XIA 3 482317540Recalls & reported problems
Stryker Corporation · POLYAXIAL SCREW · Class 2 · Implant
All models of XIA 3 (93)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04546540562340), received between Mar 31, 2016 and May 31, 2022 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fracture3
- Material Separation2
- Connection Problem1
- Detachment of Device or Device Component1
- Disconnection1
- Material Integrity Problem1
- Mechanical Problem1
- Unintended Movement1
Patient problems
- No Known Impact Or Consequence To Patient3
- Pain2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 31, 2022Material Separation; Detachment of Device or Device Component; Material Separation; Detachment of Device or Device Component
IT WAS REPORTED THAT THE TULIP OF A XIA 3 POLYAXIAL SCREW BROKE OFF FROM THE SCREW SHAFT POST-OPERATIVELY. REVISION SURGERY HAS NOT BEEN SCHEDULED OR PERFORMED AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14547556 ↗InjuryAug 27, 2020Fracture; Fracture
A PHYSICIAN REPORTED A PATIENT HAS BEEN EXPERIENCING BACK PAIN AND IMAGING INDICATES TWO FRACTURED XIA 3 POLYAXIAL SCREWS AT S1. REVISION SURGERY HAS BEEN PERFORMED. THIS RECORD WILL CAPTURE THE SECOND OF TWO SCREWS.
Patient problems: Pain
MAUDE report 10463278 ↗InjuryAug 27, 2020Fracture; Fracture
A PHYSICIAN REPORTED A PATIENT HAS BEEN EXPERIENCING BACK PAIN AND IMAGING INDICATES TWO FRACTURED XIA 3 POLYAXIAL SCREWS AT S1. REVISION SURGERY HAS OCCURRED. THIS RECORD WILL CAPTURE THE FIRST OF TWO SCREWS.
Patient problems: Pain
MAUDE report 10463277 ↗MalfunctionFeb 24, 2017Disconnection; Fracture; Material Integrity Problem; Fracture; Material Integrity Problem
IT WAS REPORTED THAT; TULIP HEAD DISENGAGED FROM THE SCREW.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6358170 ↗InjuryNov 23, 2016Mechanical Problem; Unintended Movement; Mechanical Problem; Connection Problem; Unintended Movement
IT WAS REPORTED THAT; THE PROCEDURE WAS PLIF. ATTEMPT TO TIGHTEN THE L5 LEFT SCREW BY TORQUE WRENCH DURING FINAL FASTING. IT COULDN'T CONNECT THE ANTI-TORQUE-KEY ON SCREW HEAD DUE TO PRESSED BY OUTER MUSCLE. ATTEMPTED TO START A FINAL FASTEN AGAIN FOR L4 LEFT SCREW AFTER CONNECTED THE ANTI-TORQUE-KEY WITH L4 LEFT SCREW HEAD. BUT THERE WAS CUT OUT THE L5 LEFT SCREW UNDER ABOUT 12N. REMOVED THE BLOCKER AND THE ROD BOTH. INSERTED THE L5 LEFT SCREW TO PROGRESS IN A NEW DIRECTION, HOWEVER THERE WAS FOUND A MOVEMENT OF SCREW AGAIN. THEN CREATED ANOTHER NEW SCREW HOLE AND IMPLANTED OTHER SCREW WITH NO COMPRESSION BUT IT REMAINED A SLIGHT MOTION ON SCREW. COMPLETE THE PROCEDURE. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCE REPORTED AT THIS TIME.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6125368 ↗MalfunctionMar 31, 2016Material Separation
IT WAS REPORTED THAT; THE TULIP HEAD SEPARATED FROM THE SCREW WHILE REDUCING A SPONDY WITH THE CORKSCREW PERSUADER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5537687 ↗
Other ORTHOSIS, SPINAL PEDICLE FIXATION devices
- NA 04.614.508Synthes GmbH · ORTHOSIS, SPINAL PEDICLE FIXATION277 reports
- CREO 1134.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION112 reports
- CREO 1119.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION80 reports
- VIPER 186731745Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION43 reports
- CD HORIZON® Spinal System 1553201090Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION38 reports
- EXPEDIUM 179902000Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION37 reports
- CD HORIZON® Spinal System 7540220Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION36 reports
- EXPEDIUM VERSE 199721640Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION35 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page