports describe the blocker migrating out of place, fracturing, or loosening, and several describe revision surgery to remove or replace it, sometimes after infection. Reports rose from 4 in the prior 12 months to 15 in the most recent 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
193
Deaths reported
0
Injuries reported
100
Recalls naming it
0

Summary

The FDA's MAUDE database contains 193 adverse event reports that list this exact device (DI 04546540560193), received between May 17, 2016 and Jun 8, 2026 — classified as 100 injuries, 93 malfunctions.

The most frequently reported specific device problem is “Migration” (81 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 9 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 5 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration81
  • Fracture37
  • Loose or Intermittent Connection32
  • Mechanical Problem31
  • Device Dislodged or Dislocated22
  • Migration or Expulsion of Device22
  • Appropriate Device Problem Term/Code Not Available15
  • No Apparent Adverse Event15

Patient problems

  • Injury72
  • No Clinical Signs, Symptoms or Conditions49
  • No Known Impact Or Consequence To Patient34
  • Pain34
  • No Consequences Or Impact To Patient23
  • No Code Available16
  • Insufficient Information12
  • Unspecified Infection11

Recent adverse event reports

The 25 most recent of 193 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 8, 2026Migration

    A HEALTH PROFESSIONAL REPORTED THAT A PATIENT WAS REVISED APPROXIMATELY THREE WEEKS POST-OPERATIVELY TO ADDRESS A MIGRATED XIA 3 BLOCKER. THE PATIENT REPORTED EXPERIENCING A "CLICKING" SOUND DURING MOVEMENT AND X-RAY IMAGING REVEALED THE MIGRATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25437549 ↗
  • InjuryMar 25, 2026Insufficient Device Problem Information

    A HEALTH PROFESSIONAL REPORTED THAT A PATIENT IMPLANTED WITH FOUR ES2 INTEGRATED BLADE SCREWS, FOUR XIA BLOCKERS, AND TWO MANTIS RODS EXPERIENCED AN INFECTION POST-OPERATIVELY. IT WAS FURTHER REPORTED THAT THE PATIENT WAS REVISED AND THE DEVICES WERE EXPLANTED. THIS REPORT CAPTURES THE FOURTH OF FOUR XIA BLOCKERS.

    Patient problems: Unspecified Infection

    MAUDE report 24691088 ↗
  • InjuryMar 25, 2026Insufficient Device Problem Information

    A HEALTH PROFESSIONAL REPORTED THAT A PATIENT IMPLANTED WITH FOUR ES2 INTEGRATED BLADE SCREWS, FOUR XIA BLOCKERS, AND TWO MANTIS RODS EXPERIENCED AN INFECTION POST-OPERATIVELY. IT WAS FURTHER REPORTED THAT THE PATIENT WAS REVISED AND THE DEVICES WERE EXPLANTED. THIS REPORT CAPTURES THE THIRD OF FOUR XIA BLOCKERS.

    Patient problems: Unspecified Infection

    MAUDE report 24691087 ↗
  • InjuryMar 25, 2026Insufficient Device Problem Information

    A HEALTH PROFESSIONAL REPORTED THAT A PATIENT IMPLANTED WITH FOUR ES2 INTEGRATED BLADE SCREWS, FOUR XIA BLOCKERS, AND TWO MANTIS RODS EXPERIENCED AN INFECTION POST-OPERATIVELY. IT WAS FURTHER REPORTED THAT THE PATIENT WAS REVISED AND THE DEVICES WERE EXPLANTED. THIS REPORT CAPTURES THE FIRST OF FOUR XIA BLOCKERS.

    Patient problems: Unspecified Infection

    MAUDE report 24691086 ↗
  • InjuryMar 25, 2026Insufficient Device Problem Information

    A HEALTH PROFESSIONAL REPORTED THAT A PATIENT IMPLANTED WITH FOUR ES2 INTEGRATED BLADE SCREWS, FOUR XIA BLOCKERS, AND TWO MANTIS RODS EXPERIENCED AN INFECTION POST-OPERATIVELY. IT WAS FURTHER REPORTED THAT THE PATIENT WAS REVISED AND THE DEVICES WERE EXPLANTED. THIS REPORT CAPTURES THE SECOND OF FOUR XIA BLOCKERS.

    Patient problems: Unspecified Infection

    MAUDE report 24691085 ↗
  • InjuryMar 9, 2026Migration

    IT WAS REPORTED THAT A PATIENT UNDERWENT REVISION SURGERY TO ADDRESS A "MECHANICAL FAILURE" BETWEEN AN ES2 ROD, AN ES2 INTEGRATED BLADE SCREW AND A XIA BLOCKER FIVE MONTHS POST-OPERATIVELY. UPON INSPECTION OF AN ADDITIONAL RETURNED XIA 3 BLOCKER, IT WAS OBSERVED THAT THIS BLOCKER ALSO SHOWS SIGNS OF MIGRATION. THIS REPORT CAPTURES THE SECOND RETURNED XIA BLOCKER.

    Patient problems: Insufficient Information

    MAUDE report 24544813 ↗
  • InjuryNov 25, 2025Migration; Migration

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23647502 ↗
  • InjuryNov 6, 2025Migration; Migration

    IT WAS REPORTED THAT A PATIENT UNDERWENT REVISION SURGERY TO ADDRESS EIGHT MIGRATED MANTIS REDUX BLOCKERS.

    Patient problems: Insufficient Information

    MAUDE report 23487260 ↗
  • InjuryNov 6, 2025Migration; Migration

    IT WAS REPORTED THAT A PATIENT UNDERWENT REVISION SURGERY TO ADDRESS EIGHT MIGRATED MANTIS REDUX BLOCKERS.

    Patient problems: Insufficient Information

    MAUDE report 23487250 ↗
  • InjuryNov 6, 2025Migration; Migration

    IT WAS REPORTED THAT A PATIENT UNDERWENT REVISION SURGERY TO ADDRESS EIGHT MIGRATED MANTIS REDUX BLOCKERS.

    Patient problems: Insufficient Information

    MAUDE report 23487247 ↗
  • InjuryNov 6, 2025Migration; Migration

    IT WAS REPORTED THAT A PATIENT UNDERWENT REVISION SURGERY TO ADDRESS EIGHT MIGRATED MANTIS REDUX BLOCKERS. THIS REPORT CAPTURES THE FIRST OF EIGHT BLOCKERS.

    Patient problems: Insufficient Information

    MAUDE report 23487238 ↗
  • InjuryNov 6, 2025Migration; Migration

    IT WAS REPORTED THAT A PATIENT UNDERWENT REVISION SURGERY TO ADDRESS EIGHT MIGRATED MANTIS REDUX BLOCKERS.

    Patient problems: Insufficient Information

    MAUDE report 23487236 ↗
  • InjuryNov 6, 2025Migration; Migration

    IT WAS REPORTED THAT A PATIENT UNDERWENT REVISION SURGERY TO ADDRESS EIGHT MIGRATED MANTIS REDUX BLOCKERS.

    Patient problems: Insufficient Information

    MAUDE report 23487222 ↗
  • InjuryNov 6, 2025Migration; Migration

    IT WAS REPORTED THAT A PATIENT UNDERWENT REVISION SURGERY TO ADDRESS EIGHT MIGRATED MANTIS REDUX BLOCKERS.

    Patient problems: Insufficient Information

    MAUDE report 23487089 ↗
  • InjuryNov 6, 2025Migration; Migration

    IT WAS REPORTED THAT A PATIENT UNDERWENT REVISION SURGERY TO ADDRESS EIGHT MIGRATED MANTIS REDUX BLOCKERS.

    Patient problems: Insufficient Information

    MAUDE report 23487079 ↗
  • InjurySep 25, 2025Insufficient Device Problem Information; Migration; Insufficient Device Problem Information; Migration

    UPON INSPECTION OF AN ADDITIONAL RETURNED XIA 3 BLOCKER, IT WAS OBSERVED THAT THE BLOCKER SHOWED SIGNS OF MIGRATION. THIS REPORT CAPTURES THE FIRST OF TWO XIA BLOCKERS.

    Patient problems: Insufficient Information

    MAUDE report 23145098 ↗
  • InjuryApr 3, 2025Migration; Migration

    A COMPANY REPRESENTATIVE REPORTED THAT IMAGING TAKEN APPROXIMATELY SIX MONTHS POST-OPERATIVELY REVEALED THAT THREE XIA 3 TITANIUM BLOCKERS MIGRATED. THE PATIENT THEN UNDERWENT REVISION SURGERY APPROXIMATELY EIGHT MONTHS POST-OPERATIVELY. THIS RECORD CAPTURES THE SECOND OF THREE XIA BLOCKERS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21764092 ↗
  • InjuryApr 3, 2025Migration; Migration

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21764087 ↗
  • InjuryFeb 5, 2025Migration; Migration

    NO NEW INFORMATION.

    Patient problems: Insufficient Information

    MAUDE report 21313375 ↗
  • InjuryOct 3, 2024Migration; Migration

    IT WAS REPORTED THAT A PATIENT WAS REVISED TO ADDRESS A MIGRATED XIA BLOCKER APPROXIMATELY EIGHT MONTHS AFTER IMPLANTATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20363984 ↗
  • InjuryJun 28, 2023Migration; Migration

    A PATIENT WAS REVISED TO ADDRESS TWO MIGRATED XIA BLOCKERS. THE PATIENT REPORTED EXPERIENCING DISCOMFORT PRIOR TO REVISION. THIS REPORT CAPTURES THE SECOND OF TWO MIGRATED XIA BLOCKERS.

    Patient problems: Pain

    MAUDE report 17219472 ↗
  • InjuryJun 28, 2023Migration; Migration

    A PATIENT WAS REVISED TO ADDRESS TWO MIGRATED XIA BLOCKERS. THE PATIENT REPORTED EXPERIENCING DISCOMFORT PRIOR TO REVISION. THIS REPORT CAPTURES THE FIRST OF TWO MIGRATED XIA BLOCKERS.

    Patient problems: Pain

    MAUDE report 17219305 ↗
  • InjuryMay 10, 2023Migration; Migration

    IT WAS REPORTED THAT 2 MONTHS AFTER IMPLANTATION, A RIGHT-SIDE XIA 3 BLOCKER AND A LEFT-SIDE XIA 3 BLOCKER MIGRATED, AND A LEFT-SIDE XIA 3 POLYAXIAL SCREW FRACTURED. THE PATIENT WAS REVISED APPROXIMATELY 1 YEAR LATER TO ADDRESS THE MIGRATED BLOCKERS AND FRACTURED SCREW. THE PATIENT EXPERIENCED MILDLY INCREASING LOW BACK AND PELVIC PAIN. THIS REPORT IS FOR THE LEFT-SIDE XIA 3 BLOCKER.

    Patient problems: Pain

    MAUDE report 16907214 ↗
  • InjuryMay 10, 2023Migration; Migration

    IT WAS REPORTED THAT 2 MONTHS AFTER IMPLANTATION, A RIGHT-SIDE XIA 3 BLOCKER AND A LEFT-SIDE XIA 3 BLOCKER MIGRATED, AND A LEFT-SIDE XIA 3 POLYAXIAL SCREW FRACTURED. THE PATIENT WAS REVISED APPROXIMATELY 1 YEAR LATER TO ADDRESS THE MIGRATED BLOCKERS AND FRACTURED SCREW. THE PATIENT EXPERIENCED MILDLY INCREASING LOW BACK AND PELVIC PAIN. THIS REPORT IS FOR THE RIGHT-SIDE XIA 3 BLOCKER.

    Patient problems: Pain

    MAUDE report 16907205 ↗
  • MalfunctionFeb 3, 2023Fracture; Fracture

    A COMPANY REPRESENTATIVE REPORTED THAT A XIA BLOCKER CRACKED INTRA-OPERATIVELY WHILE FINAL TIGHTENING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16299958 ↗

Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices

All APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.