Skip to content

RADIUS 486613602Recalls & reported problems

Stryker Corporation · VITALLIUM SPINAL ROD, WITH HEX · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04546540604859), received between Mar 7, 2017 and Apr 5, 2024 — classified as 4 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture6
  • Break2
  • Bent1
  • Migration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient2
  • Failure of Implant1
  • Injury1
  • Pain1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 5, 2024Fracture; Fracture

    IT WAS REPORTED THAT APPROXIMATELY 8 YEARS POST-OPERATIVELY, IMAGING REVEALED A XIA POLYAXIAL SCREW TIP FRACTURE AND THE FRACTURE OF TWO RADIUS VITALLIUM RODS. THIS REPORT CAPTURES THE FIRST OF TWO RADIUS VITALLIUM RODS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19056054 ↗
  • InjuryApr 5, 2024Fracture; Fracture

    IT WAS REPORTED THAT APPROXIMATELY 8 YEARS POST-OPERATIVELY, IMAGING REVEALED A XIA POLYAXIAL SCREW TIP FRACTURE AND THE FRACTURE OF TWO RADIUS VITALLIUM RODS. THIS REPORT CAPTURES THE SECOND OF TWO RADIUS VITALLIUM RODS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19056052 ↗
  • InjuryMay 3, 2023Fracture; Migration; Fracture; Migration

    A PATIENT REPORTED A FRACTURED RADIUS VITALLIUM ROD ON THE RIGHT SIDE OF THEIR CONSTRUCT. IMAGING SHOWS THE FRACTURED DEVICE HAS MIGRATED AND PATIENT REPORTS THAT THEY ARE AMBULATORY BUT IN PAIN. REVISION SURGERY IS NOT CURRENTLY PLANNED.

    Patient problems: Pain

    MAUDE report 16862848 ↗
  • InjuryFeb 25, 2019Break; Fracture; Fracture

    IT WAS REPORTED THAT THE ROD WAS BROKEN POST- OPERATIVELY. PATIENT UNDERWENT REVISION SURGERY FOR ROD BREAKAGE AND POSSIBLE NON-UNION.

    Patient problems: Failure of Implant; Injury

    MAUDE report 8366832 ↗
  • MalfunctionApr 10, 2018Fracture; Break; Fracture

    IT WAS REPORTED THAT DURING A REVISION SURGERY FOR T4- ILIAC, THE ROD BROKE WHEN THE SURGEON WAS BENDING IT OVER THE PATIENT. THE SURGEON USED ANOTHER ROD TO COMPLETE THE SURGERY. THERE ARE NO REPORTS OF ADVERSE HEALTH CONSEQUENCES TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7419352 ↗
  • MalfunctionMar 7, 2017Bent; Fracture; Fracture

    IT WAS REPORTED THAT; DURING SURGERY, THE ROD WAS CUT ON THE BACK TABLE WITH TOP CUTTER. WHEN THE ROD WAS BENT ON THE MAYO WITH FRENCH BENDER, IT SNAPPED IN HALF. ANOTHER ROD WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6384299 ↗

Other ORTHOSIS, SPINAL PEDICLE FIXATION devices

All ORTHOSIS, SPINAL PEDICLE FIXATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page