Skip to content

MDM 626-00-52HRecalls & reported problems

Howmedica Osteonics Corp. · Liner- Cementless · Class 2 · Implant

All models of MDM (6)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04546540666048), received between Nov 11, 2019 and Jul 28, 2025 — classified as 3 injuries.

The most frequently reported specific device problem is “Unstable” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unstable2
  • Device Dislodged or Dislocated1
  • Osseointegration Problem1

Patient problems

  • Joint Laxity2
  • Ambulation or Postural Difficulties1
  • Inadequate Osseointegration1
  • Injury1
  • Joint Dislocation1
  • Unequal Limb Length1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 28, 2025Unstable; Unstable

    AS REPORTED: "REVISION OF LEFT TOTAL HIP DONE TODAY. THIS PATIENT PRESENTED WITH A PERIPROSTHETIC FRACTURE OF HIS RIGHT HIP WHICH WAS COMPETITOR TOTAL HIP. ORIGINALLY THE LEFT HIP WAS NOT A PROBLEM. THE RIGHT HIP WAS REVISED ON (B)(6) 2025 TO ADDRESS THE FRACTURE USING COMPETITOR PRODUCTS. AFTER THAT SURGERY IT WAS DISCOVERED THAT THE LEFT HIP HAD DISLOCATED. THE LEFT HIP WAS REDUCED AND TRANSFERRED FROM TRAUMA SERVICE TO JOINT REPLACEMENT SERVICE. IT WAS DISCOVERED THAT THE 28MM HEAD HAD DISASSOCIATED FROM THE MDM POLY LINER. IT IS THOUGHT THAT THIS DISASSOCIATION OCCURS WHEN THE HIP WAS REDUCED. THE REVISION TODAY WAS TO ADDRESS THE INSTABILITY OF THE LEFT HIP. MDM IMPLANTS WERE REMOVED AND A CONSTRAINED LINER WAS PLACED. IT WAS ALSO DECIDED TO GO FROM A +4 NECK LENGTH TO A +8 TO ADDRESS LIMB LENGTH DISCREPANCY. NO FURTHER INFORMATION AVAILABLE FROM SURGEON OR HOSPITAL." REP CONFIRMED THAT INTRA-OPERATIVELY, THE POLY WAS FOUND FLOATING IN THE JOINT.

    Patient problems: Joint Dislocation; Joint Laxity

    MAUDE report 22638120 ↗
  • InjuryMay 9, 2022Unstable; Device Dislodged or Dislocated; Unstable; Device Dislodged or Dislocated

    THIS REPORT IS FOR THE (B)(6) 2022 REVISION. IN PROVIDING INFORMATION FOR A REVISION OF THE PATIENT'S LEFT HIP ON (B)(6) 2022 THE, REP CONFIRMED THAT ON (B)(6) 2022 THE HIP HAD BEEN REVISED AS WELL. AN MDM METAL LINER, ADM/ MDM POLY INSERT, FEMORAL HEAD, AND RESTORATION MODULAR PROXIMAL BODY WERE REVISED TO ANOTHER PROXIMAL BODY AND HEAD, AND A CONSTRAINED INSERT. UPDATE REASON FOR REVISION: "LEG LENGTH INEQUALITY AND INSTABILITY WITH HISTORY OF DISLOCATION. MDM HEAD DISLOCATION OUT OF MDM LINER."

    Patient problems: Ambulation or Postural Difficulties; Joint Laxity; Unequal Limb Length

    MAUDE report 14342959 ↗
  • InjuryNov 11, 2019Osseointegration Problem; Osseointegration Problem

    ON (B)(6) 2016: RAS IMPLANTED. COMPLEX REVISION FOR PELVIC DISCONTINUITY. PATIENT HAD 4 REVISIONS PRIOR TO THIS, HISTORY OF RA. ON (B)(6) 2018: PATIENT SEEN IN CLINIC AND REPORTED THAT SHE FELT HIP SHIFT AND DISLOCATE. LOOSENING OF ACETABULAR CUP IDENTIFIED ON X-RAY. RADIOLOGY REPORT: SLIGHT DISTRACTION OF THE ACETABULAR SCREWS IN COMPARISON TO THE PREVIOUS EXAMINATION THAT WOULD BE CONSISTENT. ON (B)(6) 2019: REMOVAL OF ACETABULUM IMPLANT AND FEMORAL HEAD EXCHANGE TO BIPOLAR IMPLANT. PATIENT FACTORS: SEDENTARY, WHEELCHAIR BOUND, HISTORY OF RECURRENT DISLOCATION. COMPETITOR STEM AND HEAD NOTED IN PER PRODUCT GRID.

    Patient problems: Inadequate Osseointegration; Injury

    MAUDE report 9307403 ↗

Other Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate devices

All Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate devices →