Excelsior SL-10 M0031681920Recalls & reported problems
Stryker Corporation · 2 Tip Pre-Shaped J Microcatheter · Class 2
All models of Excelsior SL-10 (7)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04546540688248), received between Aug 6, 2018 and Sep 10, 2026 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Peeled/Delaminated” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Peeled/Delaminated3
- Adverse Event Without Identified Device or Use Problem1
- Component Missing1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Thrombosis1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 10, 2026Peeled/Delaminated
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50714424 ↗MalfunctionApr 13, 2026Peeled/Delaminated
THE SUBJECT CATHETER WAS RETURNED FOR ANALYSIS AND THE DEVICE INVESTIGATION REVEALED THAT THE SUBJECT CATHETER PTFE POLYTETRAFLUOROETHYLENE INNER LINING WAS PEELING. THERE WERE NO CLINICAL CONSEQUENCES TO THE PATIENT REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24866000 ↗MalfunctionApr 22, 2025Peeled/Delaminated; Peeled/Delaminated
IT WAS REPORTED THAT THE GUIDEWIRE ENCOUNTERED STRONG RESISTANCE WHEN ADVANCED INTO THE SUBJECT CATHETER HUB. THE PHYSICIAN FOUND OUT THE SUBJECT CATHETER INNER LUMEN HAS SOME PROBLEM (THE COATING IS DAMAGED). THE SUBJECT DEVICE WAS REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21878817 ↗MalfunctionOct 20, 2021Component Missing; Component Missing
IT WAS REPORTED THAT THE SUBJECT MICROCATHETER WAS INSPECTED AND CONFIRMED TO BE IN GOOD CONDITION DURING/AFTER UNPACKING AS DESCRIBED IN DEVICE DIRECTIONS FOR USE (DFU). AFTER THE SUBJECT MICROCATHETER WAS PLACED IN THE PATIENT, ADVANCE ALONG VESSEL COULD NOT BE FOLLOWED ON THE ANGIOGRAPHY DEVICE. THE PHYSICIAN THINKS MICROCATHETER HAS NO MARKER. THEREFORE, THE SUBJECT MICROCATHETER WAS REMOVED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THE PATIENT WAS STABLE AFTER THE PROCEDURE. NO OTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12668796 ↗InjuryAug 6, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CASE WAS RUPTURED C2-3 ANEURYSM STENT ASSISTED COIL EMBOLIZATION, THE STENT WAS DEPLOYED WITHOUT ANY PROBLEMS. THE STENT DELIVERY WIRE (SDW) WAS ATTEMPTED TO BE WITHDRAWN BUT THE SDW AND THE STENT GOT STUCK EACH OTHER. THE MICROCATHETER (SUBJECT DEVICE) WAS ADVANCED TO RELEASE THE STICKING, BUT IT ALSO STUCK WITH THE SDW. THE HUB OF THE MICROCATHETER WAS CUT OFF AND SNARE DEVICE WAS NAVIGATED TO THE SDW VIA THE MICROCATHETER OUTER SHAFT. THE SNARE DEVICE CAUGHT THE SDW, THEN THE SDW AND THE MICROCATHETER WERE REMOVED FROM THE PATIENT WHILE INVOLVING A SLIGHT STENT DISPLACEMENT. THE REMOVED SDW TIP WAS FOUND STRETCHED AND SEPARATED. COILING WAS STARTED, BUT A DISTAL BLOOD FLOW OF THE ANEURYSM CAME NOT TO BE SHOWN IN ANGIOGRAPHY DURING COILING. CROSS FLOW FROM OPPOSITE SIDE WAS SUFFICIENT, SO THE PROCEDURE WAS FINISHED WITHOUT RECANALIZATION OF THE DISTAL BLOOD FLOW. THE PATIENT'S CONDITION IS BRAIN DEATH DUE TO THE PREEXISTING SUB ARACHNOID HEMORRHAGE WORSENING.
Patient problems: Thrombosis
MAUDE report 7754655 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗