Skip to content

Restoration 6276-1-025Recalls & reported problems

Howmedica Osteonics Corp. · Modular Hip System · Class 2 · Implant

All models of Restoration (86)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04546540641090), received between Jan 14, 2015 and Aug 18, 2025 — classified as 6 injuries.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Fracture2
  • Break1
  • Corroded1
  • Degraded1
  • Detachment of Device or Device Component1
  • Device Damaged by Another Device1

Patient problems

  • Ambulation or Postural Difficulties2
  • Bone Fracture(s)1
  • Foreign Body Reaction1
  • Insufficient Information1
  • Metal Related Pathology1
  • No Clinical Signs, Symptoms or Conditions1
  • No Code Available1
  • Pain1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 18, 2025Fracture; Fracture

    NO NEW INFORMATION.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22834516 ↗
  • InjuryApr 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SURGEON ADVISED IN THE SCRUB BAY THAT THE 'REVISION FROM THURSDAY WAS INFECTED'. WE HAD TAKEN SWABS DURING HIS REVISION SURGERY AND THEY HAD GROWN CULTURES. DOCTOR TELLS ME WE ARE NO LONGER DOING THIS PATIENTS SUBSEQUENT REVISION HIP REPLACEMENT. THE PLAN IS NOW TO MANAGE WITH ANTIBIOTICS AND NOT REVISE.

    Patient problems: Unspecified Infection

    MAUDE report 21890169 ↗
  • InjuryNov 26, 2024Corroded; Degraded; Corroded; Degraded

    IT IS REPORTED BY COUNSEL THAT CLAIMANT UNDERWENT RIGHT THA ON (B)(6) 2014, AND WAS IMPLANTED WITH A LFIT ANATOMIC V40 FEMORAL HEAD. SUBSEQUENTLY HE DEVELOPED PAIN IN HIS GROIN, LATERAL HIP AND BUTTOCKS WITH ANY MOTION AND FELT HE WAS GETTING WEAK AND GETTING WORSE. ALLEGEDLY HIS COBALT CHROME LEVELS WERE NOT SIGNIFICANTLY ELEVATED BUT MRI WAS CONCERNING FOR POSSIBLE ADVERSE LOCAL TISSUE REACTION. HE WAS REVISED ON (B)(6) 2021 FOR ALLEGED TRUNNIONOSIS AND HEAD/TAPER CORROSION.

    Patient problems: Foreign Body Reaction; Metal Related Pathology; Pain

    MAUDE report 20784923 ↗
  • InjuryMar 3, 2021Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE PATIENT'S RIGHT HIP WAS REVISED DUE TO THE HEAD BREAKING OFF THE TRUNNION. PATIENT IS ACTIVE AND INVOLVED WITH MANUAL LABOR. A FEMORAL HEAD AND RESTORATION MODULAR PROXIMAL BODY WERE REVISED. DUE TO INTRA-OPERATIVE DAMAGE TO THE STEM/ NECK JUNCTION, THE RESTORATION MODULAR DISTAL STEM WAS ALSO REVISED, ADDING 60 TO 90 MINUTES TO THE PROCEDURE (INTRA-OP EVENT REPORTED). REP CAN PROVIDE ADDITIONAL PICTURES AND OTHERWISE CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON. UPDATE 11/FEBRUARY/2021 WG: THE REP IS UNAWARE IF THE SHELL OR LINER, WHICH ARE COMPETITOR DEVICES, WERE REVISED.

    Patient problems: Ambulation or Postural Difficulties

    MAUDE report 11405370 ↗
  • InjuryNov 15, 2020Device Damaged by Another Device; Device Damaged by Another Device

    IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED DUE TO DISSOCIATION OF THE CONSTRAINED LINER FROM THE SHELL. SURGEON REPORTED THAT DUE TO IMPINGEMENT WITH THE STEM, THE LINER WAS LEVERED OUT OF THE SHELL. ALSO, THE SHELL WAS IN A NEUTRAL POSITION. A SHELL AND 2 SCREWS (IMPLANTED IN 2011), RESTORATION MODULAR PROXIMAL BODY (IMPLANTED IN 2013), LFIT FEMORAL HEAD AND CONSTRAINED LINER (BOTH IMPLANTED IN 2019) WERE REVISED. REP CONFIRMED THAT THERE ARE NO ALLEGATIONS AGAINST THE REVISED FEMORAL HEAD, AND THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON. CORRECTION (B)(6) 2021 WG: IN REPORTING A 5TH REVISION ON (B)(6) /2021 FOR THIS PATIENT'S LEFT HIP, REP CORRECTED INFORMATION FOR THE PATIENT'S PREVIOUS REVISION (THIS PI). A CONSTRAINED LINER AND FEMORAL HEAD WERE THE ONLY DEVICES REVISED.

    Patient problems: Ambulation or Postural Difficulties

    MAUDE report 10841562 ↗
  • InjuryJan 14, 2015Fracture; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT WAS BROUGHT TO OR DUE TO TIP OF THE GREATER TROCHANTER BROKE OFF ON A LEFT HIP. SURGEON SUPPORTED THAT WITH CABLES AND A PLATE.

    Patient problems: Bone Fracture(s); No Code Available

    MAUDE report 4416295 ↗

Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices

All Prosthesis, hip, semi-constrained, metal/polymer, cemented devices →