Skip to content

Restoration 6276-1-221Recalls & reported problems

Howmedica Osteonics Corp. · Modular Hip System · Class 2 · Implant

All models of Restoration (86)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04546540641212), received between Jan 12, 2018 and May 9, 2022 — classified as 7 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated2
  • Insufficient Device Problem Information2
  • Unstable2
  • Inadequacy of Device Shape and/or Size1
  • Material Erosion1
  • Unintended Movement1

Patient problems

  • Injury5
  • Ambulation or Postural Difficulties2
  • Joint Dislocation1
  • Joint Laxity1
  • Unequal Limb Length1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 9, 2022Unstable; Device Dislodged or Dislocated; Unstable; Device Dislodged or Dislocated

    THIS REPORT IS FOR THE (B)(6) 2022 REVISION. IN PROVIDING INFORMATION FOR A REVISION OF THE PATIENT'S LEFT HIP ON (B)(6) 2022 THE, REP CONFIRMED THAT ON (B)(6) 2022 THE HIP HAD BEEN REVISED AS WELL. AN MDM METAL LINER, ADM/ MDM POLY INSERT, FEMORAL HEAD, AND RESTORATION MODULAR PROXIMAL BODY WERE REVISED TO ANOTHER PROXIMAL BODY AND HEAD, AND A CONSTRAINED INSERT. UPDATE REASON FOR REVISION: "LEG LENGTH INEQUALITY AND INSTABILITY WITH HISTORY OF DISLOCATION. MDM HEAD DISLOCATION OUT OF MDM LINER."

    Patient problems: Ambulation or Postural Difficulties; Joint Laxity; Unequal Limb Length

    MAUDE report 14342958 ↗
  • InjuryAug 17, 2021Device Dislodged or Dislocated; Unintended Movement; Device Dislodged or Dislocated; Unintended Movement

    PATIENT IS A 57-YEAR-OLD FEMALE WHO PREVIOUSLY UNDERWENT RIGHT TOTAL HIP REPLACEMENT - OUTSIDE CORS SCOPE (GERMANY IN 2008). SHE UNDERWENT ACETABULAR REVISION - OUTSIDE CORS SCOPE (ROI 2011). PATIENT FELL AND HAD A PPFX OF HER RIGHT FEMUR (B2) UNDERWENT REVISION IN (B)(6) 2020. PATIENT DISLOCATED IN BED AND REQUIRED REVISION DUE TO INSTABILITY AND SUBSIDENCE OF FEMORAL COMPONENT PERFORMED IN (B)(6) 2020. MDM POLYETHYLENE, HEAD, AND RMS COMPONENTS EXCHANGED.

    Patient problems: Joint Dislocation

    MAUDE report 12334692 ↗
  • InjuryDec 23, 2019Material Erosion; Material Erosion

    IT WAS REPORTED THAT DURING REVISION OF RESTORATION MODULAR PROXIMAL BODY, THE HEAD AND TRUNNION EXHIBITED TRUNNIONOSIS.

    Patient problems: Injury

    MAUDE report 9512723 ↗
  • InjuryJun 25, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    SALES REP REPORTED THAT A PATIENT UNDERWENT A REVISION SURGERY DUE TO POSSIBLE TRUNNIONOSIS. THE ORIGINAL CASE TOOK PLACE IN 2008.

    Patient problems: Injury

    MAUDE report 8731816 ↗
  • InjuryFeb 21, 2019Unstable; Unstable

    THIS PI IS FOR THE REVISION OF THE PATIENT'S RIGHT HIP DUE TO INSTABILITY. IN REPORTING AN INTRA-OPERATIVE INSTRUMENT BREAKAGE, REP MENTIONED REVISED DEVICES. SPOKE TO REP. PATIENT'S RIGHT HIP WAS REVISED DUE TO INSTABILITY. SURGEON DID NOT REPORT ANY POSSIBLE CAUSE OR CONTRIBUTOR TO INSTABILITY. INTRA-OPERATIVELY, NOTHING WAS FOUND LOOSE, WORN, OR BROKEN. A 21 +20 RESTORATION MODULAR CONE BODY, SIZE 40 METAL HEAD, AND 40E POLY LINER WERE REVISED. REP REPORTED THAT X-RAYS, MEDICAL RECORDS, AND ADDITIONAL INFORMATION ARE NOT AVAILABLE.

    Patient problems: Injury

    MAUDE report 8358936 ↗
  • InjuryDec 12, 2018Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    IT WAS REPORTED THAT PATIENT'S RIGHT HIP WAS REVISED. AS REPORTED BY REP: "PATIENT APPEARED TO HAVE OFFSET ISSUES ON X-RAY". A 21MM +20 V40 RESTORATION PROXIMAL BODY AND 36 +0 BIOLOX V40 HEAD WERE REVISED TO A 25MM +20 V40 RESTORATION PROXIMAL BODY AND ANOTHER 36 +0 BIOLOX V40 HEAD.

    Patient problems: Injury

    MAUDE report 8156282 ↗
  • InjuryJan 12, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT PATIENT'S LEFT HIP WAS REVISED. EXPLANTED DEVICES ARE UHR UNIVERSAL HEAD, CONE BODY, AND LFIT V40 FEMORAL HEAD (BOTH IMPLANTED (B)(6) 2017). INITIAL SURGERY (B)(6)2017 WAS PERFORMED DUE TO A FAILED COMPETITOR IM HIP NAIL. 1ST AND 2ND REVISION WAS DUE TO NON-COMPLIANT PATIENT VS HIP INSTABILITY.

    Patient problems: Ambulation or Postural Difficulties; Injury

    MAUDE report 7186658 ↗

Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices

All PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page