Skip to content

Excelsior SL-10 M0031681810Recalls & reported problems

Stryker Corporation · 1 Tip Straight Microcatheter · Class 2

All models of Excelsior SL-10 (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04546540688200), received between Jun 21, 2019 and Mar 19, 2025 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Peeled/Delaminated” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Peeled/Delaminated3
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1
  • Perforation of Vessels1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 19, 2025Peeled/Delaminated; Peeled/Delaminated

    ANALYSIS OF THE RETURNED DEVICE REVEALED THAT SUBJECT CATHETER PTFE POLYTETRAFLUOROETHYLENE INNER LINING WAS PEELING. THERE WAS NO CLINICAL CONSEQUENCE TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21644335 ↗
  • MalfunctionNov 20, 2024Peeled/Delaminated; Peeled/Delaminated

    THE CATHETER (SUBJECT DEVICE) WAS RETURNED FOR ANALYSIS AND WAS EXAMINED. DURING DEVICE INVESTIGATION AND ANALYSIS, IT WAS DISCOVERED THAT CATHETER (SUBJECT DEVICE) PTFE INNER LINING WAS IDENTIFIED IN THE SHAFT. THE PHYSICIAN REPLACED IT WITH A NEW DEVICE AND CONTINUED THE PROCEDURE WITHOUT CLINICAL CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20735713 ↗
  • InjuryMar 9, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PROCEDURE OF THROMBECTOMY FOR LEFT M1-DISTAL WAS PERFORMED ON (B)(6) 2020. THE MODIFIED RANKIN SCALE (MRS) OF THE PATIENT BEFORE THE PROCEDURE WAS 2. THE ALBERTA STROKE PROGRAM EARLY CT SCORE (ASPECTS) BEFORE PROCEDURE WAS 8. THE THROMBOLYSIS IN CEREBRAL INFARCTION (TICI) WAS 0 BEFORE PROCEDURE. THE NATIONAL INSTITUTES OF HEALTH STROKE SCALE (NIHSS) SCORE BEFORE THE PROCEDURE WAS 26. A RETRIEVER AND ASPIRATION CATHETER WERE USED TO COMPLETE THE 1ST AND 2ND PASSES. HOWEVER, THE MODIFIED TREATMENT IN CEREBRAL INFARCTION (MTICI) WAS NOT IMPROVED AND REMAINED 0 AFTER THE 2ND PASS. 3RD PASS WAS PERFORMED USING A MICROCATHETER (SUBJECT DEVICE), STENT RETRIEVER AND A GUIDEWIRE. VESSEL PERFORATION OCCURRED DURING THE 3RD PASS; HOWEVER, THE PHYSICIAN COULD NOT IDENTIFY WHICH DEVICE CAUSED THE PERFORATION. AFTER THE PROCEDURE, THE FINAL TICI WAS 2A AND NIHSS WAS 30. THE MRS SCORE BECAME WORSE TO 4 ON (B)(6) 2020 AND THE MRS SCORE BECAME WORSE TO 5 ON (B)(6) 2021. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Perforation of Vessels

    MAUDE report 11445808 ↗
  • MalfunctionJun 21, 2019Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT DURING A PROCEDURE, WHEN A SURGEON WAS REMOVING A MANDREL FROM THE SUBJECT DEVICE MICROCATHETER, THE RESISTANCE WAS FELT AND NOTICED THAT THE COATING WAS PEELING. ANOTHER NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO CLINICAL CONSEQUENCES TO THE PATIENT DUE TO THE REPORTED EVENT

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8724071 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →