Restoration 1235-2-502Recalls & reported problems
Howmedica Osteonics Corp. · ADM- Cup w/HA · Class 2 · Implant
All models of Restoration (86)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04546540638731), received between Jul 1, 2019 and Apr 10, 2020 — classified as 2 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated1
- Positioning Problem1
- Use of Device Problem1
Patient problems
- Injury2
- Ambulation or Postural Difficulties1
- Joint Dislocation1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 10, 2020Use of Device Problem; Positioning Problem; Use of Device Problem; Positioning Problem
STEM SUBSIDENCE OF 13 MM OF SECUR-FIT ADVANCED STEM FOR CLINICAL STUDY PATIENT. UPDATE 15/JANUARY/2020 WG: AS REPORTED: "UPON REVIEW OF THE SUBJECT'S SIX-WEEK POSTOPERATIVE X-RAYS, IT WAS NOTED THAT THE STEM MOVED 13MM. THE EVENT WAS NOTED AS FEMORAL COMPONENT SUBSIDENCE, AND CONSIDERED BY THE STUDY INVESTIGATOR TO BE DEVICE RELATED AND SERIOUS. THIS SUBSIDENCE RESULTED IN A REVISION OF THE FEMORAL BEARING HEAD ON (B)(6) 2014, AND WAS CONSIDERED RESOLVED AT THAT TIME." LEFT HIP. UPDATE 13 MARCH 2020 BASED ON MEDICAL REVIEW: "THE X-RAY COPIES OF THE POSTOPERATIVE FILMS FROM THE PRIMARY SURGERY SHOWED A SUBSIDED STEM AND A SOMEWHAT OVERSIZED AND NOT FULLY SEATED CUP. THE STEM LOOKED UNDERSIZED. THE SUBSIDENCE LED TO LIMB LENGTH DISCREPANCY AND ULTIMATELY TO FEMORAL HEAD REVISION TO IMPROVE THE LENGTH DISCREPANCY.
Patient problems: Ambulation or Postural Difficulties; Injury
MAUDE report 9948399 ↗InjuryJul 1, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated
AS REPORTED: "LEFT THA REVISION DUE TO DISLOCATION. PATIENT EXPERIENCED DISLOCATION. DURING THE ATTEMPT TO REDUCE HER HIP, THE FORCE BEING APPLIED PRIED THE ADM INSERT OFF OF THE 28MM HEAD". SPOKE TO REP. THE RESTORATION ADM INSERT DISASSOCIATED OUT OF THE RESTORATION ADM SHELL. PATIENT WENT TO ER WHERE A CLOSED REDUCTION WAS ATTEMPTED. DURING THE REDUCTION THE HEAD DISLOCATED OUT OF THE LINER AND THE PATIENT WAS REVISED.
Patient problems: Injury; Joint Dislocation
MAUDE report 8750668 ↗
Other Prosthesis, hip, hemi-, femoral, metal devices
- UHR UH1-48-28Howmedica Osteonics Corp. · Prosthesis, hip, hemi-, femoral, metal2 recalls3 reports
- "UNITED" USTAR II Knee System 2515-3111United Orthopedic Corp. · Prosthesis, Hip, Hemi-, Femoral, Metal2 recalls0 reports
- "UNITED" USTAR II Knee System 2515-3213United Orthopedic Corp. · Prosthesis, Hip, Hemi-, Femoral, Metal2 recalls0 reports
- STAGEONE 431207Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal2 recalls0 reports
- "UNITED" USTAR II Knee System 2515-3115United Orthopedic Corp. · Prosthesis, Hip, Hemi-, Femoral, Metal2 recalls0 reports
- ACTIS 101011050Depuy (Ireland) · Prosthesis, hip, hemi-, femoral, metal73 reports
- UHR UH1-46-26Howmedica Osteonics Corp. · Prosthesis, hip, hemi-, femoral, metal1 recall11 reports
- UHR UH1-45-26Howmedica Osteonics Corp. · Prosthesis, hip, hemi-, femoral, metal1 recall11 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page