XIA 3 482318540Recalls & reported problems
Stryker Corporation · POLYAXIAL SCREW · Class 2 · Implant
All models of XIA 3 (93)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04546540598585), received between Jun 30, 2016 and Jun 28, 2022 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Material Separation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Separation2
- Device Slipped1
- Migration1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 28, 2022Material Separation; Material Separation
IT WAS REPORTED THAT THE TULIP OF A XIA 3 TITANIUM POLYAXIAL SCREW DISENGAGED DURING IMPLANTATION. THE SHAFT OF THE SCREW REMAINS IN THE PATIENT'S BONE AND THE PROCEDURE WAS COMPLETED FOLLOWING A FIVE-MINUTE SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14847976 ↗InjuryMar 10, 2021Material Separation; Migration; Material Separation; Migration
IT WAS REPORTED THAT THE TULIP OF A XIA 3 TITANIUM POLYAXIAL SCREW AT S1 DISENGAGED 2 MONTHS POST-OPERATIVELY. UPON REVISION, THE SURGEON FOUND THE TULIP OF THE SCREW TO HAVE BROKEN OFF FROM THE SCREW THREADS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11455195 ↗MalfunctionJun 30, 2016Device Slipped
IT WAS REPORTED THAT; TULIP POPPED OFF SCREW WHEN IT WAS FINAL TIGHTENED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5762244 ↗
Other ORTHOSIS, SPINAL PEDICLE FIXATION devices
- NA 04.614.508Synthes GmbH · ORTHOSIS, SPINAL PEDICLE FIXATION277 reports
- CREO 1134.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION112 reports
- CREO 1119.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION80 reports
- VIPER 186731745Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION43 reports
- CD HORIZON® Spinal System 1553201090Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION38 reports
- EXPEDIUM 179902000Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION37 reports
- CD HORIZON® Spinal System 7540220Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION36 reports
- EXPEDIUM VERSE 199721640Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION35 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.