The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04546540638908), received between Jun 12, 2020 and Jun 12, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1

Patient problems

  • Fall1
  • Injury1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 12, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT'S RIGHT HIP WAS REVISED. SURGEON REPORTED HIS OPINION THAT THE PATIENT FELL, CAUSING THE STEM TO KNOCK THE ADM/ MDM POLY INSERT, PARTIALLY DISLODGING IT (NO DISLOCATION OR DISSOCIATION), AND CONTINUED TO IMPINGE ON ONE SIDE OF THE LINER, CAUSING WEAR. THE ADM/ MDM POLY INSERT AND 28MM METAL HEAD WERE REVISED TO ANOTHER ADM/ MDM POLY INSERT WITH A CERAMIC HEAD AND SLEEVE. REP CONFIRMED THERE ARE NO ALLEGATIONS AGAINST THE REVISED FEMORAL HEAD. REP CAN PROVIDE ADDITIONAL PICTURES AND CATALOG NUMBERS/ LOT CODES FOR THE DEVICES, AND CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Fall; Injury

    MAUDE report 10147521 ↗

Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices

All Prosthesis, hip, semi-constrained, metal/polymer, cemented devices →