Skip to content

Restoration 6276-5-011Recalls & reported problems

Howmedica Osteonics Corp. · Modular Hip System · Class 2 · Implant

All models of Restoration (86)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04546540641809), received between Dec 10, 2018 and Dec 10, 2018 — classified as 1 injury.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1

Patient problems

  • Injury1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 10, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT PATIENT'S LEFT HIP WAS REVISED DUE TO DISLOCATION. THE HEAD DISLOCATED FROM THE LINER AS A RESULT OF THE PROXIMAL BODY DISASSOCIATING FROM THE PLASMA STEM AND PULLING THE HEAD OUT OF THE LINER. PATIENT HAD BEEN PERFORMING A NON-COMPLIANT ACTIVITY (BENDING OVER) WHEN DISLOCATION/ DISASSOCIATION OCCURRED. SURGEON REPORTED HE WAS UNSURE IF THE DISASSOCIATION OCCURRED BECAUSE OF SURGEON ERROR OR DEVICE FAILURE OF THE LOCKING MECHANISM. REP NOTED THAT DURING IMPLANTATION, SURGEON DID NOT USE TORQUE WRENCH, BUT APPLIED FORCE BY FREE-HAND TIGHTENING THE LOCKING MECHANISM. A BIOLOX CERAMIC HEAD AND RESTORATION PROXIMAL BODY WERE REVISED TO THE SAME CATALOG NUMBERS OF DIFFERENT LOT CODES AS THE SURGEON DECIDED NOT TO RE-IMPLANT THE REVISED DEVICES.

    Patient problems: Injury

    MAUDE report 8146433 ↗

Other Prosthesis, hip, hemi-, femoral, metal devices

All Prosthesis, hip, hemi-, femoral, metal devices →

Sources and updates