Skip to content

Restoration 6276-7-015Recalls & reported problems

Howmedica Osteonics Corp. · Modular Hip System · Class 2 · Implant

Adverse event reports
9
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 04546540643735), received between Jun 14, 2018 and Mar 31, 2025 — classified as 9 injuries.

The most frequently reported specific device problem is “Fracture” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture6
  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Crack1
  • Insufficient Device Problem Information1
  • Loss of Osseointegration1
  • Osseointegration Problem1

Patient problems

  • Insufficient Information3
  • Pain3
  • Ambulation or Postural Difficulties2
  • Injury2
  • Failure of Implant1
  • Inadequate Osseointegration1
  • No Information1
  • Unspecified Infection1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 31, 2025Fracture; Fracture

    PATIENT REPORTED THAT THEY HAD A SECOND REVISION WITH STRYKER IMPLANTS ON (B)(6) 2024 FOR PAIN. THEY FURTHER STATED THAT IT WAS NOTED DURING THE PROCEDURE THAT THE TIP OF THE PROSTHETIC IN THE FEMUR HAD BROKEN OFF AND COME LOOSE. THE BROKEN PIECE AT THE VERY DISTAL END OF THE FEMORAL COMPONENT WAS LEFT IN THE PATIENT, RIGHT HIP.

    Patient problems: Pain

    MAUDE report 21730307 ↗
  • InjuryMar 18, 2025Fracture; Fracture

    IT WAS REPORTED BY THE PATIENT THAT ON (B)(6) 2014, HE UNDERWENT REVISION SURGERY AND WAS IMPLANTED WITH A RESTORATION MODULAR HIP SYSTEM. IN LATE ON (B)(6) 2024 HE STARTED TO EXPERIENCE EXTREME PAIN IN HIS RIGHT UPPER LEG FEMUR AREA AND HAD DIFFICULTY STANDING OR WALKING. HE WAS REVISED ON (B)(6) 2024, ALLEGEDLY DUE TO THE TIP OF THE PROSTHESIS BROKE OFF INSIDE HIS FEMUR.

    Patient problems: Ambulation or Postural Difficulties; Pain

    MAUDE report 21631315 ↗
  • InjuryJan 7, 2025Fracture; Fracture

    HIP REVISION SURGERY WAS REQUIRED DUE TO FRACTURE OF THE ROD IMPLANTED IN PRIMARY SURGERY. THERE WERE NO ADVERSE EVENTS IN THE REVISION SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 21089637 ↗
  • InjuryMay 16, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT HAD AN INDEX RIGHT THA ON (B)(6) 2022. THE PATIENT DEVELOPS PJI REQUIRING A RIGHT THA REVISION ON (B)(6) 2022. ALL COMPONENTS WERE EXCHANGED. REPORTED AS CORS PER 11001HIP. THE PATIENT REQUIRES A SECOND RIGHT THA REVISION DUE TO PERSISTENT PJI. ALL COMPONENTS WERE EXCHANGED.

    Patient problems: Unspecified Infection

    MAUDE report 16936894 ↗
  • InjuryJan 11, 2022Fracture; Fracture

    IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED DUE TO THE RESTORATION MODULAR DISTAL STEM BREAKING IN THE MIDDLE OF THE DISTAL STEM. SURGEON REPORTED A POSSIBLE CONTRIBUTING FACTOR BEING THE SELECTED STEM WAS TOO SMALL FOR THE PATIENT'S ANATOMY (INCORRECT SELECTION, NOT INCORRECT SIZE). THE MODULAR STEM AND FEMORAL HEAD WERE REVISED. REP CONFIRMED THERE ARE NO ALLEGATIONS AGAINST THE REVISED FEMORAL HEAD OR PROXIMAL BODY.

    Patient problems: Ambulation or Postural Difficulties; Insufficient Information

    MAUDE report 13222533 ↗
  • InjuryNov 4, 2020Fracture; Fracture

    RESTORATION MODULAR BROKEN STEM REQUIRING SURGERY. IN SURGERY ETO WAS REQUIRED TO GET STEM OUT. UPDATE: "GUN SHOT IN (B)(6) THAT REQUIRED A THA. HAD SEVERAL SURGERIES INCLUDING A 2 STAGE REVISION FOR INFECTION. CURRENTLY HAS STABLE LABS AND NO INFECTION AT TIME OF SURGERY. STEM IS BROKEN ABOUT 1.5¿ DOWN FROM THE BODY. LEFT HIP".

    Patient problems: Failure of Implant; Insufficient Information; No Information

    MAUDE report 10786964 ↗
  • InjuryOct 23, 2020Osseointegration Problem; Insufficient Device Problem Information; Osseointegration Problem; Insufficient Device Problem Information

    IT WAS REPORTED BY SCRUB TECHNICIAN THAT REVISION SURGERY PERFORMED DUE TO PATIENT FEEL PAIN IN THE HIP FOR A WHILE AFTER HAVING HIP RECONSTRUCTION SURGERY LONG TIME AGO, AFTER X-RAY, X-RAY FILM REVEALED THAT HIP COMPONENT WAS LOOSE. (STRYKER STAFF AWARENESS DATE IS THE REVISION SURGERY AS EXACT PRIMARY DEVICE TYPE IS UNKNOWN).

    Patient problems: Inadequate Osseointegration; Pain

    MAUDE report 10725267 ↗
  • InjuryApr 1, 2019Break; Break

    REVISION OF BROKEN RESMOD STEM. UPDATE PER SALES REP: "THE DISTAL STEM WAS BROKEN APPROXIMATELY 1/3 FROM THE END, THE BODY AND JUNCTION WERE NOT INVOLVED." RIGHT SIDE.

    Patient problems: Injury

    MAUDE report 8470312 ↗
  • InjuryJun 14, 2018Crack; Fracture; Loss of Osseointegration; Crack; Fracture; Loss of Osseointegration

    FAILED REVISION TOTAL HIP. PATIENT WAS BEING WATCHED FOR LOOSE STEM BY SURGEON AND WAS PLANNING ON REVISION UPON RETIREMENT. DUE TO STEM FAILURE REVISION WAS NECESSARY AT THIS TIME BY A DIFFERENT SURGEON.

    Patient problems: Injury

    MAUDE report 7602934 ↗

Other Prosthesis, hip, hemi-, femoral, metal devices

All Prosthesis, hip, hemi-, femoral, metal devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.