Restoration 6276-5-011Recalls & reported problems
Howmedica Osteonics Corp. · Modular Hip System · Class 2 · Implant
All models of Restoration (86)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04546540641809), received between Dec 10, 2018 and Dec 10, 2018 — classified as 1 injury.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated1
Patient problems
- Injury1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryDec 10, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT PATIENT'S LEFT HIP WAS REVISED DUE TO DISLOCATION. THE HEAD DISLOCATED FROM THE LINER AS A RESULT OF THE PROXIMAL BODY DISASSOCIATING FROM THE PLASMA STEM AND PULLING THE HEAD OUT OF THE LINER. PATIENT HAD BEEN PERFORMING A NON-COMPLIANT ACTIVITY (BENDING OVER) WHEN DISLOCATION/ DISASSOCIATION OCCURRED. SURGEON REPORTED HE WAS UNSURE IF THE DISASSOCIATION OCCURRED BECAUSE OF SURGEON ERROR OR DEVICE FAILURE OF THE LOCKING MECHANISM. REP NOTED THAT DURING IMPLANTATION, SURGEON DID NOT USE TORQUE WRENCH, BUT APPLIED FORCE BY FREE-HAND TIGHTENING THE LOCKING MECHANISM. A BIOLOX CERAMIC HEAD AND RESTORATION PROXIMAL BODY WERE REVISED TO THE SAME CATALOG NUMBERS OF DIFFERENT LOT CODES AS THE SURGEON DECIDED NOT TO RE-IMPLANT THE REVISED DEVICES.
Patient problems: Injury
MAUDE report 8146433 ↗
Other Prosthesis, hip, hemi-, femoral, metal devices
- UHR UH1-48-28Howmedica Osteonics Corp. · Prosthesis, hip, hemi-, femoral, metal2 recalls3 reports
- "UNITED" USTAR II Knee System 2515-3111United Orthopedic Corp. · Prosthesis, Hip, Hemi-, Femoral, Metal2 recalls0 reports
- "UNITED" USTAR II Knee System 2515-3213United Orthopedic Corp. · Prosthesis, Hip, Hemi-, Femoral, Metal2 recalls0 reports
- STAGEONE 431207Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal2 recalls0 reports
- "UNITED" USTAR II Knee System 2515-3115United Orthopedic Corp. · Prosthesis, Hip, Hemi-, Femoral, Metal2 recalls0 reports
- ACTIS 101011050Depuy (Ireland) · Prosthesis, hip, hemi-, femoral, metal73 reports
- UHR UH1-46-26Howmedica Osteonics Corp. · Prosthesis, hip, hemi-, femoral, metal1 recall11 reports
- UHR UH1-45-26Howmedica Osteonics Corp. · Prosthesis, hip, hemi-, femoral, metal1 recall11 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page